Effects of the soft tissue artefact on the hip joint kinematics during unrestricted activities of daily living
Introduction
Skin marker-based motion capture (MC) is one of the most commonly used methods for human motion analysis. Yet, its accuracy is limited by the soft tissue artefact (STA), which is caused by the relative movement between skin markers and bone due to inertial effects, skin sliding and muscle contraction (Camomilla et al., 2017a, Leardini et al., 2005). STA was identified to be the largest source of error of skin markers-based motion capture (Benoit et al., 2006, Peterset al., 2010) and it is hard to resolve in a non-invasive manner.
Quantification of the effects of STA on the kinematics of the hip joint is important for reliable interpretation of results from clinical gait analysis (Benedetti et al., 2017, Ku et al., 2015, Wrenet al., 2011), for detection of femoro-acetabular impingement (Rylander et al., 2013), for wear analysis of the prosthetic hip (Hua et al., 2016, Mellonet al., 2013), and for validation of the estimates from musculoskeletal models. In studies on simulated gait propagation of STA had the highest impact among all sources of uncertainty on the hip joint moment outputs, and produced variations of 37.2 ± 20 N on the predicted muscle forces (Myers et al., 2015). Lamberto et al. (2017) reported that STA causes variations of up to 1.8 times the body weight for hip contact forces and of 30% to 50% for some muscle forces. Stagni et al. (2000) estimated that misplacements of the hip joint center (HJC) by 30 mm produce errors as high as 22% and 15% in hip flexion-extension and adduction-abduction moments, respectively. These studies concluded that propagation of STA to the estimates of current biomechanical models should be compensated for, especially when accurate subject-specific kinetics and bone strains are required.
For compensation of STA, a thorough in-vivo characterization of STA and of its effects during motion activities are required. Recent methods measured STA in-vivo by comparing the skin marker positions with the 3D poses of the underlying joint segment retrieved from X-ray video fluoroscopic images. While several studies have used this technique to quantify the STA for the knee joint during different activities (Akbarshahi et al., 2010, Barré et al., 2013, Garling et al., 2007, Kuo et al., 2011, Da Li et al., 2017, Stagniet al., 2005, Tsaiet al., 2011), fewer have been performed for the hip joint. Recently, Fiorentino et al. (2017) have used dual-plane fluoroscopy and CT to quantify STA and its effects on the angles and range of motion of the native hip joint during level and inclined treadmill gait, and during hip adduction and hip rotation motions. They reported large STA, especially for the markers attached to the femur, and underlined its dependency on anatomical direction, subject and activity. STA caused an underestimation of the overall range of motion, with values up to 21.8 deg for the internal-external rotation during hip rotation. The effects of STA on the HJC location were also analyzed in-vivo using video-fluoroscopy. Functional methods for determination of the HJC location were reported to carry errors of 11 ± 3 mm, with STA the primary contributor (Fiorentino, Kutschke, et al., 2016). During dynamic activities, STA caused about 2.2 mm of spurious HJC motion, contributing to overall HJC errors of 16.6 ± 8.4 mm (Fiorentino, Atkins, et al., 2016).
To date, characterization of STA is still a complex challenge due to its dependency on subject (Benoit et al., 2006), on motion activity (Holden et al., 1997, Leardini et al., 2005) and on skin-marker configuration (Schwartz et al., 2004), its non-linearity with respect to the pose parameters of the underlying bone (Tsaiet al., 2009, Barré et al., 2013), and the scarcity of reference in-vivo estimations (Cereatti et al., 2017, Peterset al., 2010). Consequently, propagation of STA on the kinematic estimates has not been thoroughly characterized. The objective of this study was to extend the existing knowledge of the effects of STA on the kinematics of the hip joint, by quantifying them for a sample total hip arthroplasty (THA) population, for a broader range of activities of daily living (ADLs). A moving fluoroscopic system was used to measure the hip joint during unconstrained motions such as overground gait, stair descent and chair rise, which were not previously assessed due to the use of stationary X-ray imaging equipment. Errors due to STA were quantified across the whole motion, including the relative contribution of each joint segment. The overall error on the HJC location was also analyzed with respect to both its components: a static component, arising from STA during the functional calibration task and dependent on the functional algorithm, and a dynamic component, cause by inertial STA occurring during motion.
Section snippets
Subjects and measurements
15 participants with successful unilateral primary THA (9 men, 6 women, average age 65 ± 7.4, mass 77.4 kg ± 10.6, height 174.7 cm ± 9.1, BMI 25.3 kg/m2 ± 2.2, average follow-up time 31 months ± 10) provided their written informed consent to the study, which was approved by the local cantonal ethics committee (BASEC-No. 2016-00438).
Each subject was measured with moving video-fluoroscopy (VF) and optical motion capture at a reference standing upright position, and while performing four ADLs:
Results
Soft tissue artefact had a statistically significant effect across subjects on the ROM and the angles of the hip joint, on the positions of the pelvis and the thigh, and on the location of the HJC, depending on the activity and on the specific phase of the motion cycle.
The ROM retrieved by MC was generally reduced (Fig. 1, Supplementary Table 1). ROM for FE was underestimated by on average 4.1, 6.5, 8.0 and 6.9 deg for gait, stair decent, chair rise and putting on socks respectively,
Discussion
Effects of STA on the kinematics of the hip joint, of the pelvis and of the thigh and on the location of the HJC were quantified during each phase of the motion cycle of four ADLs.
Due to STA, the MC-measured ROM of the hip joint angles was reduced for all activities and all anatomical axes, except for the AA and the IE axes during gait and for the IE axis during stair descent, confirming the findings from a previous study (Fiorentino et al., 2017). Mean ROM errors over all trials were largest
Declaration of Competing Interest
We declare that we have no financial or personal relationships with other people or organizations that could inappropriately influence (bias) our work.
Acknowledgments
This work was supported by the European Union Seventh Framework Programme FP7-NMP-2012 LARGE-6, “LifeLongJoints” (grant number 310477). The digital models of the hip implants were kindly provided by Smith&Nephew. We would like to thank Dmitriy Alexeev, Enrico De Pieri and Renate List, for their technical support.
References (43)
- et al.
Non-invasive assessment of soft-tissue artifact and its effect on knee joint kinematics during functional activity
J. Biomech.
(2010) SIAMOC position paper on gait analysis in clinical practice: General requirements, methods and appropriateness. Results of an Italian Consensus Conference
Gait and Posture
(2017)- et al.
Effect of skin movement artifact on knee kinematics during gait and cutting motions measured in vivo
Gait and Posture
(2006) - et al.
A soft tissue artefact model driven by proximal and distal joint kinematics
J. Biomech.
(2014) - et al.
A model of the soft tissue artefact rigid component
J. Biomech.
(2015) - et al.
Human movement analysis: The soft tissue artefact issue
J. Biomech.
(2017) - et al.
Soft tissue displacement over pelvic anatomical landmarks during 3-D hip movements
J. Biomech.
(2017) Standardization proposal of soft tissue artefact description for data sharing in human motion measurements
J. Biomech.
(2017)A procedure for determining rigid body transformation parameters
J. Biomech.
(1995)- et al.
Determining 3D kinematics of the hip using video fluoroscopy: guidelines for balancing radiation dose and registration accuracy
J. Arthroplasty
(2017)