A publishing partnership

ARE CORONAL LOOPS ISOTHERMAL OR MULTITHERMAL?

, , , , , , , and

Published 2009 January 15 © 2009. The American Astronomical Society. All rights reserved.
, , Citation J. T. Schmelz et al 2009 ApJ 691 503 DOI 10.1088/0004-637X/691/1/503

0004-637X/691/1/503

ABSTRACT

Surprisingly few solar coronal loops have been observed simultaneously with TRACE and SOHO/Coronal Diagnostics Spectrometer (CDS), and even fewer analyses of these loops have been conducted and published. The SOHO Joint Observing Program 146 was designed in part to provide the simultaneous observations required for in-depth temperature analysis of active region loops and determine whether these loops are isothermal or multithermal. The data analyzed in this paper were taken on 2003 January 17 of AR 10250. We used TRACE filter ratios, emission measure loci, and two methods of differential emission measure analysis to examine the temperature structure of three different loops. TRACE and CDS observations agree that Loop 1 is isothermal with log T = 5.85, both along the line of sight as well as along the length of the loop leg that is visible in the CDS field of view. Loop 2 is hotter than Loop 1. It is multithermal along the line of sight, with significant emission between 6.2 < log T< 6.4, but the loop apex region is out of the CDS field of view so it is not possible to determine the temperature distribution as a function of the loop height. Loop 3 also appears to be multithermal, but a blended loop that is just barely resolved with CDS may be adding cool emission to the Loop 3 intensities and complicating our results. So, are coronal loops isothermal or multithermal? The answer appears to be yes.

Export citation and abstract BibTeX RIS

1. INTRODUCTION

Standard models for quiescent coronal loops (e.g., Rosner et al. 1978) assumed a single magnetic flux tube where the plasma was in hydrostatic equilibrium and had a uniform cross section. These models predicted a set of well known scaling laws and a significant loop temperature gradient. Actual measurements of temperature and density are vital for understanding the physics of loops, and observations by the Coronal Diagnostics Spectrometer (CDS) on SOHO have produced a variety of results. For example, the loops analyzed by Del Zanna (2003) and Del Zanna & Mason (2003) were isothermal along the line of sight and had a temperature gradient along their length, but the one analyzed by Brkovic et al. (2002) was isothermal in both directions. The loop analyzed by Schmelz et al. (2001) and Schmelz & Martens (2006) was multithermal along the line of sight, with a temperature distribution that increased as a function of loop height.

The Brkovic et al. (2002) result agrees with those obtained by Lenz et al. (1999) using TRACE data. This triggered a resurgence of modeling analysis as the observed loops had substantially smaller temperature gradients than expected. The Schmelz et al. (2001) results support nanoflare models and require us to refine our view of coronal loops. Throughout this paper, we refer to a loop as a coherent structure seen in an observation and a strand as a magnetic flux tube for which the heating and plasma properties have an approximately uniform cross section. Nanoflare models (e.g., Klimchuk & Cargill 2001) as well as these CDS results indicate that some or perhaps even many loops are composed of an unknown number of strands.

The data used in this analysis were acquired as part of SOHO JOP 146, designed to provide simultaneous TRACE and CDS observations of active region coronal loops and address the apparent controversy that arose from results such as those discussed above. These results may, however, be related to the fact that different types of loops were observed, but only additional observations and analyses can confirm this. The observations are described in Section 2, data analysis in Section 3, results in Section 4, and conclusions in Section 5.

2. OBSERVATIONS

Cirtain et al. (2007) give details of the SOHO JOP 146 data set, including timelines, calibration, coalignment, as well as the initial stages of analysis. TRACE (Handy et al. 1999) is a Cassegrain telescope with a 30 cm aperture and a field of view of 8farcm5 × 8farcm5. It images the corona using three EUV passbands centered at 171, 195, and 284 Å. There are also UV passbands that can image the photosphere and the transition region, but these data were not used in this analysis. TRACE has state-of-the-art pixels (0farcs5) and a spatial resolution of about of 1''. It is capable of producing one EUV image per minute.

