Single probe for real-time simultaneous monitoring of neurochemistry and direct-current electrocorticography

https://doi.org/10.1016/j.bios.2015.09.021Get rights and content

Highlights

  • We developed a probe for simultaneous neurochemicals and DC-ECoG monitoring.

  • Two on-probe reference electrodes were included to improve sensor accuracy.

  • Transient neurochemical and DC potential changes can be detected during the passage of SD.

Abstract

We report a novel single neural probe for real-time simultaneous monitoring of multiple neurochemicals and direct-current electrocorticography (DC-ECoG). A major advance of this probe is the inclusion of two iridium oxide reference electrodes to improve sensor accuracy. The ECoG reference electrode is identical to the ECoG recording electrodes to significantly improve DC stability, while the reference for electrochemical sensors has 10-fold lower polarization rate to minimize the small current-induced drift in the reference electrode potential. In vitro, the single probe selectively measured oxygen (r2=0.985±0.01, concentration range=0-60 mmHg, limit of detection=0.4±0.07 mmHg) and glucose (r2=0.989±0.009, concentration range=0–4 mM, limit of detection=31±8 µM) in a linear fashion. The performance of the single probe was assessed in an in vivo needle prick model to mimic sequelae of traumatic brain injury. It successfully monitored the theoretically expected transient brain oxygen, glucose, and DC potential changes during the passage of spreading depolarization (SD) waves. We envision that the developed probe can be used to decipher the cause-effect relationships between multiple variables of brain pathophysiology with the high temporal and spatial resolutions that it provides.

Introduction

The brain is the most complex organ in the human body made up of a network of billions of neurons. The communications between the neurons are inherently electrical and chemical in nature through approximately 100 billion neurons with 100 trillion synapses. Hence, there is a considerable need in neuroscience research for in situ, simultaneous acquisition of the electrical and chemical signals. After acute injury to the brain, neurons at the damage epicenter depolarize, and this depression of activity spreads outward in waves through the cortex (Ayata and Lauritzen, 2015, Dreier, 2011, Leao, 1944). Decades of research show that cortical spreading depolarization (SD) can cause progressive lesion growth after ischemic, hemorrhagic, or traumatic brain injury by increasing tissue metabolic burden and diminishing perfusion (Chase, 2014; Hartings et al., 2009, Hartings et al., 2011; Hinzman et al., 2014; Lauritzen et al., 2011; Takano et al., 2007). Detection of such processes through neuromonitoring offers the prospect of targeting patient therapies to detected intracranial pathology. However, not all pathologic brain activity is harmful, and accurate multimodal monitoring of SD and other events is needed to determine which ones require treatment. Furthermore, simultaneous electrical and chemical measures in the same brain tissue location can provide insight to cause–effect relationships among the multiple variables of interest and inform the best therapeutic approach.

Previous attempts at simultaneous electrocorticography (ECoG) and neurochemical signal recordings were based on the use of multiple, separate electrodes, which limits spatial and temporal resolution since separate sensors are not located within the same microenvironment (Dash et al., 2009, Naylor et al., 2011, Odijk et al., 2015, Rogers et al., 2013). Consequently, the development of a single probe capable of simultaneously detecting electrical and chemical signals is of great significance. Implantable micromachined microelectrode arrays represent a versatile and powerful tool to achieve this in vivo (Fekete, 2015, Tseng and Monbouquette, 2012, Wise et al., 2008). The techniques allow for the development of smaller electrodes that provide improved spatial and temporal resolution as well as multiple sensing electrodes with precisely defined geometries. Micro-electrochemical sensors allow the continuous, precise and fast in situ measurement of metabolic parameters directly within the tissue (Cordeiro et al., 2015; Hinzman et al., 2015a, Hinzman et al., 2015b; Kealy et al., 2013; Rocchitta et al., 2013; Wei et al., 2014; Weltin et al., 2014).

