Single or triple gradients?

https://doi.org/10.1016/j.jmr.2008.04.029Get rights and content

Abstract

Pulsed Field Gradients (PFGs) have become ubiquitous tools not only for Magnetic Resonance Imaging (MRI), but also for NMR experiments designed to study translational diffusion, for spatial encoding in ultra-fast spectroscopy, for the selection of desirable coherence transfer pathways, for the suppression of solvent signals, and for the elimination of zero-quantum coherences. Some of these experiments can only be carried out if three orthogonal gradients are available, while others can also be implemented using a single gradient, albeit at some expense of performance. This paper discusses some of the advantages of triple- with respect to single-gradient probes. By way of examples we discuss (i) the measurement of small diffusion coefficients making use of the long spin-lattice relaxation times of nuclei with low gyromagnetic ratios γ such as nitrogen-15, and (ii) the elimination of zero-quantum coherences in Exchange or Nuclear Overhauser Spectroscopy (EXSY or NOESY) experiments, as well as in methods relying on long-lived (singlet) states to study very slow exchange or diffusion processes.

Introduction

Among the key ingredients that have made Magnetic Resonance Imaging (MRI) possible are carefully engineered coils designed to induce accurate Pulsed Field Gradients (PFGs). Three orthogonal gradients Gx = dBz/dx, Gy =  dBz/dy, and Gz = dBz/dz can be generated by suitable Maxwell or Helmholtz coils designed to avoid generating non-linear (quadratic and higher) derivatives of the main Bz field [1]. Such coils have evolved from shim-coils that were originally designed to improve the homogeneity of static fields for high-resolution spectroscopy, and from PFG coils that were designed to measure diffusion coefficients [2], [3]. Ideally, the orthogonality of modern PFG coils precludes mutual interference, and so-called active shields, i.e., coils with similar geometry but larger dimensions and opposite polarity, make it possible to minimize the magnetic coupling with the environment, particularly with the innermost shield of the main magnet bore. Once optimized for MRI, sophisticated designs for PFG coils were soon ‘re-imported’ into high-resolution spectroscopy. They have now become ubiquitous tools for the accurate measurement of translational diffusion coefficients, for spatial encoding in ultra-fast NMR, for elimination of unwanted zero-quantum coherences, for suppression of strong solvent signals [4], and for selection of coherence transfer pathways, either simply by dephasing or ‘purging’ unwanted coherences, or by selecting suitable coherence transfer echoes [5] to retain desirable pathways. In some instances, the use of gradients makes extensive phase cycles superfluous, e.g., in the case of heteronuclear correlations, a pair of gradients with strengths inversely proportional to the gyromagnetic ratios of the two involved nuclei ensures clean selection of the desired pathway. In homonuclear systems, the selection of coherences is in some cases impossible to perform using phase cycles alone. For instance, the distinction between longitudinal two-spin order and zero-quantum coherences requires pulsed field gradients.

Although some of these methods can only be applied if three orthogonal gradients are available, in other cases it has been possible to design a poor man’s version using a single gradient. The limitation to a single gradient may lead to a sacrifice in performance, e.g., to imperfect selection of coherence transfer pathways. Poor suppression of water signals makes some experiments very difficult to implement on single-gradient probes, even when cumulative dephasing of the water magnetization is achieved by an appropriate choice of the signs of consecutive gradients [6]. Ultra-fast three- or four-dimensional NMR [7], [8] requires two or three orthogonal gradients. The simultaneous activation of two or three orthogonal gradients makes it possible to generate tilted gradients, in a manner that may allow one to control deleterious effects of demagnetizing dipolar fields [9], [10], [11]. On the other hand, triple gradients may entail a minor loss of sensitivity due to unavoidable technical compromises in the design of very crowded probes. Similar issues may arise for solid-state NMR with Magic Angle Spinning (MAS): while a single PFG with a tilted gradient Gz=dBz/dz parallel to the spinning axis z is reasonably straightforward to implement, perpendicular, rotor-synchronized, time-dependent gradients designed to appear static in the frame of the rotor are more challenging [12].

