Determining the Artifact Sensitivity of Recent Pulse Oximeters during Laboratory Benchmarking

  • Published:
Journal of Clinical Monitoring and Computing Aims and scope Submit manuscript

Abstract

This study aimed to investigate and compare the performance of the algorithms contained in the newest generation of pulse oximeters (Masimo SET® in IVY2000, Nellcor Oxismart® N-3000, Agilent M3rev. B) against a traditional pulse oximeter (Agilent CMS rel. A.0). The benchmark was performed in an efficient and reproducible way in the laboratory environment using patient signal recordings complemented by a two-hand volunteer motion test. A novel method of creating artifact/reference signal pairs from a clinical database, the noise-mix-composition (NMC), was developed. The new method enabled the simulation of critical clinical situations in a more realistic way than the usual two hand volunteer studies. An advantage of the laboratory tests over live clinical studies was that a continuoussaturation reference was available, allowing accurate on-going determination of the SpO2 error. A new quantitative performance measure, the non-performance index (NPI), was developed and applied to the benchmark results. It covers the 3 performance aspects of a pulse oximeter: (1) SpO2 accuracy, (2) pulse rate accuracy and (3) drop out times. These factors were weighted according to clinical importance determined by a survey. During the restricted conditions of steady state and forced motion test on healthy volunteers Masimo/Ivy’s pulse oximeter performed best with a 2.6 fold improvement over the conventional technology. Clear improvements were also found for Agilent’s M3 (1.6 fold) and Nellcor’s N-3000(1.6 fold). In contrast, the clinically oriented NMC study yieldedthe best performance improvement – as measured in NPI numbers– for Agilent’s M3 rev. B (1.6 fold) and due to more frequent SpO2 errors only 1.5 for Masimo and 1.3 for N-3000.A large difference was found for the dropout rate: the lowest was achieved by Masimo (3.0% of total time), the largest by NellcorN-3000 (24.1% of total time), a factor which was rated high by clinicians. Very pronounced improvements (between 2.3 and 3.4 fold) on all of the newer devices were found for the pulse rate. The NMC turned out to be a very useful tool for generating a standard signal setfor algorithm development and benchmarking purposes that eliminates repetitive clinical testing in early stages. The applicability of its results needs confirmation by clinical live studies.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Similar content being viewed by others

REFERENCES

  1. Bohnhorst B, Poets CF. Major reduction on alarm frequency with a new pulse oximeter. Intens Care Med 1998; 24(2): 277–8.

    Article  CAS  Google Scholar 

  2. Zabar R, Zmora E. Nurses response to false alarms from monitoring systems in the NICU. Pediatr Res 1997; 41: 174A.

    Google Scholar 

  3. Visram AR, Jones RDM, Irwin MG, Bacon-Shone J. Use of two pulse oximeters to investigate a method of movement artifact rejection using photo-plethysmographic signals. Brit J Anaesth 1994; 72: 288–392

    Google Scholar 

  4. Lawless ST. Crying wolf: False alarms in a pediatric intensive care unit. Crit Care Med 1994; 22: 981–985.

    PubMed  CAS  Google Scholar 

  5. Nellcor. Die nächste Generation von Nellcor Pulsoximetern. Ein technischer Überblick zu Nellcor Symphony N-3000, Nellcor GmbH, Idstein, Germany, 1994.

    Google Scholar 

  6. Masimo. Masimo Signal Extraction Technology (SET). Technical Bulletin 1, Masimo Co., Irvine, CA, 1998.

    Google Scholar 

  7. Kästle S, Blank H, Blank M. Method and apparatus for determining the concentration of a component. European Patent Office EP 0870466 B1, 1998.

  8. Barker SJ, Shah NK. The effects of motion on the performance of pulse oximeters in volunteers. Anesthesiology 1997; 86: 101–108.

    Article  PubMed  CAS  Google Scholar 

  9. Bohnhorst B, Peter CS, Poets CF. Pulse oximeters' reliability in detecting hypoxemia and bradycardia: Comparison between Nellcor N-200, N-3000 (Oxismart) and Masimo SET. Presented at The Society of Neonatal Intensive Care Conference, Frankfurt, Germany, September 1998. Online at http://www.masimo.com/PDF/b14.pdf.

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Kästle, S.W., Konecny, E. Determining the Artifact Sensitivity of Recent Pulse Oximeters during Laboratory Benchmarking. J Clin Monit Comput 16, 509–522 (2000). https://doi.org/10.1023/A:1011420228523

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1023/A:1011420228523

Navigation