CDS (Harrison et al. 1995) is a Wolter-Schwartschild II grazing incident telescope, with a full geometric area of 289.28 cm2 and an effective focal length of 257 cm. A light beam is focused at an entrance slit and then split in two, with one component feeding into the Normal Incidence Spectrometer (NIS) while the other beam is directed to the Grazing Incidence Spectrometer (GIS). Since all data for this project were taken exclusively from the NIS component, the GIS will not be discussed further. NIS data are structured into cubes by scanning the slit across the Sun from west to east onto a 1024 × 1024 pixel detector. The instrument has 1farcs68 pixels and a spatial resolution of about 5''–6''. These observations used the 4'' slit. For each slit position, two spectra are generated at each pixel, the first covering the wavelength range from 310 to 380 Å, and the second covering 517–633 Å. Thus a unique spectrum exists for each scan position and each pixel along the slit. NIS rarely transmits data from the entire wavelength range back to Earth due to telemetry limitations. Instead, small wavelength bands, usually centered on strong emission lines, are transmitted.

Two types of CDS rasters were taken as part of SOHO JOP 146: context rasters and sparse rasters. The 4'' × 240'' slit was used in both cases. The context raster field of view was 240'' × 240''. It was composed of 60 exposures, each with duration of 60 s for a total accumulation time of approximately 85 minutes. For the sparse rasters, exposures were not taken at every possible slit position, but instead with 12'' or 16'' steps in between each exposure position. This significantly reduced the time required to complete a raster. The total field of view was 102'' × 240'', and data were collected at fixed positions on the solar disk every 7 minutes.

The data analyzed in this paper were taken on 2003 January 17 of AR 10250. The start time for the CDS observations was 06:51:27 UT. The observing sequences ran for 7 hr and the context rasters were interspersed with the sparse rasters. TRACE images were collected in all three EUV filters. The 171 and 195 Å images were taken at full resolution. The 284 Å images were binned to decrease the exposure time, but unfortunately, were not useful for the present analysis.

3. ANALYSIS

Some of the JOP 146 data have been analyzed by Cirtain et al. (2007) and Schmelz et al. (2007b). Cirtain et al. (2007) used an emission measure (EM) loci technique (Jordan at al. 1987) to determine where and when the background-subtracted loop plasma was isothermal. Their analysis used many CDS lines observed in 26 sparse rasters taken from 10:43 to 16:45 UT on 2003 January 17. They found that about half of the pixels investigated were consistent with isothermal plasma, but that these results change with both space and time. Some (but not all) pixels along the loop were isothermal some of the time.

Schmelz et al. (2007b) used this data set to investigate various concerns related to the temperature analysis of CDS data as well as the inherent assumptions made in differential emission measure (DEM) techniques. They used the CDS context raster from 14:24 UT to address the relatively slow cadence and a large pixel size of the CDS instrument, the inherent smoothing required in all automatic inversion DEM techniques, as well as limb versus disk solar loop observations. They found that CDS DEM analysis was reliable and provided meaningful physical results for coronal loops. This paper is an extension of these two previous analyses.

In coronal equilibrium, where upward transitions result from collisions between electrons and ions and downward transitions are spontaneous, the intensity I of an optically thin spectral line of wavelength λij can be written as

Equation (1)

where h is the Plancks constant, c is the speed of light, A is the elemental abundance, T is the electron temperature, G(T, ne) is the contribution function, ne is the electron density, l is the line-of-sight element, and L is the depth of the observing region. If the emitting plasma can be characterized by a single temperature and density, the contribution function is taken out of the integral, which is then referred to as the emission measure (EM), the amount of emitting material along the line of sight:

Equation (2)

The EM loci method uses the observed intensity of each spectral line and plots the EM as a function of temperature (for different values of the electron density). If all the resulting curves intersect at a single point (within the uncertainties), then the plasma may be isothermal and the approximation is justified. If, however, the curves do not intersect, then the plasma is multithermal and the analysis becomes more complex. In this case, we need to define the DEM, Φ(T), the amount of emitting material as a function of temperature along the line of sight:

Equation (3)

and

Equation (4)

The DEM problem is mathematically ill posed so small changes in the observed intensities can produce large variations in the DEM curves, and a vast range of solutions can reproduce the spectra within arbitrarily small observational errors. As a result, DEM solutions have been criticized severely in the literature and rightly so (see e.g., Judge & McIntosh 1999). Enforcing a positive solution helps a bit, and no one disputes the fact that although a purely mathematical solution can have negative values, a DEM curve representing astrophysical plasma must be strictly positive. Many DEM methods also rely on smoothing. Again, in a purely mathematical solution, adjacent temperature bins can be independent of each other, but these solutions might not be physically meaningful. Different smoothing methods, however, often result in different DEM results. For an detailed discussion of DEM uncertainties and smoothing, see Kashyap & Drake (1998).

To get around these problems, authors sometimes impose ad hoc assumptions such as the overly simplistic two-Gaussian temperature distribution for the loop strands modeled by Aschwanden et al. (2007). Artificial limits on the DEM temperature range are also employed when the resulting DEM curves are not constrained by the data itself. Excluding temperature bins, however, can artificially limit the range of possible solutions. The complicating factor is the intrinsically high dynamic range of the DEM curves. A suitably large combination of emissivity and DEM can easily overcome the small instrument response. Del Zanna et al. (2003) describe a good example of this analysis for polar plumes. Schmelz et al. (2007a) show just how large the DEM uncertainties can be for TRACE loop data when no ad hoc assumptions are imposed. These results caution against such analyses as they tend to over interpret the data.

This paper uses two methods to evaluate the DEM based on the CDS data. The first uses the forward folding technique with a manual manipulation of the curve as described by Schmelz et al. (1999). Although this method is time consuming, it forces the user to understand both the limitations on the data and the assumptions going into the analysis. The user has control of the final DEM shape and no smoothing is required beyond that imposed by the resolution of the contribution functions (0.1 dex). The second method uses the automatic inversion technique described by Warren (2005) and Brooks & Warren (2006). The DEM curve is represented with a series of spline knots that are repositioned interactively for more control over the smoothness of the DEM curve. This method represents the best of both worlds: the quickness of automatic inversion and the control of manual manipulation. In both cases, the best fit is determined from a χ2 minimization of the differences between the intensities observed by CDS and those predicted by Equation (4).

4. RESULTS

Our results used the context rasters from 14:24 UT mentioned in the previous section to analyze three loops that were part of AR 10250. The emission lines used in this study are listed in Table 1. These range from Mgvi at 349.2 Å with a peak formation temperature of log T = 5.6 to Fe xix at 592.2 Å with a peak formation temperature of log T = 6.80, although no photons were detected at these high temperatures. Figure 1 shows the CDS field of view in a series of spectral lines with different peak formation temperatures. Figure 2 shows the pixels along each loop as well as the background pixels chosen for detailed analysis. We used version 5.1 of the CHIANTI atomic physics database (Dere et al. 1997; Landi et al. 2006) and the hybrid coronal abundances (Fludra & Schmelz 1999).

Figure 1.

Figure 1. CDS images taken on 2003 January 17 at 06:51:27 UT of AR 10250 in order of increasing peak formation temperature: (a) Mg vi (log T = 5.6); (b) Ne vi (log T = 5.6); (c) Ca x (log T = 5.85); (d) Mg ix (log T = 6.0); (e) Mg x (log T = 6.05); (f) Si x (log T = 6.15); (g) Fe xiv (log T = 6.3); (h) Si xii (log T = 6.3); and (i) Fe xvi (log T = 6.4). These are negative images.

Standard image High-resolution image
Figure 2.

Figure 2. Subset of the CDS images from Figure 1 showing the loop pixels and background pixels chosen for detailed analysis. (a) CDS Mg ix at 368.1 Å with log T = 6.0 showing the pixels for Loop 1; (b) CDS Si xii at 520.7 Å with log T = 6.3 showing the pixels for Loop 2 (right) and Loop 3 (left). These are negative images.

