Dental unit water systems (DUWS) are generally contaminated with bacteria.7,9,13,15 The normal method of measuring bacterial contamination is by total viable counts (TVC) following culture on microbiological media in a dedicated laboratory. This is a time-consuming and labour intensive process, and there is an inevitable delay before the data are available and can be acted upon. In order for the bacterial counts to be as accurate as possible, it is essential to ensure that the samples are processed without delay after collection. This may prove difficult if the site of collection is remote from the microbiology laboratory.

In the brewing and food industry, bacterial and other cellular contamination is estimated by measuring total adenosine triphosphate (ATP).4,5 Hygiene monitoring using ATP also offers a simple and valuable means of monitoring dental practice cleaning routines.6 This test is relatively simple to perform and can be done at the site of sampling. The simplicity and rapidity of the ATP test system are features that could potentially make this a feasible approach to measure contamination of DUWS at the chairside. This would be particularly useful in confirming the efficacy of decontamination procedures.

Another potential method of estimating TVC could be to measure endotoxin concentration. Much of the bacterial contamination of DUWS is by Gram-negative bacteria, which on lysis would release endotoxin. Thus, endotoxin estimation could potentially also be used as a measure of bacterial contamination of DUWS. This assay is used extensively in the pharmaceutical industry for testing the production of endotoxin-free water.

The purpose of this study was to:

  • Evaluate the use of rapid, commercial kits to estimate ATP

  • Assess a laboratory-based endotoxin assay

  • Determine whether the data generated by ATP and endotoxin analysis would correlate with TVC in DUWS, and

  • Investigate if they have potential as simpler approaches to determine levels of DUWS contamination.

Materials and methods

Water samples from DUWS

For a pilot study, ten water samples from a general dental practice were analysed in order to establish the relationship between TVC and total cell counts. Subsequently, in the study of the water in GDPs, 57 water samples were collected aseptically from both the triple spray handpiece and the air rotor connector in 25 primary care surgeries in South West England. The sterile sample bottles contained 0.1 g of sodium thiosulphate to neutralise any chlorine present in the sample (Abinghurst Ltd, UK). The water samples were transported in a cool box to the laboratory and were tested within 3 h of collection.

TVC of water samples

A water sample of 100 ml was passed through a 0.45 μm cellulose acetate filter (Techware, Poole, UK). The filter was then placed in a sterile container, containing 10 ml of sterile water, and agitated vigorously to re-suspend any organisms trapped on the filter surface. Ten-fold dilutions from neat to 10−2 were made of the supernatant fluid in PBS containing sodium thiosulphate (3.5 g per l) and 0.1 ml of each dilution was then surface plated onto Yeast Extract agar (Biomerieux, UK) for total viable counts (TVC) and MacConkey agar (Biomerieux, UK) for coliform counts. All plates were then incubated at 37°C for 72 h. Colonies were counted using a plate counter and the TVC for each sample was calculated. The level of detection was 10 CFU ml−1.

Total microbial cell count

Samples taken during the pilot study were analysed in counting chambers under light microscopy using the improved Neubauer, Kova and Thoma slides. In addition, the total number of microbial cells present in 1.0 ml of each water sample were also analysed using flow cytometry (Beckman Coulter Epics XL) with fluorescent beads as a background control. Bacterial and bead populations were separated on a forward scatter v side scatter plot. Maximum total events were set at 15,000. Data were collected for each sample to determine cell numbers per ml.

ATP assays

ATP was measured in the original water sample, using the Aqua-Trace ™ system (Biotrace, Cardiff, UK). This detects and quantifies luminescence of a luciferase/ATP reaction in a photometer and gives a value in relative light units (RLU) proportional to the quantity of ATP present. The device is in the form of a plastic swab that contains reagents to lyse intact cells to release the ATP and is designed to sample the correct volume of water for the reaction. The swab is then placed back into the plastic container and pushed to the bottom of the tube which contains the reagents for the reaction. The tube is then shaken gently and placed into the portable Uni-Lite ™ reading device (Biotrace, Cardiff, UK) which measures the amount of light released by the reaction, and the value is printed out. Before each batch of tests was performed the Uni-Lite device was calibrated with no Aqua-Trace reaction to measure any background luminescence (<10 RLU). The detection limit of the system is approximately 103 CFU ml−1.

Endotoxin assay

Separate water samples were collected into plastic endotoxin- and pyrogen-free universal containers. The samples were filtered through a 0.45 μm cellulose acetate filter (Sartorius Ltd., UK), which removes intact bacteria but allows free endotoxin to remain in suspension, and stored at −4°C until tested. The water samples were analysed using a Limulus Amoebocyte Lysate (LAL) gel-clot reagent (Charles River Laboratories) and recorded in endotoxin units ml−1 (EU ml−1). Initially, the filtered water samples were screened at 150 EU ml−1 by diluting the samples 1:600. Any positive reactions were further diluted to give sensitivities of 300, 600 and 1,200 EU ml−1. Those samples that gave negative reactions were reanalysed to test at sensitivities of 25, 50 and 100 EU ml−1. These were titrated against a control standard Escherichia coli endotoxin of known potency (0.25 EU ml−1) (Charles River Laboratories). Tests were incubated at 37°C for 60 minutes; positive reactions were recorded for those tubes that contained a firm gel-clot that remained in place when the tube was inverted. All tests and dilutions were performed in endotoxin-free glass test tubes (Charles River Laboratories) using LAL reagent water (<0.001 EU ml−1) (Charles River Laboratories). Each batch of tests contained both positive and negative controls and the method used for the assay meant that the sensitivity for the test was > 25 EU ml−1.

