The effects of ionizing radiation on eight cardiac pacemakers and the influence of electromagnetic interference from two linear accelerators

https://doi.org/10.1016/S0167-8140(85)80011-6Get rights and content

Summary

Eight cardiac pacemakers were irradiated in a cobalt-60 beam. Two out of six demand-type pacemakers showed an alarming decrease in pulse repetition frequency when irradiated to dose levels that are used in radiotherapy (<100 Gy). Two modern programmable pacemakers showed a failure at a dose of 97 and 147 Gy, respectively. The dose levels at which these failures occurred were low enough to recommend that cardiac pacemakers should always be kept outside the radiation beam.

The signals induced by electromagnetic interference (EMI) from two linear accelerators were measured using a simulation model of a pacemaker. In the laboratory, 22 modern-type pacemakers were tested with these signals to determine the sensitivity for the electromagnetic fields in the treatment rooms. It was observed that an inhibition of one pacemaker pulse was to be expected on one of the two linear accelerators when switching the machine on and off. No permanent effects were found. These findings resulted in the recommendation in our department not to use this treatment machine for radiation therapy of pacemaker-bearing patients.