This paper discusses advantages of triple- over single-gradient probes. By way of example, we shall discuss two applications of broad interest:

  • (i)

    The encoding and decoding of the phases of coherences by gradients

    This is relevant both for spin-echo diffusion experiments and for heteronuclear correlation experiments. We shall discuss the measurement of small diffusion coefficients via longitudinal magnetization of low-γ nuclei [13] that have long spin-lattice relaxation times T1, such as nitrogen-15;

  • (ii)

    The elimination of zero-quantum coherences while retaining longitudinal magnetization terms

    Both EXSY and NOESY experiments [14], [15], [16] are discussed, as well as the elimination of zero-quantum coherences while retaining longitudinal two-spin order terms 2IzSz for broadband excitation of long-lived (singlet) spin states [17], [18], [19], [20], [21], [22], [23], [24], which can be used to study slow dynamics [25] and slow diffusion [26], [27].

Section snippets

Coherence encoding in diffusion experiments

For the history of Diffusion Ordered Spectroscopy (DOSY) and earlier measurements of diffusion coefficients by magnetic resonance, the reader is referred to reviews by C.S. Johnson [28] and W.S. Price [29], [30]. Fig. 1 shows a recent variant of the stimulated echo designed for slowly diffusing systems [13]. Typically, for complexes of 15N-enriched outer-membrane proteins with detergents having molecular masses of about 45 kDa, T1(15N)  1 s, whereas T1(1H)   100 ms. This makes it attractive to store

Suppression of zero-quantum coherences

In many NMR experiments, the excitation of specific coherences over a broad range of shifts and couplings may be necessary. In the case of homonuclear spin systems, it is difficult to discriminate between zero-quantum coherences and longitudinal two-spin terms using radio-frequency pulses alone. Pulsed field gradients (PFGs) can make this distinction, and the best option is to use three orthogonal gradients. The amplitudes and lengths of the gradients must be set with care in order to achieve

Elimination of zero-quantum coherences

In a homonuclear pair of J-coupled spins I and S, a spin-echo sequence [(π/2)x  τ1   (π)x  τ1] with a duration 2τ1 = 1/(2J) leads to antiphase terms 2IxSz + 2IzSx (at time point b in Fig. 3). A (π/4)y pulse transforms these antiphase terms into a superposition of longitudinal two-spin order, zero-quantum, and double-quantum coherences at point c of the sequence in Fig. 3:2IxSz+2IzSx2IxSx-2IzSz=ZQx+DQx-2IzSzwhere the zero- and double-quantum coherences are defined as usual [44]:ZQx=1/2(2IxSx+2IySy)DQx=

Experimental

All experiments were carried out on 500 MHz (B0 = 11.75 T) spectrometers in Fällanden and Lausanne, both equipped with prototypes of cryoprobes with three orthogonal gradients. The test sample was a monosaccharide with a five-membered furanose ring where all hydrogen atoms - except those in positions H5 and H5- were replaced by deuterium atoms so that the lifetime of the singlet state was TS  26  s, while the longitudinal relaxation time, T1, was 0.7 s, as described in more detail elsewhere [25].

Results and discussion

An imperfect cancellation of ZQx terms after point c of the sequence in Fig. 3 results in variations of the singlet state populations at the beginning of the RF irradiation interval. This is due to destructive interference of the longitudinal two-spin order and zero-quantum coherences at point d.

We have calibrated the filter described in Materials and Methods using a sequence that excites both 2IzSz and ZQx. The latter is partly eliminated by the filter, while the former is transformed into a

Conclusions

It has been shown that probes with triple gradients are of great advantage for the accurate determination of slow diffusion constants with the X-STE method. A reliable calibration of Thrippleton–Keeler (TK) filters, designed to eliminate zero-quantum coherences, has been presented. By calibrating these filters against their intended target, i.e., against zero-quantum coherences, it has been shown that their performance is improved when at least two orthogonal gradients are available. These

Acknowledgments

This work was supported by the Swiss Commission for Technology and Innovation (CTI), the Swiss National Science Foundation (FNRS), the Ecole Polytechnique Fédérale de Lausanne (EPFL), and the French CNRS.

References (47)

  • S. Macura et al.

    Separation and suppression of coherent transfer effects in two-dimensional NOE and chemical-exchange spectroscopy

    J. Magn. Reson.

    (1982)
  • M. Rance et al.

    A systematic-approach to the suppression of J-cross peaks in 2D exchange and 2D NOE spectroscopy

    J. Magn. Reson.

    (1985)
  • K. Gopalakrishnan et al.

