Introduction

X-ray orbital radiography (plain and computed tomography (CT) imaging) to detect and/or localize intraocular foreign bodies (IOFBs) is commonly performed on ophthalmic patients with a history of exposure to high-velocity particles. The booklet ‘Making the Best Use of a Department of Clinical Radiology’ issued by the Royal College of Radiologists (MBUR RCR)1 states that a single ‘soft’ lateral X-ray is the only projection required to exclude a metallic foreign body, and that images with eye movement are indicated only in those cases where the intraocular position of a radiologically proven foreign body is uncertain. These guidelines also suggest that CT orbital imaging in cases or suspected cases of IOFB should be undertaken only when the plain X-ray orbital radiograph fails to show a strongly suspected foreign body, which may not be metallic, when multiple foreign bodies are present, or when it remains uncertain whether a radiologically proven foreign body is intraocular. The MBUR RCR guidelines aim to limit patients' exposure to radiation and to reduce non-contributory yet costly examinations in the Accident and Emergency and/or Radiology Departments.

Bray and Griffith2 reported that plain X-rays of the orbits are unnecessary in patients exposed to high-velocity particles where there is no clinical evidence of ocular penetration, and suggested that plain X-ray orbital radiography should be restricted to those patients with clinical evidence of ocular penetration. They did not evaluate the role of CT orbital imaging in cases and suspected cases of IOFB.

The purpose of this study was to evaluate the roles of plain X-ray and CT orbital radiography in patients where an IOFB is present or suspected.

Methods

The records of all patients who underwent X-ray orbital radiography (plain films±CT imaging) between August 2001 and July 2002, where an IOFB was present or suspected, in the Radiology Department of the Royal Victoria Eye and Ear Hospital, Dublin, Ireland, were retrieved. The data recorded for each subject can be classed as demographic, clinical, radiological, and therapeutic.

A standard plain X-ray orbital radiograph was taken according to a strict protocol, which included a non-screened lateral view, with occiptomental eye-moving images obtained only if the initial view was suggestive of an IOFB.

In compliance with the MBUR RCR guidelines,1 CT orbital imaging was undertaken only in cases or suspected cases of IOFB following a plain orbital X-ray radiograph. All scans were performed on a spiral CT scanner (Siemens Somatom) using a 3 mm slice thickness and a pitch of 1 : 1. All the images (plain X-ray or CT) were reported by one of two consultant radiologists.

Results

Two hundred and ten patients with an IOFB or suspected IOFB underwent X-ray orbital radiography (plain X-ray±CT imaging) during the study period. Of these, the medical records were available in 204 cases (197 males and 7 females). The mean (±SD) age was 39 (±16.8) years, with a range of 7–91 years.

The clinical details and radiological findings are given, for cases where plain X-ray orbital radiography and CT orbital imaging were undertaken, in Tables 1 and 2, respectively. In brief, of the 21 CT orbital scans undertaken, 11 (52.4%), 8 (38.1%), and 2 (9.5%) were performed in patients in whom an IOFB was seen, was not seen, and was indeterminate, respectively, on plain X-ray orbital radiography. Sensitivity, specificity, and predictive values of plain X-ray and CT orbital radiography in cases and suspected cases of IOFB are given in Table 3.3

Table 1 Clinical details and radiological findings in 204 cases and suspected cases of intraocular foreign body that underwent plain X-ray orbital radiography
Table 2 Clinical details and radiological findings in 21 cases and suspected cases of intraocular foreign body that underwent CT orbital imaging
Table 3 Sensitivity, specificity, and predictive values of plain X-ray and CT orbital radiography in cases and suspected cases of IOFB

Discussion

X-ray orbital radiography, whether plain X-ray or CT imaging, is commonly performed on ophthalmic patients with a history of exposure to high-velocity particles and a suspected or clinically visible IOFB. The sensitivity of plain orbital X-rays in these cases has been the subject of investigation in the past, with previous reports indicating that radiopaque foreign bodies are detected by plain X-ray radiography in 70–90% of cases.2, 4 Of note, failure of plain X-ray to detect an IOFB is a well-recognized phenomenon, and reflects the variable sizes, compositions, and locations of IOFBs.5, 6, 7

Interestingly, 87% of the plain orbital X-rays undertaken in this study were performed in the absence of clinically evident ocular penetration, with a yield of 0 (0%), consistent with an earlier report by Bray and Griffith.2 Of the 27 (13%) plain X-ray orbital radiographs performed in patients with clinically evident ocular penetration following exposure to high-velocity particles, 19 (70%) exhibited a clinically visible IOFB. In these cases, therefore, the clinical findings were simply confirmed by plain X-ray orbital radiography, and the contribution to the management of these cases by such imaging rests on its documentary value and on its ability to exclude multiple IOFBs. Furthermore, in eight cases of clinically confirmed ocular penetration following exposure to high-velocity particles, in the absence of a clinically visible IOFB, plain X-ray orbital radiography did furnish the clinician with radiological evidence of an IOFB in seven (87%) cases.

