Abstract
Subcutaneous injection by means of prefilled syringes allows patients to self-administrate high-concentration (100 g/L or more) protein-based drugs. Although the shear flow of concentrated globulins or monoclonal antibodies has been intensively studied and related to the injection force proper of SC processes, very small attention has been paid to the extensional behavior of this category of complex fluids. This work focuses on the flow of concentrated bovine serum albumin (BSA) solutions through a microfluidic “syringe-on-chip” contraction device which shares some similarities with the geometry of syringes used in SC self-injection. By comparing the velocity and pressure measurements in complex flow with rheometric shear measurements obtained by means of the “Rheo-chip” device, it is shown that the extensional viscosity plays an important role in the injection process of protinaceous drugs.
Article Highlights
-
A microfluidic “syringe on chip” device mimicking the injection flow of protinaceous drugs has been developed.
-
The velocity field of concentrated BSA solutions through the “syringe on chip” is Newtonian-like.
-
The extensional viscosity of concentrated protein solutions should also be considered when computing injection forces through needles.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
1 Introduction
Macromolecules such as monoclonal antibodies (mAbs) are increasingly being used for treating a large variety of diseases, including immunodeficiency, leukemia and arthritis [17], while globular proteins such as recombinant serum albumin are often employed as drug stabilizers [28, 59]. Subcutaneous (SC) injection by means of prefilled syringes coupled with autoinjector devices constitutes a viable alternative to intravenous delivery, as it allows the patients to self-administrate the protinaceous drug at home. Perceived pain and back pressure typically limit the highest deliverable volume to 1.5 mL [37, 58, 62]; therefore, biopharmaceuticals need to be concentrated above 100 g/L in order for the drug to be effective. Such large values of protein concentrations typically translate into large solution viscosity (up to 1 Pa s [2, 63]) as well as into shear thinning behavior [8, 11, 18, 68] at the shear rates typical of SC injection (\(10^{4} \le {\dot{\gamma }} \le 10^{5}\) s\(^{-1}\) for 27-G needle syringes, see [1]).
Several previous works have produced a calculation of the injection force of concentrated mAbs through syringes by assuming either Newtonian or shear-thinning viscosity [1, 7, 48]. It must however be observed that the flow of protein solutions through needles can be considered as a complex flow, because it contains not only a shear component, but also a well-defined extensional contribution which is localized in the contraction part between the syringe barrel and the needle. Thus, in order to develop a complete understanding of the origin of the forces arising when a biopharmaceutical drug is delivered by means of SC injection, one must characterize not only the shear, but also the extensional viscosity of concentrated protein solutions. While there is a large amount of works which focus on bulk shear viscosity [29, 36, 53, 68], high-frequency viscoelasticity [42, 43, 65] or interfacial shear rheometry [26, 57] of dense protein solutions, very little attention has so far been paid to the elongational behavior of this category of complex fluids.
The extensional viscosity of complex fluids is directly related to the amount of deformation underwent by macromolecules under flow. Following de Gennes’ work [10], a polymer in extensional flow undergoes a transition from coiled to uncoiled state if it is deformed with a strain rate whose magnitude is comparable to the inverse of the longest polymer relaxation time, and for a long enough time such that a sufficient amount of strain can be accumulated. For example, \(\lambda\)-bacteriophage DNA has a relaxation time of approximately 0.1 s [45], and hence it reaches the uncoiled state if extensional rates of 10 s\(^{-1}\) or above are applied. While there is a general consensus in the literature on the flow conditions that need to be achieved in order to stretch DNA and linear polymer chains, the data related to proteins are more controversial. Jaspe and Hagen [27] estimated that horse cytochrome unravels under purely extensional flow at strain rates of \(10^{7}\) s\(^{-1}\) or above. Churchich et al. [9] suggest that also the relaxation time of a much larger globulin such as bovine serum albumin (BSA) is on the order of \(10^{-7}\) s. This is in contrast to the experimental work of Bekard et al. [5], which suggest that bovine serum albumin (BSA) unfolds in Couette flows at shear rates as low as \(10^{2}\) s\(^{-1}\). By means of a microfluidic device, Schneider [54] demonstrated that the von Willebrand factor, a protein commonly found in human blood, can unravel at shear rates of about \(10^{3}\) s\(^{-1}\). Computer simulations [60, 61] showed that the unraveling under shear flow of ubiquitin and integrin gives rise to many more intermediates than in the case of homopolymers. The dynamics of unraveling also depend on whether the protein is anchored or not, and on which specific protein site is anchored. More recently, Dobson et al. [13] showed that BSA and monoclonal antibodies can also aggregate when they undergo multi-pass extensional flow at strain rates \({\dot{\varepsilon }} \cong 10^{4}\) s\(^{-1}\). The authors suggest that the aggregation of protein molecules at large extensional rates is caused by flow-driven protein unfolding, which is in turn related to strong hydrodynamic forces.