Standard image High-resolution image

Table 1. CDS Spectral Lines

λ (Å) Ion log T Atomic   Transition Locations Where Line Is Present
349.164 Mg vi 5.60 2s2 2p3 2D5/2 ... 2s 2p4 2D5/2      
558.592 Ne vi 5.60 2s2 2p 2P1/2 ... 2s 2p2 2D3/2 Loop 1    
562.798 Ne vi 5.60 2s2 2p 2P3/2 ... 2s 2p2 2D5/2 Loop 1    
557.766 Ca x 5.85 3s 2S1/2 ... 3p 2P3/2 Loop 1 Ftpt 2  
368.070 Mg ix 6.00 2s2 1S0 ... 2s 2p 1P1 Loop 1 Ftpt 2 Loop 3
349.860 Si ix 6.05 2s2 2p2 3P2 ... 2s 2p3 3D1,2,3 Loop 1 Ftpt 2 Loop 3
624.943 Mg x 6.05 1s2 2s 2S1/2 ... 1s2 2p 2P1/2   Ftpt 2 Loop 3
347.409 Si x 6.15 2s2 2p 2P1/2 ... 2s 2p2 2D3/2   Ftpt 2 Loop 3
356.030 Si x 6.15 2s2 2p 2P3/2 ... 2s 2p2 2D3/2,5/2   Ftpt 2 Loop 3
346.853 Fe xii 6.15 3s2 3p3 4S3/2 ... 3s 3p4 4P1/2   Ftpt 2 Loop 3
352.107 Fe xii 6.15 3s2 3p3 4S3/2 ... 3s 3p4 4P3/2   Ftpt 2 Loop 3
364.468 Fe xii 6.15 3s2 3p3 4S3/2 ... 3s 3p4 4P5/2   Loop 2 Loop 3
348.184 Fe xiii 6.20 3s2 3p2 3P0 ... 3s 3p3 3D1   Loop 2 Loop 3
550.032 Al x 6.20 1s2 2s 2S1/2 ... 1s2 2p 2P3/2   Loop 2 Loop 3
334.180 Fe xiv 6.30 3s2 3p 2P1/2 ... 3s 3p2 2D3/2   Loop 2 Loop 3
353.837 Fe xiv 6.30 3s2 3p 2P3/2 ... 3s 3p2 2D5/2   Loop 2 Loop 3
520.666 Si xii 6.30 1s2 2s 2S1/2 ... 1s2 2p 2P1/2   Loop 2 Loop 3
335.410 Fe xvi 6.40 3s 2S1/2 ... 3p 2P3/2   Loop 2 Loop 3
360.759 Fe xvi 6.40 3s 2S1/2 ... 3p 2P1/2   Loop 2 Loop 3
592.236 Fe xix 6.80 2s2 2p4 3P2 ... 2s2 2p4 1D2      

Download table as:  ASCIITypeset image

4.1. An Isothermal Loop

We begin our analysis with the loop seen in Figure 2(a). The footpoint area of this loop is visible in Mg vi at 349.2 Å with a peak formation temperature of log T = 5.6, but the main leg of the loop is best seen in Mg ix at 368.1 Å with a peak formation temperature of log T = 6.0, and fades in lines with higher temperatures. We chose six pixels along the loop for detailed analysis. The CDS temporal resolution is relatively slow and there is often concern that the target loop could be evolving while the CDS slit steps across the region. This is not a problem for Loop 1, however, because it is orientated vertically, that is, along the CDS slit. In this configuration, all the pixels analyzed for Loop 1 were observed at about the same time and can therefore be compared directly with each other.

Proper background subtraction is an essential step in the analysis, and a lot of work was done investigating potential background pixels or collection of pixels. The background structure varies from position to position and from spectral line to spectral line. A potential background pixel might appear dark at one wavelength and not at another. There are also loops of varying temperatures close to each other so cuts across the loops do not reveal the true background. These cuts intersect other loops, which are visible in maps made in different spectral lines with different peak formation temperatures. That is why CDS is such a valuable tool for this type of analysis.