Previous reports have shown that CT orbital imaging is superior to orbital plain X-ray orbital radiography for the detection and localization of IOFBs, although CT is associated with greater exposure to radiation than plain X-ray radiography.8, 9, 10 In contrast to plain X-ray orbital radiography, the majority (12; 57%) of orbital CT scans in our study were performed on patients with clinically evident ocular penetration. However, five (42%) of these cases had a clinically visible IOFB and radiological confirmation of the IOFB on plain X-rays of the orbit. In these cases, multiple foreign bodies were not evident on the plain X-ray orbital radiographs, and the value of CT images in this setting is therefore questionable, given that localization of the IOFB was achieved clinically.

Similar to plain X-ray orbital radiography, CT findings correlated well with clinical findings, in that all CT orbital images that were performed on eyes exposed to high-velocity particles with clinically evident ocular penetration, whether or not an IOFB was clinically visible, provided radiological evidence of an IOFB, whereas no CT scans performed in the absence of clinically evident ocular penetration following exposure to high-velocity particles showed radiological evidence of an IOFB.

The MBUR RCR guidelines stipulate that CT imaging should be undertaken only when the preceding plain X-ray orbital radiograph fails to show a strongly suspected foreign body, which may not be metallic, when multiple foreign bodies are present, or when it is not certain whether a foreign body already demonstrated is intraocular.1 Our experience suggests that, where there is clinical evidence of ocular penetration but an IOFB is not clinically visible, CT imaging would be necessary irrespective of the result of the preceding plain X-ray orbital radiograph. This is so for two reasons. First, CT imaging would identify an IOFB that might have gone undetected on orbital plain X-ray. Second, when an IOFB is demonstrated on the plain X-ray orbital radiograph, CT imaging would be necessary for its accurate localization.

It could be argued that plain X-ray orbital radiography before CT imaging acts as a guide to the radiologist whether 3 or 6 mm thick CT sections are to be taken. Six millimetre scans, which are associated with less radiation dosage than 3 mm scans, would be sufficient to detect and localize an IOFB that is visible on the plain X-ray orbital radiography.8 Modern CT scanners (multidetector CT scanning (MCT)) preclude the need for such measures because of their ability to obtain much thinner scans (as thin as 1 mm) with much less radiation exposure than conventional scans.11, 12 The logistical burden of arranging pre-CT plain X-ray orbital radiography, and having such images reviewed, may adversely affect the speedy management of patients, especially in eye-dedicated units that may be remote from radiological services.

Conversely, where a metallic IOFB is clinically visible, plain X-ray orbital radiography without subsequent CT orbital imaging may serve to ensure that multiple IOFBs are not present. However, in the presence of a reliable history that excludes multiple IOFBs, the value of such imaging is questionable.

In conclusion, a thorough ophthalmic examination, with particular attention directed towards the likelihood of ocular penetration, and including gonioscopy (where appropriate) and detailed dilated fundoscopy remains the mainstay of management of ophthalmic patients exposed to high-velocity particles. The results of this series confirm other authors' conclusions2 that patients without clinical evidence of ocular penetration do not need to undergo orbital imaging of any kind. However, such a recommendation applies only to those patients who are exposed to high-velocity particles in the recent past, where clinical evidence of ocular penetration (eg subconjunctival haemorrhage) is not likely to have resolved. Where a metallic IOFB is clinically visible, plain X-ray orbital radiography without subsequent CT orbital imaging may be required to ensure that multiple IOFBs are not present, before proceeding to surgery (Figure 1). Given that in no case did plain X-ray of the orbit detect an IOFB that went undetected on subsequent CT imaging, and that the detection of an IOFB on plain X-ray radiography does not contribute to the decision-making process with respect to proceeding to CT imaging where there is a clinically evident ocular penetration but an IOFB is not clinically visible, it would appear that the standard imaging guidelines by the RCR need to be amended with respect to the stipulation that CT imaging must be preceded by plain X-ray orbital radiography.1 Finally, if a non-radiopaque IOFB is still strongly suspected following negative CT imaging, other imaging modalities such as ocular ultrasound or magnetic resonance imaging may need to be considered.4, 13, 14

Figure 1
figure 1

Flow diagram of the proposed X-ray (plain X-ray and/or CT imaging) orbital radiography pathway in cases and suspected cases of intraocular foreign body (IOFB).