Layer-by-layer assembly of chemical reduced graphene and carbon nanotubes for sensitive electrochemical immunoassay

https://doi.org/10.1016/j.bios.2012.02.007Get rights and content

Abstract

In this work, uniform and stable multi-walled carbon nanotubes (MWCT) and chemically reduced graphene (GR) composite electrode interface was fabricated by using layer-by-layer assembly method. The performances of these GR–MWCT assembled electrode interfaces were studied by electrochemical impedance spectroscopy (EIS) and cyclic voltammetry (CV). It was demonstrated that the assembled composite film significantly improved the interfacial electron transfer rate compared with that of GR or MWCT modified electrode. Based on the GR–MWCT assembled interface, a sandwich-type electrochemical immunosensor was constructed using human IgG as a model target. In this assay, human IgG was fixed as the target antigen, the HRP-conjugated IgG as the probing antibody and hydroquinone as the electron mediator. The detection limit of the immunosensor was 0.2 ng mL−1 (signal-to-noise ratio of 3). A good linear relationship between the current signals and the concentrations of Human IgG was achieved from 1 ng mL−1 to 500 ng mL−1. Moreover, this electrochemical immunosensor exhibited excellent selectivity, stability and reproducibility, and can be used to accurately detect IgG concentration in human serum samples. The results suggest that the electrochemical immunosensor based on GR–MWCT assembled composite will be promising in the point-of-care diagnostics application of clinical screening of multiple diseases.

Highlights

▸ We constructed a uniform and highly porous GR–MWCT composite electrode interface. ▸ The GR–MWCT modified GC electrodes exhibited significant improvement on electron transfer rate, mass transfer. ▸ We fabricated a sandwich-type electrochemical immunosensor based on the GR–MWCT electrode interface. ▸ The immunosensor shows excellent selectivity, stability and reproducibility with the detection limit of 0.2 ng mL−1.

Introduction

Sensitive and accurate detection of disease-related proteins is a fundamental requirement in the field of modern biomedicine, clinical diagnostics and therapeutic analysis, which will offer opportunities for the understanding of disease related biological processes (Kitano, 2002, Weston and Hood, 2004, Xiao et al., 2005). Immunoassay, which is one of the most commonly used methods based on the highly specific interaction between antigen and antibody, is conducted on the clinical serum sample to measure disease-associated biomarkers. Until now, several immunoassay methods including chemiluminescence immunoassay (Fu et al., 2006, Voller et al., 1978), enzyme-linked immunosorbent assay (ELISA) (Yates et al., 1999), radioimmunoassay (Goldsmith, 1975), fluorescence immunoassay (Cesaro-Tadic et al., 2004, Matsuya et al., 2003), mass spectrometric immunoassay (Aebersold and Mann, 2003, Hu et al., 2007), electrophoretic immunoassay (Schmalzing and Nashabeh, 1997) and immune polymerase chain reaction (PCR) assay (Saito et al., 1999) have been developed. Recently, electrochemical immunosensor has attracted much attention especially in the point-of-care applications due to the high sensitivity, simplicity, ease of miniaturization, and low-cost of both the sensors and the instrumentation, and several electrochemical immunosensors have been reported (Ala-Kleme et al., 2006, Bourigua et al., 2010, Chen et al., 2010b, Miao and Bard, 2004, Pumera et al., 2007, Wan et al., 2010). Although great progresses have been made, it is still a challenge to develop novel detection technologies and enhance the detection sensitivity due to the increasing demand for inexpensive, operationally simple, early and ultrasensitive profiling of early clinical disease diagnostic including cancer biomarkers.

For an electrochemical immunosensor, its performance is critically dependent on the properties of electrode interface. The structure, conductivity, surface area and biocompatibility of the electrode interface play a central role on the electron transfer, mass transport, biomolecule activities on the interface electrochemical signal transduction (Chen et al., 2006, Jacobs et al., 2010, Lai et al., 2011, Lu et al., 2006, Warsinke, 2009). On the other hand, stable immobilization and retaining the activity of biomolecules on the electrode interface are of fundamental importance for engineering reliable biosensors (Chen and Li, 2006). Recently, nanomaterials, the building blocks of nanotechnology, have opened new horizons for electrochemical interface as well as high performance devices due to their good biocompatibility, enhanced electrocatalytic activity, large surface area, excellent conductivity, and so on. The organization of nanomaterials would open a new pathway for engineering intelligent multifunctional interfaces for the construction of high performance biosensors (Jie et al., 2008, Jie et al., 2010, Pinijsuwan et al., 2008, Wu et al., 2010).

Carbon nanomaterials have attracted considerable attention in electrochemical biosensors because of their extraordinary physical properties and remarkable conductivities. Carbon nanotubes are formed by rolling graphite sheets, and they have quasi-one dimensional (1D) structures. Due to their great chemical stability, large aspect ratio, excellent electrical conductivity, and extremely high mechanical strength and stiffness, carbon nanotubes are widely used in the development of high-performance electrochemical sensors (Ahammad et al., 2009, Bareket et al., 2010, Jacobs et al., 2010, Nie et al., 2009). Recently, graphene, another new carbon materials, attracted much attention since its fascinating two-dimensional (2D) structure, unusual electrochemical properties, large accessible surface area as well as good biocompatibility (Chen et al., 2010a, Du et al., 2011, Geim and Novoselov, 2007, Jacobs et al., 2010, Liu et al., 2008, Lu et al., 2009, Wang et al., 2010, Yang et al., 2010). Graphene presents excellent electron transfer ability for enzymes and excellent catalytic behavior towards small molecules such as dopamine, H2O2, O2, NADH and TNT (Kong et al., 2011, Lu et al., 2011, Tang et al., 2009, Wang et al., 2009, Zhou et al., 2009). Chemically reduced graphene (GR), which is fabricated through chemical reduction of colloidal dispersion of graphene oxide, possesses the outstanding electronic features. In addition, a small quantity of functional groups such as hydroxyl, carboxyl and epoxy also remain on its edges and surface, which make GR negatively charged and is ideal to be functionalized to build multifunctional nanostructure hybrid materials in the applications of high efficiency photoelectronic energy conversion module and sensing devices (Bai et al., 2011, Du et al., 2011, Wan et al., 2011).