The best loop-background contrast for Loop 1 (in Mg ix) was about 2.7:1, so we knew background subtraction was going to be challenging. The crowded field added to the difficulties. In our original analysis, we selected an individual background pixel for each loop pixel. Each pixel pair was at approximately the same y-coordinate, but the background pixel was about 10 pixels to the right of the loop pixel. This method gave us results that seemed inconsistent with the loop appearance: we got significant values for background-subtracted intensities for lines where the loop appeared invisible. We then modified our original approach and followed the method of Cirtain et al. (2007). We used a single background pixel for loop pixels A–D, which appear to have a background with a minimum structure, and a second background pixel for loop pixels E–F, which appear to be superimposed on a background of mass. This approach subtracted too much flux from several of the pixels, and we finally settled on a single background pixel (see Figure 2). This method gave us the results that were most consistent with the loop appearance in the different spectral lines.

Our results for Loop 1 are shown in Figure 3. The left column shows the EM loci plots for each pixel, starting at the footpoint and moving up the loop leg (which corresponds to pixels with lower y-values in Figures 1 and 2). These plots can be used to determine if the background-subtracted loop plasma is isothermal. The plots in the middle column show the DEM analysis done with forward folding and a manual manipulation of the curve while those in the right column show the DEM analysis done with automatic inversion.

Figure 3.

Figure 3. Temperature results for the six pixels in Loop 1 after background subtraction, including EM loci plots (left), DEM curves with forward folding (middle), and DEM curves with automatic inversion (right). The plus signs represent the ratio of the observed to calculated radiance scaled by the emission measure and plotted at the peak formation temperature. Arrows indicate lines that fall below the y-axis minimum, but still give an excellent fit. Note: these symbols are not included for the plots in the middle column because the spike DEM shape can fall between the data points, changing all the plus signs to arrows, and giving the incorrect impression of a bad fit.

Standard image High-resolution image

The EM loci plots and the forward-folding DEM curves indicate that the loop plasma at these positions is isothermal with log T = 5.85. The automatic-inversion DEM curves are a bit broader, but have the same peak temperature. The discrepancies are due in part to the inherent differences in the methods, but also to the smoothing incorporated into the automatic inversion. Our results confirm those of Schmelz et al. (2007b) who showed that the forward-folding method adds no artificial width to the DEM curve beyond the temperature resolution element imposed by the steps taken in the spectral line response functions (log T = 0.1). However the automatic inversion DEM method produces a natural temperature width related to the inherent smoothing associated with the inversion, so it is not possible to obtain a result that represents a truly isothermal plasma.

It is unfortunate that there are no density-sensitive ion pairs remaining after background subtraction. Figure 4 plots what information we have about density. It shows the predicted (from forward folding) to observed intensity ratios for densities of (a) 108, (b) 109, and (c) 1010 cm−3. We did this analysis for all of the pixels and show a typical example. We chose to display the results for pixel C because it was far enough away from the footpoint and in the middle of the distribution of loop pixels. Unfortunately, most of the lines are not density sensitive in this range, but the Si x line indicates that the plasma density is unlikely to be as low as 108 cm−3, but 109 and 1010 look equally good. Since the loop apex is well out of the CDS field of view, it is not possible to say with certainty that this loop is overdense, but assuming that the apex temperature is the same as that for the positions analyzed here, then the loop is indeed overdense with respect to results expected from scaling laws (see e.g., Warren & Winebarger 2003).

Figure 4.

Figure 4. Loop 1 density analysis. The predicted (from forward folding) to observed intensity ratios for the spectral lines observed in pixel C for (a) 108, (b) 109, and (c) 1010 cm−3. The number in the upper-right corner of each panel is the reduced χ2.

