Reliability of surface electromyography of the masseteric and anterior temporal areas

https://doi.org/10.1016/0003-9969(90)90098-UGet rights and content

Abstract

This was studied in 37 patients suffering from myofascial pain-dysfunction. Bipolar surface electrodes were used to record tonic, resting EMG activity for 2 consecutive trials. This protocol was repeated at a second recording session 2 weeks later. A custom-made, plastic template was constructed for each subject during the first visit in order to relocate the electrodes accurately at the second recording session. Correlation coefficients were calculated for each muscle area. Within-session (same day) r values ranged from 0.7620 to 0.8884 for the masseteric area and from 0.8686 to 0.9109 for the anterior temporal area. Across-session (different day) r values ranged from 0.5645 to 0.6503 for the masseteric area and 0.3309 to 0.4844 for the anterior temporal area. The lower correlation between different day recordings could reflect a methodological shortcoming in relocating the electrodes, particularly on the anterior temporal area. The greater variability recorded from the anterior temporal area could also reflect the dynamic role of the temporalis muscle in maintaining the postural rest position of the mandible.

References (32)

  • L. Angelone et al.

    An approach to quantitative electromyography of the masseter muscle

    J. dent. Res.

    (1960)
  • Autogenic Systems Incorporated

    Instruction Manual for the Autogen 1500c

    (1977)
  • J.C. Barbenel

    Analysis of forces at the temporo-mandibular joint during function

    Dent. Practit.

    (1969)
  • J.V. Basmajian et al.
  • B.H. Burdette et al.

    The effects of treatment on masticatory muscle activity and mandibular posture in myofascial pain-dysfunction patients

    J. dent. Res.

    (1988)
  • B.H. Burdette et al.

    Electromyographic activity of masticatory muscles of MPD patients during opening

    J. dent. Res.

    (1989)
  • Cited by (42)

    • Coordination of the non-paretic leg during hemiparetic gait: Expected and novel compensatory patterns

      2012, Clinical Biomechanics
      Citation Excerpt :

      Nonetheless, we present the P-values to allow the reader to assess the strength of evidence for the changes observed in the coordination of the non-paretic leg. Another limitation is that absolute magnitude of EMG (in μV) recorded via surface electrodes is influenced by several intrinsic (thickness of sub-cutaneous fat, distribution of motor units etc.) and extrinsic factors (location and area of the electrode and distance between them), and thus is limited in its ability to provide strong evidence for changes in the magnitude of muscle activity between subjects (Burdette and Gale, 1990; Farina, 2006). Nevertheless, we believe that despite the limitations, integrated absolute EMG provides evidence for increased activity that is confirmatory of the increased propulsion generation typically measured in the non-paretic leg (Bowden et al., 2006; Kim and Eng, 2004; Olney et al., 1991) and it is also useful when interpreted in the context of the increased duration of activity observed for many muscles.

    • Influence of different upper cervical positions on electromyography activity of the masticatory muscles

      2012, Journal of Manipulative and Physiological Therapeutics
      Citation Excerpt :

      The diameter of each of the 2 circular conductive areas was 1 cm, with an interelectrode distance of 2 cm. Despite the influence of factors including changes in head and body posture, skin resistance, temperature and humidity, as well as muscle fatigue, emotional factors, topographical location of the electrodes over the muscle area, and the removing and replacing of the electrodes on the reliability and reproducibility of surface EMG,28,29 several studies have reported good to excellent reliability and reproducibility for this technique.30-32 For the measurement of mandibular excursions, subjects sat in a chair at approximately a 90° angle to the examiner with the jaw muscles in a passive state.

    • The clinical usefulness of surface electromyography in the diagnosis and treatment of temporomandibular disorders

      2006, Journal of the American Dental Association
      Citation Excerpt :

      It is possible that the displacement could alter the SEMG recordings significantly, thus resulting in different conclusions about the muscle activity. Burdette and Gale81 found that SEMG recordings were altered significantly with changes to IED, even though they tried to reliably relocate the surface electrodes with a custom-made template. Other researchers found that alteration to the IED created a greater variation in surface recordings from the deeper layers of the muscle fibers (masseter) than from the superficial layers (anterior temporal),82,83 implying that for accurate measurement of different muscles, IED must be individualized depending on the depth of the fibers.

    • Masticatory muscle activity assessment and reliability of a portable electromyographic instrument

      2001, Journal of Prosthetic Dentistry
      Citation Excerpt :

      As expected, these subjects also showed significant differences between left- and right-sided EMG recordings with unilateral, right-sided only chewing activity. The level of reliability of the portable EMG integrator is similar to the values found in other studies with other portable EMG recording equipment in a controlled laboratory environment.7,10,11 After completion of the last phase of data collection in this investigation, several of the portable EMG units began to spontaneously accumulate EMG activity during a period of no muscle activity.

    • Role of Electromyography in Dental Research: A Review

      2023, Journal of Research in Dental and Maxillofacial Sciences
    View all citing articles on Scopus
    View full text