Recently, several neural probes have been developed for simultaneous electrophysiological and neurochemical signal recordings (Johnson et al., 2008, Morales-Villagrán et al., 2008, Song et al., 2012, Wei et al., 2015). By placing both ECoG recording electrodes and electrochemical sensors in the area of interest, near real-time monitoring of multiple parameters is possible. However, it is challenging to measure multiple variables simultaneously, especially for direct-current electrocorticography (DC-ECoG), since crosstalk between sensors introduces noise and instability. For example, DC-ECoG recording electrodes should have a very low polarization rate and stable baseline for detecting the DC shift of SD. Currently, microelectrodes for SD recording are limited to glass micropipettes with Ag/AgCl wire due to their non-polarizable property (Hinzman et al., 2015a, Hinzman et al., 2015b, Odijk et al., 2015, Rogers et al., 2013). No study to date has examined the accuracy and stability of micromachined probes for simultaneous DC-ECoG and electrochemical signal recordings.

The reference electrode is a key component for both ECoG electrodes and electrochemical sensors (Shinwari et al., 2010). However, these electrodes are not usually integrated into a single sensing probe. In previous work, an Ag/AgCl wire, screw, or platinum wire was used as a remote reference for the ECoG electrode and was placed in a comparable but inactive area such as the contralateral hemisphere, musculature, or subdermal space (Lewitus et al., 2011, Muthuswamy et al., 2005). Most electrochemical sensors used for in vivo studies also rely on a separate reference electrode, usually a Ag/AgCl wire (Thomas et al., 2009; Wassum and Tolosa, 2012). Ag/AgCl, however, has been shown to cause widespread necrosis in biocompatibility studies (Stensaas and Stensaas, 1978) and the implantation of multiple electrodes can cause more damage to the brain. A single probe combining all sensing and biocompatible reference electrodes can overcome this important issue, which is particularly important in the clinical setting. Such an integrated probe for in vivo application not only reduces tissue damage, but may also reduce noise arising from external sources and within the brain, providing improved signal-to-noise ratios (SNR) and detection limits (Li et al., 2009a, Peteu et al., 1996, Tolosa et al., 2013).

Previous studies have shown that iridium oxide (IrOx) electrodes can be a suitable replacement for Ag/AgCl wire reference electrodes for electrochemical sensors and can be used as a quasi-reference electrode for in vivo neuroscience applications (Tolosa et al., 2013, Yang et al., 2004). IrOx exhibits excellent biocompatibility and mechanical stability; it is stable over a range of potentials and in the presence of the ions and other components of brain extracellular fluid. Although the IrOx potential shows strong pH dependence (Ges et al., 2005), the small dynamic range of normal brain pH (7.15–7.4) makes this issue negligible (less than 30 mV potential drift) in most cases. However, IrOx has drawbacks as well. Compared to Ag/AgCl, IrOx microelectrodes have much higher impedance which not only slows the response of the potentiostats, but also increases the susceptibility of the system to environmental noise. In addition, unlike the nearly ideal non-polarization liquid–solid interface which Ag/AgCl forms in solutions, IrOx is a polarizable material, such that the passage of small currents can alter its potential. So far, the stability of the IrOx electrode potential was only studied in phosphate-buffered saline (PBS) solution; its stability as a reference electrode under continuous electrochemical sensor operation has not been reported.

In this paper, we present a new micromachined single probe for real-time simultaneous monitoring of multiple neurometabolic and DC-ECoG signals. ECoG microelectrodes are designed to be placed in gray matter to detect cortical SD, whereas electrochemical sensors are designed to be placed in white matter to measure the focal brain oxygen and glucose concentrations simultaneously. In place of separate Ag/AgCl reference electrodes, two biocompatible IrOx reference electrodes are incorporated on-probe to achieve improved DC stability and accuracy. The ECoG reference electrode is identical to the ECoG recording electrodes so that both electrodes detect the same background noise and hence increase the DC stability and SNR. The reference electrode for the electrochemical sensors has 10-fold lower polarization rate than the ECoG reference to minimize the open circuit potential drift due to small current-induced polarization. To assess the performance of the developed single probe, it was tested in vivo using a needle prick model that induces SD, thereby mimicking the electrophysiologic and neurochemical changes of developing brain injury. We demonstrate that this single probe can accurately monitor brain glucose, oxygen tension, and DC-ECoG simultaneously.