Standard image High-resolution image

The isothermal results for Loop 1 are consistent with the TRACE data (Figure 5). The flux from the ensemble of TRACE pixels corresponding to the CDS background pixel is subtracted from the flux of the ensembles for the various loop pixels in both the 195 and 171 Å TRACE images. We propagated the TRACE uncertainties through the pixel-averaging process, background subtraction, and 195/171 division. We found that the TRACE ratios for the various Loop 1 pixels were not statistically different. The results are shown in Figure 6, which plots the 195/171 response as a function of log T. The solid flat lines represent the flux ratios for the background-subtracted pixels, where the spread is a measure of the uncertainty. The intersections are very close to a local minimum in the response ratio, and all the results are consistent with the CDS temperatures of log T = 5.85. This shows that the TRACE image ratio technique works well if the plasma is isothermal, but see the next two subsections for different results (the dashed lines are for Loop 3).

Figure 5.

Figure 5. TRACE 195 Å image showing pixels chosen for detailed analysis for Loops 1 (left) and 3 (right). This is a negative image.

Standard image High-resolution image
Figure 6.

Figure 6. TRACE 195/171 response function ratio vs. log T. The flat lines represent the flux ratios for the background-subtracted pixels. The solid lines are for Loop 1 and the dashed lines are for Loop 3. The spread in each set is a measure of the uncertainty. The intersections of the solid lines are all close to a local minimum with T = 0.7 ± 0.2 MK (log T = 5.85), which is consistent with the isothermal CDS result for Loop 1. The intersections for the dashed lines are at T = 1.1−1.2 MK, consistent with the multithermal CDS result for Loop 3 (see the text).

Standard image High-resolution image

4.2. A Multithermal Loop

We continue our analysis with the loop in Figure 2(b) (right). The footpoint area of this loop is visible in almost every spectral line in the data set. This indicates to us that perhaps many individual loops with different temperatures or temperature distributions are anchored at this position. As we move up the leg of the loop, we see that it remains visible in all the hotter lines (log T > 6.1), but is not visible in the cooler lines. It is most distinct in Si xii at 520.7 Å with a peak formation temperature of log T = 6.3, and is still visible in the two Fe xvi lines, but not in Fe xix where a significant intensity would most likely indicate the flaring plasma. As with Loop 1, we chose several pixels along the loop for detailed analysis. Note that Loop 2 is also orientated along the CDS slit, so all the pixels were observed at about the same time and can be compared directly with each other.

The best loop–background contrast for Loop 2 (in Si xii) was about 2:1, so background subtraction was even more challenging than it was for Loop 1. We first chose the same background pixel, but all the spectral lines, including Si xii where the loop was clearly visible, had insignificant fluxes after background subtraction. We had subtracted too much flux. We experimented with different individual pixels and with the pixel-pair method that had not worked with Loop 1. We settled on a single pixel (see Figure 2(b), right); as with Loop 1, this approach seemed the most consistent with the loop appearance in the different spectral lines.

Our results for Loop 2 are shown in Figure 7. The format is the same as that of Figure 3; from left to right, we have the EM loci plots for each pixel, the forward-folding DEM results, and the automatic-inversion DEM results. These results are different from those for Loop 1: the curves for the different spectral lines do not intersect at a single point in the EM loci plots, indicating that the background-subtracted loop plasma is not isothermal. The two DEM methods agree (with this result and with each other). A multithermal distribution is required to reproduce the observed spectral line intensities. As mentioned above, the footpoint pixel seen in (A) may be an area where many individual loops with different temperatures or temperature distributions are anchored. This could account for the very broad distribution. For the loop-leg pixels (B)–(F), the cool temperature end of the DEM distributions are strongly constrained since the loop is not seen in any lines with peak formation temperature log T< 6.15. The high-temperature constraint is somewhat weaker. The loop is visible in Fe xvi (log T = 6.40) but not in Fe xix (log T = 6.80), so the DEM curves could be a bit wider than those shown in Figure 7, but not narrower.

Figure 7.

Figure 7. Temperature analysis. Same as Figure 3, but for Loop 2.

