Elsevier

Radiotherapy and Oncology

Volume 81, Issue 2, November 2006, Pages 168-175
Radiotherapy and Oncology

EPID and IMRT
Dosimetric pre-treatment verification of IMRT using an EPID; clinical experience

https://doi.org/10.1016/j.radonc.2006.09.008Get rights and content

Abstract

Background and purpose

In our clinic a QA program for IMRT verification, fully based on dosimetric measurements with electronic portal imaging devices (EPID), has been running for over 3 years. The program includes a pre-treatment dosimetric check of all IMRT fields. During a complete treatment simulation at the linac, a portal dose image (PDI) is acquired with the EPID for each patient field and compared with a predicted PDI. In this paper, the results of this pre-treatment procedure are analysed, and intercepted errors are reported. An automated image analysis procedure is proposed to limit the number of fields that need human intervention in PDI comparison.

Materials and methods

Most of our analyses are performed using the γ index with 3% local dose difference and 3 mm distance to agreement as reference values. Scalar parameters are derived from the γ values to summarize the agreement between measured and predicted 2D PDIs. Areas with all pixels having γ values larger than one are evaluated, making decisions based on clinically relevant criteria more straightforward.

Results

In 270 patients, the pre-treatment checks revealed four clinically relevant errors. Calculation of statistics for a group of 75 patients showed that the patient-averaged mean γ value inside the field was 0.43 ± 0.13 (1 SD) and only 6.1 ± 6.8% of pixels had a γ value larger than one. With the proposed automated image analysis scheme, visual inspection of images can be avoided in 2/3 of the cases.

Conclusion

EPIDs may be used for high accuracy and high resolution routine verification of IMRT fields to intercept clinically relevant dosimetric errors prior to the start of treatment. For the majority of fields, PDI comparison can fully rely on an automated procedure, avoiding excessive workload.

Section snippets

IMRT pre-treatment verification: current practice

In our institute, the fluoroscopic Theraview NT (TNT) EPID (Cablon Medical-Theraview Technology, Leusden, The Netherlands) is used for portal imaging. This system is equipped with a low-noise cooled CCD camera. Its stable response [0.4% (1 SD)], the simultaneous integration of signal in 1024 × 1024 pixels, and a dead time between the acquisition of frames of only 0.2 ms make this EPID well suited for dosimetric measurements in IMRT fields produced with dynamic multileaf collimation [9].

For the

Intercepted errors

In 270 patients, severe errors in the fluence delivery were detected four times: once a wrong plan was accidentally sent to the accelerator, and three times one of the leaves was malfunctioning. For the wrong treatment plan all five treatment fields showed a γ image with most values larger than one, clearly showing that there was a serious problem. Fig. 1 shows dose profiles for one of the leaf pairs, for the wrong (1A) and the correctly transferred (1B) treatment plan, respectively. Similar

Discussion

In the literature, there is no consensus of the criteria to be used for calculation of the γ index when comparing two dose distributions. For our clinical practice, the 3% local/3 mm criterion has been selected. However, in other studies, the dose difference criterion is taken with respect to the maximum global dose instead [2], [23]. Low and Dempsey suggest using 5% global/2–3 mm criteria for IMRT verification in clinical practice, based on their experience that for a 3% dose difference the

Conclusion

Fluoroscopic EPIDs may be used for high accuracy and high resolution routine verification of IMRT fields to intercept clinically relevant dosimetric errors prior to the start of treatment. Such errors have been found for 4 patients in a total of 270. An automated image analysis procedure has been proposed to limit the involved workload. Only the images that do not pass the tests are selected for further review by a physicist. With the proposed test criteria only 1/3 of the images require this

Acknowledgements

We thank Martijn Hol and Paul Jacobs for development of software tools and Adri van den Biggelaar and Henk de Swarte for carrying out the pre-treatment measurements. This work was financially supported by the Dutch Cancer Society (Grant DDHK 2004-3107).

References (28)

  • E.M. Franken et al.

    A novel approach to accurate dosimetry using CCD-camera based EPIDs

    Med Phys

    (2006)
  • B.J.M. Heijmen et al.

    Portal dose measurement in radiotherapy using an electronic portal imaging device

    Phys Med Biol

    (1995)
  • T. LoSasso et al.

    Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy

    Med Phys

    (1998)
  • D.A. Low et al.

    A technique for the quantitative evaluation of dose distributions

    Med Phys

    (1998)
  • Cited by (68)

    • Improving intensity-modulated radiation therapy quality assurance by adopting statistical process control

      2022, Cancer/Radiotherapie
      Citation Excerpt :

      Caution should be exercised when using EPIDs as they are not ideal absolute dosimeters. Calibration and commissioning should be performed before use, following vendor recommendations [1,6–10]. The accuracy and stability of the EPID should be regularly monitored.

    View all citing articles on Scopus
    View full text