Section snippets

Reagents and apparatus

All chemicals were analytical reagent grade or higher purity and were used as received. Chitosan (MW~105, 75–85% deacetylation), glucose oxidase (GOD) from Aspergillus niger (EC 1.1.3.4), glutaraldehyde (25% in water), β-d-glucose, potassium chloride (K2PtCl6), sulfuric acid (99.999%), o-phenylenediamine (oPD), iridium tetrachloride hydrate, oxalic acid dehydrate (99%), hydrogen peroxide (30% solution in water), anhydrous potassium carbonate, and phosphate buffered saline (pH 7.4) were

Potential crosstalk and solution

All microelectrodes on the neural probe are in direct contact with the conductive media. There is a chance that the low-frequency current generated from the electrochemical biosensors affects the ECoG recording and the electrochemical reference electrodes if these sensors are in close proximity. Fig. 2A illustrates the crosstalk paths between the electrochemical biosensors and the ECoG electrodes via their equivalent circuit model. IN represents neuronal activity current and IL stands for the

Conclusion

In the present study, a single neural probe for real-time simultaneous monitoring of multiple neurometabolic and DC-ECoG signals has been successfully developed. The use of two on-probe iridium oxide reference electrodes greatly improved the DC stability of baseline ECoG signals and also increased the accuracy of electrochemical sensors by minimizing the current-induced drift, thereby improving the capacity for long-term monitoring. As proof-of-concept, the performance of the probe was tested

Acknowledgments

This research was supported by the U.S. Department of Defense CDMRP PH/TBI Research Program under Project no. PT090526P4 (W81XWH-10-1-0978).

References (44)

  • C.A. Cordeiro et al.

    Biosens. Bioelectron.

    (2015)
  • Z. Fekete

    Sens. Actuator B—Chem.

    (2015)
  • I.A. Ges et al.

    Biosens. Bioelectron.

    (2005)
  • J.A. Hartings et al.

    Co-operative study on brain injury depolarisations

    Lancet Neurol.

    (2011)
  • J.M. Hinzman et al.

    Exp. Neurol.

    (2015)
  • J.M. Hinzman et al.

    Biosens. Bioelectron.

    (2015)
  • M.D. Johnson et al.

    J. Neurosci. Methods

    (2008)
  • J. Kealy et al.

    J. Neurosci. Methods

    (2013)
  • D. Lewitus et al.

    Acta Biomater.

    (2011)
  • C. Li et al.

    Biosens. Bioelectron.

    (2007)
  • C. Li et al.

    Biosens. Bioelectron.

    (2009)
  • A. Morales-Villagrán et al.

    J. Neurosci. Methods

    (2008)
  • J. Muthuswamy et al.

    J. Neurosci. Methods

    (2005)
  • E. Naylor et al.

    J. Electroanal. Chem.

    (2011)
  • M. Odijk et al.

    Sens. Actuator B—Chem.

    (2015)
  • S.F. Peteu et al.

    Biosens. Bioelectron.

    (1996)
  • Y. Song et al.

    Biosens. Bioelectron.

    (2012)
  • T.T.-C. Tseng et al.

    J. Electroanal. Chem.

    (2012)
  • V.M. Tolosa et al.

    Biosens. Bioelectron.

    (2013)
  • W. Wei et al.

    Biosens. Bioelectron.

    (2014)
  • A. Weltin et al.

    Biosens. Bioelectron.

    (2014)
  • C. Ayata et al.