Standard image High-resolution image

This broad DEM result is reminiscent of the analysis of a loop on the limb observed by CDS on 1998 April 20 and analyzed by Schmelz et al. (2001) and Schmelz & Martens (2006). That loop was also most distinct in the Si xii line at 520.7 Å, but there was concern that the broad DEM might be the result of the long path length for a loop on the limb. Loop 2 appears to have similar properties, but as is obvious from the images in Figure 1, it is situated on the solar disk. Of course it is still possible that the broad DEM distributions seen in Figure 7 are the result of loops of different temperatures along the line of sight and unresolved by the moderate pixel size of CDS, but these distributions could also result from a single loop composed of isothermal strands of different temperatures.

Our density analysis for Loop 2 is shown in Figure 8. The format is similar to that of Figure 4 for Loop 1, showing the predicted (from forward folding) to observed intensity ratios for the spectral lines detected in pixel C for densities of (a) 108, (b) 109, and (c) 1010 cm−3. The two coolest lines (Al xi and Fe xii) and the three hottest lines (Si xii and the two Fe xvi lines) are not sensitive to density in this range, but the intermediate lines (Fe xiii and the two Fe xiv lines) are. These lines indicate that the plasma density is unlikely to be as high as 1010 cm−3 or as low as 108 cm−3. The best-fit density was 4 × 108 cm−3.

Figure 8.

Figure 8. Density analysis. Same as Figure 4, but for Loop 2.

Standard image High-resolution image

Unfortunately, Loop 2 does not show up in the TRACE images. This is simply a matter of bad timing. During the lifetime of Loop 2, TRACE was imaging with the 171 and 284 Å channels. The response of the former is too cool and the confusion in the latter for this particular active region was too high.

4.3. A Curved Loop

The last loop selected for analysis is shown in Figure 2(b) (left). This loop is seen in many spectral lines with a wide range of temperatures and it is the most distinct loop structure in the TRACE image. The best loop–background contrast for Loop 3 (in Si xii) was about 2.3:1, and background subtraction was just as challenging as it was for Loops 1 and 2. As before, we eventually settled on a single pixel because this approach seemed the most consistent with the loop appearance.

Our results for Loop 3 are shown in Figures 9 and 10. The format is the same as that of Figures 3 and 7. The EM loci plots show that the curves for the different spectral lines do not intersect at a single point, indicating that the background-subtracted loop plasma is not isothermal and that a multithermal distribution is required. However, we found two different DEM models that could reproduce the observed spectral line intensities. One is very broad (Figure 9) and the other has two more narrow components (Figure 10). This result shows that the DEM loop analysis is not always clean and that in this particular case, the CDS data are not good enough to distinguish between these models.

Figure 9.

Figure 9. Temperature analysis. Same as Figure 3, but for Loop 3 (broad DEM model).

Standard image High-resolution image
Figure 10.

Figure 10. Temperature analysis. Same as Figure 3, but for Loop 3 (two-component model).

Standard image High-resolution image

Our density analysis for Loop 3 is shown in Figure 11. The format is similar to that of Figures 4 and 8, showing the predicted (from forward folding) to observed intensity ratios for the spectral lines observed in pixel C for densities of (a) 108, (b) 109, and (c) 1010 cm−3. As with Loop 2, these lines indicate that the plasma density is unlikely to be as high as 1010 cm−3 or as low as 108 cm−3, but 109 cm−3 looks quite good. The best-fit density was 3 × 109 cm−3.

Figure 11.

Figure 11. Density analysis. Same as Figure 4, but for Loop 3.

Standard image High-resolution image

The TRACE analysis for Loop 3 is shown in Figure 6. As with Loop 1, the flux from the ensemble of TRACE pixels corresponding to the CDS background pixel is subtracted from the flux of the ensembles for the various loop pixels in both the 195 and 171 TRACE images. The dashed lines represent the flux ratios for the background-subtracted pixels, and the intersections with the response ratio range from T = 1.1−1.2 MK. This result may appear inconsistent with that of CDS, but a possible explanation was provided by Weber et al. (2005). If an isothermal approximation is used when the plasma is actually multithermal, then the resulting 171 Å–195 Å ratio will be 0.81. This value corresponds to the ratio of the areas under the 171 and 195 Å response curves and implies a plasma temperature of 1.2 MK in the isothermal approximation. In other words, a TRACE temperature of 1.2 MK is consistent with multithermal plasma, and therefore, with the results obtained with CDS.

