Clinical Studies
Use of radial artery applanation tonometry and a generalized transfer function to determine aortic pressure augmentation in subjects with treated hypertension

https://doi.org/10.1016/S0735-1097(98)00411-2Get rights and content
Under an Elsevier user license
open archive

Abstract

Objectives. The purposes of this study were to investigate the use of radial artery applanation tonometry and a generalized transfer function for the assessment of central aortic pressure augmentation in subjects taking commonly used antihypertensive agents (angiotensin-converting enzyme inhibitors, beta-adrenergic blockers, Ca2+antagonists, diuretic therapy).

Background. Applanation tonometry of the radial artery with a generalized transfer function has been proposed as a means of assessing central aortic blood pressure. Recently, a commercial apparatus based on this technique has become available; we therefore examined the effect of a generalized transfer function on derived central aortic pressure compared with measured brachial blood pressures and also investigated the potential of this technique to assess the influence of differing drug therapy.

Methods. Two hundred and sixty-two hypertensive patients on stable medication were studied using the PWV Medical Blood Pressure Analysis System (version 2, DAT-1).

Results. In univariate analysis, augmentation index showed association with age, sex, height and heart rate. In multivariate analysis, diastolic blood pressure and age (positively), height and heart rate (negatively) and sex were significantly associated. After adjustment for these variables, pressure augmentation was not associated with any antihypertensive treatment investigated. Linear relationships were demonstrated between brachial blood pressures and corresponding central pressures derived by transfer function methods.

Conclusions. Our findings suggest that if adjustment for central-peripheral pressure difference is necessary, simple linear relationships may be sufficient. Age, heart rate and height but not the class of antihypertensive medication affected the degree of pressure augmentation observed using this technique.

Abbreviations

ACE
angiotensin-converting enzyme
AI
augmentation index
AP
augmentation pressure
DBP
diastolic blood pressure
PWV
pulse wave velocity
SBP
systolic blood pressure
TF
transfer function

Cited by (0)

This study was supported by an equipment grant from Servier Laboratory (Australia) Pty Ltd, Servier House, 13-15 Cato St, Hawthorn, Australia.