Statistical analysis

Statistical analysis was carried out using Pearsons (rank-order) Correlation Coefficient Test.

Results

Pilot study

Microbial numbers were too low in the water samples to obtain reliable cell counts using the conventional counting chambers and therefore flow cytometry was used to assess the total cell count more accurately. Total microbial counts as measured by flow cytometry were generally higher than TVC by 50 to 500 fold, with extremes of 7-, 1,925- and 10,709-fold (Table 1). TVC exceeded 100 CFU ml−1 in 9/10 samples taken during the pilot study. One sample was from a stagnant unit (number 5, Table 1) and resulted in both a high TVC, total count and ATP. The ATP assay was shown to have good reproducibility when replicate samples from the same water source were analysed (Table 2).

Table 1 Comparison of total viable counts (TVC), total cell counts (as measured by flow cytometry), and ATP on ten water samples taken from three GDP DUWS
Table 2 Multiple ATP analyses of four water samples to evaluate the reproducibility of Aquatrace Bioluminescence systems

Practitioner study

The TVC ranged from not detectable to 2.16 × 104 CFU ml−1 (mean = 1.3 × 103) (Fig. 1). None of the samples produced growth on MacConkey agar, while 10 samples gave no colonies on Yeast Extract agar (Fig. 1). The ATP assay was quick and easy to use, and values ranged from 22 to 958 RLU. The automatic print-out meant that there was a record of the data. In contrast the endotoxin assay test was semi-quantitative and required specialist laboratory glassware and equipment to perform. Results took at least one hour to obtain and for economic reasons were best performed on a batch basis. Only 11/57 samples had detectable levels of endotoxin (range = 0–300 EU ml−1); these positive samples were not related to high TVCs or the presence of coliforms.

Figure 1
figure 1

Scatter plot of TVC versus ATP from 50 water samples from GDP DUWS

Analysis of the relationship between TVC and ATP analysis was performed using Microsoft Excel (Fig. 1). Statistical analysis using Pearsons (rank-order) Correlation Coefficient Test indicated that the correlation, at 0.006, was not significant (p = 0.98) and that the correlation, at 0152, between the TVC and total cell counts in the pilot study was also not significant (p = 0.676).

Discussion

A recent study in the UK has shown that as many as 93% of GDP DUWS would fail to deliver water that would conform to EU potable water standard of 100 CFU ml−1.15 In addition, 87% of the units sampled would have failed to meet current ADA DUWS standards of 200 CFU ml−1.2 Chairside tests would offer the opportunity to both monitor water quality and assess the efficacy of any adopted preventative regime. Conventional microbiology is time-consuming, expensive and inevitably there is a lag between sample taking and provision of results, during which time patients will continue to be treated using water that may not fulfil published guidelines.

The majority of the organisms found in the planktonic phase of DUWS are derived from a biofilm that forms on the internal surfaces of tubing within the dental unit.3,10,14,15 The accurate recovery of viable microorganisms from aquatic systems is problematical. For example, it has been reported that it may only be possible to recover 5% of the organisms from DUWS by conventional cultural techniques12 with recovery from other ecosystems being as low as 0.001%.1 Therefore, the extent of the contamination of DUWS may be underestimated when based on TVC. Our pilot study confirmed this, and found the ratio of cultured/total cell counts to be in the same range as that reported for other environmental aquatic ecosystems.1 This may be one reason why the TVC, ATP and endotoxin assay did not correlate. The problem in culturing all bacteria has led to the proposal that ATP levels may be more representative of the contamination of the water than the TVC. However, there were samples where high TVC were found, but only a relatively low ATP value was recorded. This may be due to the rapid degradation of ATP by the release of cellular ATP'ases. More work is required to explain the lack of direct correlation between TVC and ATP before this approach could be adopted.

The endotoxin assay method used in this study was relatively expensive and required specialist expertise and equipment and therefore may not be practical for use in the dental surgery. Very few of the samples showed detectable levels of endotoxin and, in the majority of those that did, the level was found to be low. Other workers have reported comparable endotoxin levels in DUWS and have attributed clinical significance to their findings.11 However, endotoxin levels of up to 300 EU ml−1 have been found in the saliva of normal healthy subjects.8 It is unlikely, therefore, that endotoxin concentrations of <300 EU ml−1 will be of any clinical relevance.

The ATP assay method used here was found to be very simple to use, was portable and gave virtually instant results (within one minute) making it very suitable for use at the chairside. However, statistical analysis of the data showed no statistical correlation between TVC and ATP, while a large proportion of water samples with TVC greater than the EU potable water standard (100 CFU ml−1) gave low ATP results, making it impossible at present to set a discriminatory cut off point using this approach. Although the ATP method worked better at high TVC (ie as an indicator of gross microbial contamination) and gave good reproducibility in replicate samples, routine use of this assay may yield a large number of unacceptable false negative values (Fig. 1). Thus, although the ATP method has a number of advantages, in its present format it is too insensitive to be recommended for the routine monitoring of DUWS microbial contamination.

Conclusion

ATP levels and endotoxin levels were not found to correlate with TVCs in water derived from dental unit water systems. The ATP assay was simple to perform and could be suitable for chairside use if the sensitivity and specificity issues could be overcome. Culturable methods underestimate the total number of microbial cells in water samples but currently represents the most reliable method available for assessing microbial contamination of dental unit water lines.