Abstract
The Pacific Northwest National Laboratory has developed an Automated Radioxenon Sampler/Analyzer (ARSA) in support of the Comprehensive Nuclear-Test-Ban-Treaty (CTBT) to measure four radioxenon isotopes: 131mXe, 133mXe, 133gXe, and 135gXe. This system uses a beta-gamma coincidence counting detector to produce two-dimensional plots of gamma-energy versus beta-energy. Betas and conversion electrons (CE) are detected in a cylindrical plastic scintillation cell and gamma and X-rays are detected in a surrounding NaI(Tl) scintillation detector. The ARSA has been field tested at several locations to measure the radioxenon concentrations. Most recently it has been deployed at the Institut für Atmosphärische Radioaktivität in Freiburg, Germany. During the first 4 months of 2000 the measured 133Xe oncentrations have varied between 0.0±0.1 and 110±10 mBq/m3 air. The longer lived 131mXe (T 1/2 = 11.9 d) and short lived 135Xe (T 1/2 = 9.1 h) have also been detected in small quantities, while 133mXe concentrations have been consistent with zero. Minimum detectable concentration (MDC) calculations for 133gXe fell well below the 1 mBq per standard-cubic-meter of air requirement adopted by the CTBT Preparatory Commission.1 A description of the radioxenon detector, the concentration and MDC calculations and preliminary results of the field test in Germany are presented.
Similar content being viewed by others
References
Operational Manual for Radionuclide and the International Exchange of Radionuclide Data, Prepatory Commission for Comprehensive Nuclear-Test-Ban-Treaty Organization, CTBT/WGB/TL-11/5/Rev. 4, 1999.
J. C. Hayes, K. H. Abel, Th. W. Bowyer, T. R. Heimbigner, M. E. Panisko, P. L. Reeder, J. I. McIntyre, R. C. Thompson, L. C. Todd, R. A. Warner, 21st Seismic Research Symp.: Technologies for Monitoring CTBT, Las Vegas, NV, Sep. 21-24, 1999, PNNL-SA 32064.
T. W. Bowyer, K. H. Abel, C. W. Hubbard, M. E. Panisko, P. L. Reeder, R. C. Thompson, R. A. Warner, J. Radioanal. Nucl. Chem., 240 (1999) 109.
E. Browne, R. B. Firestone, Table of Radioactive Isotopes, John Wiley and Sons, Inc., New York, 1986.
ENDF/B-VI, National Nuclear Data Center, Brookhaven, NY.
P. L. Reeder, T. W. Bowyer, R. W. Perkins, J. Radioanal. Nucl. Chem., 235 (1998) 89.
P. L. Reeder, T. W. Bowyer, Nucl. Instr. Meth. Phys. Res., A408 (1998) 582.
T. W. Bowyer, J. I. McIntyre, P. L. Reeder, High sensitivity detection of Xe isotopes via beta-gamma coincidence counting, 21st Seismic Research Symp.: Technologies for Monitoring CTBT, Las Vegas, NV, Sep. 21-24, 1999, PNNL-SA-32048.
K. M. Biegalski, Proposal for Release 3 Formats and Protocols: Full Incorporation of Noble Gas Data into Pulse Height Data Messages, Prototype International Data Centre, Rev. 10, October 1999.
P. L. Reeder, T. W. Bowyer, J. I. McIntyre, W. K. Pitts, J. Radioanal. Nucl. Chem., 248 (2001) 617.
L. A. Curie, Analytical Chemistry, 40 (1968) 586.
J. I. McIntyre, Th. W. Bowyer, P. L. Reeder, Calculations of Minimum-Detectable-Concentrations Levels of Radioxenon Isotopes Using the PNNL ARSA System, PNNL-13102, Internal Report, Pacific Northwest National Laboratory, Richland, WA, 1999.
K. M. Foltz Biegalski, S. R. Biegalski, J. Radioanal. Nucl. Chem., 248 (2001) 673.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
McIntyre, J.I., Abel, K.H., Bowyer, T.W. et al. Measurements of ambient radioxenon levels using the automated radioxenon sampler/analyzer (ARSA). Journal of Radioanalytical and Nuclear Chemistry 248, 629–635 (2001). https://doi.org/10.1023/A:1010672107749
Published:
Issue Date:
DOI: https://doi.org/10.1023/A:1010672107749