SHORT COMMUNICATION

Comparison of optical-power meters between the NIST and the PTB

, , , and

Published under licence by IOP Publishing Ltd
, , Citation I Vayshenker et al 2000 Metrologia 37 349 DOI 10.1088/0026-1394/37/4/15

0026-1394/37/4/349

Abstract

We describe the results of a comparison of optical-power meters undertaken by the National Institute of Standards and Technology (NIST, USA) and the Physikalisch-Technische Bundesanstalt (PTB, Germany) at nominal wavelengths of 1300 nm and 1550 nm. Both laboratories used thermal detectors as reference standards, which were compared using a germanium trap detector as a transfer standard. Measurement results showed differences of less than 1 part in 103, well within the combined uncertainty for both laboratories.