The advent of microfluidics has made it possible to study flow regimes which could not be previously investigated. Due to characteristic sizes of just a few microns, microfabricated devices allow for fluids to be tested under very large (\({\dot{\gamma }} \ge 10^{4}\) s\(^{-1}\)) rates of deformation, while Reynolds numbers (hence inertial effects) are almost negligible. When macromolecular solutions are subjected to large deformation rates in microscopic flows through abrupt [30, 35, 49, 50] and hyperbolic contractions [32] or “cross-slots” [3, 24, 33], nonlinear flow effects such as asymmetries in the flow field and vortex development are often witnessed. While a considerable amount of effort has been spent in studying microscopic flows of macromolecular fluids such as polymers, DNA and other biopolymers [19,20,21, 23], or wormlike micellar solutions [22, 44, 69], very little attention has so far been paid to the behavior of concentrated protein solutions at high deformation rate, complex flow conditions which mimic practical processes. It is then natural to ask if the flow-induced protein unraveling and/or aggregation phenomena discussed above can lead to nonlinear flow effects similar to what has been previously observed for other types of macromolecules.
In the present work, the commonly made assumption that the injection force of protinaceous drugs is made only by a purely shear contribution is critically assessed. This is done by means of a microfluidic platform which mimics the flow behavior of concentrated protein solutions when flowing through syringes. Three BSA solutions in a pH \(=7.0\) buffer condition and at \(c_{2}=100, 200\) and 300 g/L were used here. Such concentrations are above the physiological value (\(7\times 10^{-4}\) M, equivalent to \(c_{2} \approx 50\) g/L) typically found in human blood [14]. Nevertheless, BSA solutions at this pH and over this range of \(c_{2}\) are expected to have a similar shear and extensional viscosity with respect to that of mAbs-based biopharmaceuticals [53]. The microscopic geometries employed here are straight channel and a “syringe-on-chip” device consisting in several sudden planar contractions with same contraction ratio as those found in 26-G, 27-G and 30-G syringe needles. From the first geometry, the steady shear viscosity \(\eta \left( {\dot{\gamma }}\right)\) corresponding to a range of deformation rates (\(10\, s^{-1} \le {\dot{\gamma }} \le 10^{4}\,\mathrm{s}^{-1}\)) comparable with that of the syringe injection process is extracted. From the “syringe-on-chip” geometry, the velocity and local strain rate fields within a similar range of nominal deformation rates are measured by means of particle-image velocimetry (PIV) [39, 41], so that it is verified whether the flow of the BSA solutions retains a Newtonian character. Finally, the extra pressure drops from the “syringe-on-chip” device are critically compared with the results obtained from the steady shear flow measurements, so that the relative importance of shear and extensional viscosity in the overall injection force can be assessed.
2 Materials
BSA (Cat. no. A7906) was obtained from Sigma-Aldrich. The protein was initially dissolved at a concentration \(c_{2}=300\) g/L in a pre-filtered buffer made by NaCl (145 mM), sodium octanoate (8 mM) and Tween 80 (0.05 g/L) at a pH \(=7.0\) and let to dissolve for 72 h at 5 \(^{\circ }\)C. The initial BSA sample was filtered with a 0.22-microns filter (Anotop 10 filters, GE Healthcare, Piscataway, CA) to remove irreversible aggregates, and then fresh buffer was used for all subsequent dilutions up to the desired \(c_{2}=100\), 200 or 300 g/L. The concentration of the protein samples was confirmed by a UV-1600PC spectrophotometer (VWR) operating at a wavelength of 280 nm. The small-angle X-ray scattering (SAXS) measurements performed by Sønderby et al. [59] on recombinant serum albumin, a globulin very similar to BSA, demonstrated that protein–protein interactions are repulsive under the chosen buffer conditions.
3 Experimental methodology
3.1 Rheo-chip platform
All microfluidic experiments presented in this work were run by means of the “Rheo-chip” platform [67]. It consists of a series of microchannels made in polymethilmetaacrilate (PMMA) by Epigem Ltd (Redcar, TS, UK). Each microchip features a delivery module where inlets, outlets and pressure taps are obtained. Pressure measurements were taken by means of Honeywell strain gauge sensors 26PCGFM6G (range 0–250 psi). The transient pressure signals were collected at 50 Hz by a cDAQ data acquisition system (National Instruments) and a LabVIEW software. The flow through microfluidic devices was controlled by means of a Nexus 6000 high force syringe pump (Chemyx Inc, Stafford, Tx) equipped with a 1-mL glass syringe (SGE Analytical, ID \(=4.5\) mm). Note that, differently from other microfluidics-based rheometric techniques available elsewhere [46], the Rheo-chip technology employs replaceable pressure sensors, so that only the microfluidic chips need to be replaced if microchannels are clogged. This makes the rheometric characterisation of precious materials such as concentrated protein solutions much more cost-effective. The schematic diagrams of the “Rheo-chip” coupled with the pressure sensors are given in Fig. 1.
