Deep Flexion-Oriented Bisurface-Type Knee Joint and Its Tibial Rotation That Attributes Its High Performance of Flexion
Section snippets
Development of BS Knee
We considered that the idea of conventional knee prostheses with just replicating the anatomy of the natural knee joint would not be good enough when a deep flexional function is pursued with a knee prosthesis of new design. After extensive consideration, we determined to add the prosthesis a structure that specifically functions for flexion. We divided the function of the femorotibial articulation into 2 parts and made these 2 parts to have 2 separate functions, weight bearing and extra
Case Material and Method
As typical cases of deep flexion, 30 knees of 23 patients who could sit on legs in oriental sitting position (seiza) were selected of 946 joints operated between July 1997 and August 2008 for the prosthetic kinematics study of this knee. The knees were examined by lateral view of x-ray. Because there is no practical method to directly measure the prosthetic kinematics in vivo, indirect techniques have been applied. There have been various reported techniques concerning with a 3-dimensional
Results
The average flexion angles of the 946 joints operated between July 1997 and August 2008 were 132.2° before operation and 138.5° 1 month after operation. Joints with more than 130° of maximum flexion accounts for 67% before operation, and the number considerably increased to 86% 1 month after operation.
The good flexion quality of the BS knee was analyzed from the point of kinematics. Although informal sitting that shifts the feet just off to the side (Fig. 3) or cross-legged sitting (Fig. 4) was
Discussion
It has been widely accepted that deep flexion of the knee could be achieved with internal rotation of the tibia. It was affirmed that the BS knee could achieve the flexion function that was the initial objective of the development. However, the 3-dimensional presentation of kinematics, simultaneous movement of flexion, and rotation were difficult. Through the development of computer technology, this proposition found a solution using a pattern matching method. There is a concern about early
References (12)
- et al.
Effect of segmentation errors on 3D to 2D registration of implant models in x-ray images
J Biomechanics
(2005) - et al.
A mechanical comparison of 2 posterior-stabilizing designs—Insall/Burstein 2 knee and bisurface knee
The Journal of Arthroplasty
(2002) - et al.
A new artificial knee joint that has an articulation designed for complete knee flexion
Japanese J Clin Biomech
(1990) - et al.
The total condylar knee prosthesis. A report of two hundred and twenty cases
J Bone Joint Surg
(1979) - et al.
The posterior stabilized condylar prosthesis: a modification of the total condylar design
J Bone Joint Surg
(1982) - et al.
Improved range of flexion after total knee arthroplasty. the total condylar knee versus the KU knee
Bulletin Hospital for Joint Diseases.
(1997)
Cited by (19)
In vitro biomechanical evaluation of tri-condylar total knee arthroplasty with posterior release for restoration of full extension
2017, Journal of Orthopaedic TranslationCitation Excerpt :However, the PS TKA raises post-related issues such as post-wear, post-fractures, or patellar clunk syndrome [7–9]. Tri-condylar TKA is another solution for improving knee function in deep flexion [10–16]. The primary feature of this design is the introduction of a third nonanatomic spherical condyle in the posterior groove and a socket surface on the tibial component, thus providing a contact between the femoral and tibial components in deep flexion.
Noise Generation With Good Range of Motion but Without Femorotibial Instability Has Small Effect on Patient Satisfaction After Total Knee Arthroplasty
2017, Journal of ArthroplastyCitation Excerpt :The patella was resurfaced. Previous studies reported that patients with this prosthesis had excellent postoperative ROM, including in the kneeling position, as well as a high level of patient satisfaction and long-term durability at a minimum of 10 years' follow-up [12,14-16]. In the comparative study, we also examined 3 knees (3%) using a Bi-Surface knee with unresurfaced patella and 18 knees (17%) with other TKA designs, to compare factors related to component materials, for instance the metal-on-polyethylene interface or the degree of articular constraint (18 women, 3 men; mean age, 71 years [range, 59-82 years]; osteoarthritis, 15 knees; rheumatoid arthritis, 6 knees).
In vivo kinematic effects of ball and socket third condyle as a post-cam mechanism in tri-condylar knee implants
2015, KneeCitation Excerpt :This third condyle is expected to replace the function of a post-cam mechanism and also to act as a load-bearing interface in flexion. In previous biomechanical studies this design has been shown to provide sufficient stability like the commonly used PS designs [18,19,22], but it is still unknown whether the third condyle actually works as a posterior stabilizer in vivo. It has been hypothesized that the ball and socket joint will work as a posterior stabilizer like other PS post-cam designs and that the ball and socket third condyle will not prevent normal axial rotation.
In vivo femoro-tibial kinematic analysis of a tri-condylar total knee prosthesis
2014, Clinical BiomechanicsCan post-cam function be replaced by addition of a third condyle in PS TKA?
2014, Journal of Arthroplasty3D in vivo femoro-tibial kinematics of tri-condylar total knee arthroplasty during kneeling activities
2014, KneeCitation Excerpt :The Bi-Surface Knee System developed by Kyocera Medical (Osaka, Japan) has a unique tri-condylar design and is classified as a posterior-cruciate-substituting TKA (Fig. 1). The unique design of this particular implant includes a ceramic femoral component to reduce wear, as well as a ball-and-socket joint in the midposterior portion of the femoral component and polyethylene insert [32–36]. Unlike other PS TKAs, whilst this ball-and-socket joint has no protruding post, it is however, expected to work as a PS cam mechanism and share a load together with the medial and lateral condyles in flexion [37,38].
No benefits or funds were received in support of the study.