In this work, we fabricated a composite electrode interface based on the quasi 1D structure of multi-walled carbon nanotubes (MWCT) and 2D plane structure of graphene by a layer-by-layer assembly method through the electrostatic adsorption between positively charged poly(diallyldimethylammonium chloride) (PDDA) and negatively charged MWCT and GR. The performance of the composite electrode interface was studied by electrochemical impedance spectroscopy and cyclic voltammetry, and improved electron transfer ability was observed compared with those only MWCTs or GR used. Moreover, a sandwich-type electrochemical immunosensor was constructed based on the assembled composite interface, using human IgG as a model target antigen, human IgG antibody as the capture probe fixed on the composite interface, the HRP-conjugated human IgG antibody as the probes and hydroquinone as the electron mediator. The results were also correlated with standard ELISA method. The as-prepared electrochemical immunosensor platform provides a strategy to fabricate high performances electrode interface and shows great promise in the point-of-care diagnostics application of clinical screening of disease-associated cancer biomarkers.

Section snippets

Regents and materials

Graphite powder (99.99995%, 325 mesh), 1-ethyl-3-(3-dimethylaminopropyl) carbodiimide hydrochloride (EDC), N-hydroxylsuccinimide (NHS), bovine serum albumin (BSA) and H2O2 (30%) were purchased from Alfa Aesar. Poly(diallyldimethylammonium chloride) (PDDA, MW ∼100,000 was purchased from Sigma–Aldrich and used as received without further purification. The goat anti-human IgG (Ab1), goat anti-human IgM (anti-IgM), human IgG (hIgG) (Ag), HRP conjugated goat anti-human IgG (Ab2-HRP), Adenosine

Results and discussion

Layer-by-layer technique offers a simple and easy way to build uniform nanostructure materials with controllable composition, structure and functions. Scheme 1 represents the assembly process of the GR–MWCT modified GC electrode. To verify the assembly of MWCT and GR, the modified GC electrode was characterized by SEM. Fig. S2A shows the SEM pictures of the GC electrode after the assembly of GR. A large amount of rolled sheets of graphene are clearly presented, indicating that GR are well

Conclusion

A uniform and highly porous GR–MWCT composite network electrode interface was fabricated by a layer-by-layer assembly method. This three-dimensional assembled composite film exhibited significant improvement on the electron transfer rate, mass transfer and specific surface area at the electrode interface compared with that of only GR or MWCTs used. Based on the assembled GR–GR–MWCT electrode interface, a sandwich-type electrochemical immunosensor is constructed using human IgG as a model

Acknowledgments

This work was financially supported by National Basic Research Program of China (no. 2011CB935700), the National Natural Science Foundation of China (no. 21128005, no. 21005046), European Commission Specific Program GlycoHIT (FP7-HEALTH-2010, no: 260600), Specialized Research Fund for the Doctoral Program of Higher Education (no. 20100002120042) and Tsinghua University Initiative Scientific Research Program.

References (46)

  • L. Bareket et al.

    Bioelectrochemistry

    (2010)
  • S. Bourigua et al.

    Biosens. Bioelectron.

    (2010)
  • D. Chen et al.

    Surf. Sci. Rep.

    (2006)
  • Z. Fu et al.

    J. Immunol. Methods

    (2006)
  • S.J. Goldsmith

    Semin. Nucl. Med.

    (1975)
  • C.B. Jacobs et al.

    Anal. Chim. Acta

    (2010)
  • G.F. Jie et al.

    Electrochem. Commun.

    (2010)
  • F.Y. Kong et al.

    Talanta

    (2011)
  • L.-M. Lu et al.

    Biosens. Bioelectron.

    (2011)
  • X. Lu et al.

    Biomaterials

    (2006)
  • M. Pumera et al.

    Sens. Actuators B: Chem.

    (2007)
  • Y. Wang et al.

    Electrochem. Commun.

    (2009)
  • Z. Xiao et al.

    Mol. Cell. Endocrinol.

    (2005)
  • R. Aebersold et al.

    Nature

    (2003)
  • A.J.S. Ahammad et al.

    Sensors

    (2009)
  • T. Ala-Kleme et al.

    Anal. Chem.

    (2006)
  • H. Bai et al.

    Adv. Mater.

    (2011)
  • S. Cesaro-Tadic et al.

    Lab Chip

    (2004)
  • D. Chen et al.

    Chem. Soc. Rev.

    (2010)
  • D. Chen et al.

    J. Phys. Chem. C

    (2006)
  • L.L. Chen et al.

    Chin. J. Anal. Chem.

    (2010)
  • D. Du et al.

    Anal. Chem.

    (2011)
  • A.K. Geim et al.

    Nat. Mater.

    (2007)
  • Cited by (0)

    View full text