5. CONCLUSIONS

We have analyzed three different loops that were observed with CDS and TRACE and obtained three different results. EM loci plots and DEM with forward folding indicate that the loop plasma at each chosen pixel of Loop 1 appears to be isothermal with log T = 5.85. DEM curves made with automatic inversion (Figure 3) and TRACE 195/171 image ratios (Figure 6) are consistent with this result. We also find that all the pixels along the leg of Loop 1 appear to have the same temperature, a result consistent with image ratio analysis (e.g., Neupert et al. 1998; Lenz et al. 1999) and inconsistent with results predicted by standard models for loops in hydrodynamic equilibrium (e.g., Rosner et al. 1978). Although CDS has moderate-size pixels, it has been used by other authors to find isothermal plasma in loops (e.g., Brkovic et al. 2002; Del Zanna 2003; Del Zanna & Mason 2003), in polar plumes (Del Zanna et al. 2003), and in an area of quiet Sun off the limb (Allen et al. 2000). The results for Loop 1 just add to this list.

Loop 2 gives a different result. DEM curves made with both forward folding and with automatic inversion indicate that the loop plasma at each pixel is multithermal. The EM loci plots are consistent with this result. This is similar to the conclusions of Schmelz et al. (2001) and Schmelz & Martens (2006) for a loop observed on the solar limb with CDS. It may be a coincidence, but both of these loops were best seen in the Si xii line, one of hotter (but nonflaring) lines available in the CDS wavelength range. This could be a filling-factor effect (Brickhouse & Schmelz 2006), meaning that there is simply more material along the line of sight at log T = 6.3 than there is at log T = 5.85, or it could be a consequence of loop cooling (Winebarger et al. 2003). Patsourakos et al. (2002) showed that slightly different initial conditions for cooling loops in an arcade were sufficient to reproduce qualitatively the fuzzy appearance of EUV post-flare loops seen for example in the TRACE 284 Å channel.

Results for Loop 3 are not as clear. The observed CDS intensities can be reproduced with a very broad DEM curve (Figure 9) or a two-component model (Figure 10). We favor the two-component model based on the CDS images. Figure 1(c) shows the active region structure in Ca x (log T = 5.85). There is a cool loop slightly offset from Loop 3, but not completely resolved with the CDS instrument. We suspect that cooler emission from this Ca x loop is blended with Loop 3 and contributing to the cooler DEM component seen in Figure 10. Subtracting the emission making up the cool component from the individual spectral line intensities (which we now assume come from the Ca x loop) leaves the Loop 3 intensities alone which can then be modeled with a single component (although a bit broader) at a temperature centered near log T = 6.3.

Although CDS has in many cases detected isothermal plasma, the moderate spatial resolution (5''–6'') of the instrument has long been a concern for loop studies. The analysis of these three loops illustrates different strengths and weaknesses. The CDS observations clearly discerned the isothermal nature of Loop 1, even though the field is crowded with other loops and plagued with a complicated background structure. The broader DEM components such as those required for Loop 2 need to be confirmed using data from spectrometers with better spatial resolution. Problems such as the apparent blending of the Ca x and Loop 3 structures, which are not quite cospatial in the CDS images, will also be resolved with future instruments.

We thank Harry Warren for the use of his DEM program, Amy Winebarger for installing this program on our computers, and University of Memphis physics students James Andrews and Jennifer Garst for help with analysis. Solar physics research at University of Memphis is supported by NSF ATM-0402729 and NASA NNG05GE68G. H.E.M. and G.D.Z. acknowledge financial support from PPARC. We are grateful to P. R. Young and the entire CDS team for their help in obtaining these observations. We also benefited greatly from discussions at the coronal loops workshops in Paris (2002 November), Palermo (2004 September), and Santorini (2007 June).

Please wait… references are loading.
10.1088/0004-637X/691/1/503