    Lifetimes of singlet-states under coherent off-resonance irradiation in NMR spectroscopy

    J. Magn. Reson.

    (2006)
  • G. Bodenhausen et al.

    Selection of coherence-transfer pathways in NMR pulse experiments

    J. Magn. Reson.

    (1984)
  • M.H. Levitt et al.

    Cogwheel phase cycling

    J. Magn. Reson.

    (2002)
  • R. Fu et al.

    Broad-band decoupling in NMR with frequency-modulated chirp pulses

    Chem. Phys. Lett.

    (1995)
  • A.J. Shaka et al.

    An improved sequence for broad-band decoupling—WALTZ-16

    J. Magn. Reson.

    (1983)
  • P.T. Callaghan

    Principles of nuclear magnetic resonance microscopy

    (1991)
  • E.O. Stejskal et al.

    Spin diffusion measurements—spin echoes in presence of a time-dependent field gradient

    J. Chem. Phys.

    (1965)
  • J.E. Tanner

    Use of stimulated echo in NMR-diffusion studies

    J. Chem. Phys.

    (1970)
  • M. Piotto et al.

    Gradient-tailored excitation for single-quantum NMR-spectroscopy of aqueous solutions

    J. Biomol. NMR

    (1992)
  • D.R. Muhandiram et al.

    Measurement of H-2 T1 and T1rho relaxation-times in uniformly C-13-labeled and fractionally H-2-labeled proteins in solution

    J. Am. Chem. Soc.

    (1995)
  • L. Frydman et al.

    The acquisition of multidimensional NMR spectra within a single scan

    Proc. Natl. Acad. Sci.

    (2002)
  • Cited by (24)

    • Ultrafast 2D NMR for the analysis of complex mixtures

      2022, Progress in Nuclear Magnetic Resonance Spectroscopy
      Citation Excerpt :

      In the Oneshot method, described by Pelta et al. [87], unbalanced bipolar gradient pulses are used for CTP selection, so that a single scan per increment is sufficient. This can also be achieved with triple-axis gradient probes, using orthogonal axes for diffusion encoding and CTP selection [88]. Joint non-uniform sampling of the gradient and indirect time dimensions can also be used to speed up the acquisition of 3D DOSY experiments [89].

    • Symmetry versus entropy: Long-lived states and coherences

      2021, Progress in Nuclear Magnetic Resonance Spectroscopy
    • Spatial encoding and spatial selection methods in high-resolution NMR spectroscopy

      2018, Progress in Nuclear Magnetic Resonance Spectroscopy
      Citation Excerpt :

      The filter has also been included in other pulse sequences that require efficient ZQC suppression [145–148], e.g., to help in obtaining in-phase signals in multidimensional correlation experiments. In some cases, the filter is used to retain multi-spin longitudinal order [149,150]. Since the underlying spatiotemporal dynamics are largely orthogonal to the rest of the sequence, these developments will not be discussed here.

    • Improving accuracy in DOSY and diffusion measurements using triaxial field gradients

      2016, Journal of Magnetic Resonance
      Citation Excerpt :

      Experiments that measure the effects of transverse displacements of spins, using transverse gradient pulses, are therefore far less vulnerable to convection than conventional experiments, which use z gradient pulses and are sensitive to vertical displacements. The relative immunity to convection when using transverse field gradient pulses to encode/decode diffusion was noted by Bodenhausen and co-workers [29–31], although the pulse sequence described in that paper only used transverse gradients to purge unwanted magnetization. ( Note however that because the active volume of an NMR sample is taller than it is wide, field gradient pulses along the z-axis are more efficient at selecting coherence transfer pathways and/or purging unwanted magnetization than transverse pulses.)

    • Long-lived coherences for line-narrowing in high-field NMR

      2011, Progress in Nuclear Magnetic Resonance Spectroscopy
      Citation Excerpt :

      It follows from Eq. (20) that ε is attenuated by uncorrelated external contributions to the relaxation rates. In analogy to relaxation rates RLLS of long-lived states, which are immune to fluctuations associated with CSA and DD tensors that are symmetrical under permutation of spins I and S [20,21], the lifetimes of long-lived coherences will be affected to a lesser extent by spin-permutation symmetric relaxation mechanisms. Fig. 5 shows signals of Gly-76 and Gly-75 in ubiquitin excited and observed with the simultaneous excitation method.

    View all citing articles on Scopus
    View full text