    Physiol. Rev.

    (2015)
  • Cited by (22)

    • Accurate and stable chronic in vivo voltammetry enabled by a replaceable subcutaneous reference electrode

      2022, iScience
      Citation Excerpt :

      Iridium oxide (IrOx) is one promising alternative to Ag/AgCl due to its biocompatibility, and has been used in sensing, stimulation, and recording studies in vivo (Li et al., 2009, 2016; Shi et al., 2021; Zheng et al., 2021). Unlike Ag/AgCl, IrOx is polarizable, although much less so than other electrode materials like platinum (Li et al., 2016). However, IrOx is highly sensitive to changes in pH (Seaton and Heien, 2021).

    • A novel approach with glass needle enclosed movable probe for in vivo real-time detection of glucose in cisternal cerebrospinal fluid

      2020, Journal of Electroanalytical Chemistry
      Citation Excerpt :

      However, it has never been achieved in real-time in vivo detection of dynamic changes in concentration of CSF glucose, which is critical for the studies of the related brain disorders. The electrochemical sensing method is becoming more and more popular over the last decade for glucose monitoring, due to its high sensitivity and selectivity, excellent temporal resolution, high portability and low cost compared with other means [12–15]. However, so far, most of electrochemical sensors are not applicable for implanting in cisterna magna or lumbar, since dura mater (which is much stronger than brain tissue) can impair the sensors during the puncture [16].

    • Concurrent measurements of neurochemical and electrophysiological activity with microelectrode arrays: New perspectives for constant potential amperometry

      2018, Current Opinion in Electrochemistry
      Citation Excerpt :

      Similarly, Wei and collaborators presented simultaneous measurements of LFP/spike activity and striatal extracellular glutamate [40] or glucose [41] using CPA with a silicon-based Pt MEA modified with nanomaterials. The quest to obtain a more integrated view of neurometabolism has also unveiled the need for multi-analyte recordings coupled to monitoring of electrical activity, recently achieved using a polyimide-based neuroprobe comprising 2 Pt sites for D-glucose and oxygen and 2 IrO sites for concurrent DC electrocorticography (DC-ECoG) [42]. The work of Millar and colleagues introduced multimodal measurements of E-chem/E-phys using a single carbon fiber microelectrode to measure monoamine efflux and single-unit neuronal activity in brain slices by alternating between recording modes [43].

    • Simultaneous in vivo ascorbate and electrophysiological recordings in rat brain following ischemia/reperfusion

      2016, Journal of Electroanalytical Chemistry
      Citation Excerpt :

      Due to the chemical and electrical nature of neuron-to-neuron communications, information on the simultaneous changes of neurochemical and electrical signals involved in the physiological and pathological processes remains very essential to understanding the neurotransmission mechanism in brain functions [16–26]. This is because the change in the levels of chemical species from the extracellular space surrounding neurons largely facilitates understanding of neurochemically associated physiological and pathological processes on a molecular basis, while the electrophysiological changes can provide immediate information on the condition of the tissue and offer additional prognostication of neurological outcome in varied neurological disorders [19,21,24,27–31]. As one kind of the most important neurochemicals, ascorbate exists as a water-soluble small molecule that has been demonstrated to play critical roles in many physiological and pathological processes [32–34].

    • The Role of Multimodal Invasive Monitoring in Acute Traumatic Brain Injury

      2016, Neurosurgery Clinics of North America
      Citation Excerpt :

      The relationship between CSD and seizures remains an intriguing question that it is hoped will be explored in future studies, particularly as intracranial EEG becomes more readily available.51 The development of single probes able to provide electrophysiologic, oxygen, and neurochemistry data could facilitate exploration of the pathophysiologic interactions and offer significant insight into the mechanisms of SBI.52 The true value of monitoring may be in understanding mechanisms of injury and in tracking pathophysiologic interactions at the tissue level.

    View all citing articles on Scopus
    View full text