3.2 Shear viscosity measurements
The steady shear viscosities of the model protein solutions were measured by means of Rheo-chips featuring a straight channel with width \(w_{c}=800\) \(\upmu\)m, depth \(h_{\text {nominal}}=50\) \(\upmu\)m and length \(L=30\) mm. Two pressure taps were located along the microchannel at a distance of \(L=30\) mm from each other, far away from the inlet and outlet, to measure the overall pressure drop \(\varDelta P_{s}\). The pressure signals were collected by imposing an initial flow rate Q\(_{\text {max}}\)=10 mL/h and then waiting for the signal to reach a steady-state value. The flow rate was then reduced to 8 mL/h, and the procedure was repeated until a steady-state pressure drop corresponding to a flow rate \(Q_{\text {min}}=0.25\) mL/h was measured [46]. For all samples studied in this paper, the nominal shear rate at walls is defined as
where \(h_{\text {true}}\) is found by fitting the steady-state pressure drop measurements of DI water with the analytical prediction for Newtonian flows through rectangular ducts given by White [64]. The corresponding shear viscosity is
where the calibration coefficient \(\frac{1}{K}=\overline{\frac{\varDelta P_{\text {buffer}}}{{\dot{\gamma }}}}\) is obtained by averaging the ratios between steady-state pressure drops of buffer and the corresponding shear rates, and \(\eta _{\text {buffer}}\cong\) 1 mPa s is the shear viscosity of buffer as measured by a TA ARES-G2 rheometer equipped with cone and plate measurement tool. The overall measurement time for each sample was 20 mins, approximately, and an amount of sample between 0.6 and 1 mL was used for each measurement. The impact of viscous heating on the rheometric measurements was estimated by the Nahme–Griffith number
where \(\kappa \approx\)0.5918 W/(m K) is the thermal conductivity of buffer and \(\beta =\frac{T}{\eta _\text {buffer}}\left| \frac{d\eta }{dT}\right| _{T=T_{0}} \approx\) 2.6 is the logarithmic derivative of shear viscosity of buffer with temperature T evaluated at T=T\(_{0}=25\) \(^{\circ }\)C. It expresses the ratio between the heat generated by viscous heating of samples and the heat released by thermal diffusion at the surface of the microchannels [46]. Na resulted to be \(\le 10^{-5}\) for all the samples and shear rates studied here, which systematically excludes the influence of thermal effects.
3.3 “Syringe-on-chip” microdevices
The two “syringe-on-chip” microdevices that are used in this work are presented in Fig. 2. They consist of planar sudden contraction devices with an upstream width \(w_{u}=800\) \(\upmu\)m, channel depth \(h=50\) \(\upmu\)m and downstream widths \(w_{d}=47\) and 38 \(\upmu\)m. The corresponding contraction ratios \(\beta\)=17.02 and 21.05 are the same as what is typically found in syringes where the plunger diameter is 4.5 mm and the needle is a 26G or 27G [66]. Hence, from here onwards such microcontractions will be labeled as “26G” and “27G” geometry, respectively. Note that, due to fact that real syringes have axisymmetric instead of planar abrupt geometries, the flow through the syringe-on-chip devices feature lower Hencky strains \(\varepsilon _{H}=\ln \beta\) compared to real syringes case, where \(\varepsilon _{H}=2\ln \beta\). Each flow geometry contains a slot for a pressure sensor as in the figure. The outlet of each microchannel is connected to a fluid reservoir by means of a stainless steel tube with length \(L_{\text {tube}}=100\) mm and inner diameter ID \(=1/16\) inches (Cat. No. 56720-U, Sigma-Aldrich). The total pressure drop \(\varDelta P_{\text {total}}\) measured by the sensor is related to the pressure drop \(\varDelta P\) across the device by \(\varDelta P_{\text {total}}=\varDelta P + \varDelta P_{\text {tube}}\), where \(\varDelta P_{\text {tube}}=\frac{128\mu LQ}{\pi ID^{4}}\) is estimated by means of the Hagen–Poiseuille law. The \(\varDelta P_{\text {tube}}\) contribution resulted to be 5\(\%\) or less of the total pressure drop. The details of the flow geometries are given in Table 1. The flow of the concentrated BSA solutions in complex geometry is characterized in terms of the Reynolds number Re
where \(<v_{d}>\), \(\rho \approx 1000\) g/L and \(\eta _{0}\) are the average speed in the downstream channel, the fluid density and zero-shear viscosity, respectively, and \(D_{h}=\frac{2hw_{d}}{h+w_{d}}\) is the hydraulic diameter of the downstream channel. In addition to Re, the Péclet number
expresses the ratio between the characteristic timescales of flow, \(\tau _{F}=\frac{1}{{\dot{\gamma }}}=\frac{w_{d}}{2<v_{d}>}\) and Brownian diffusion [34]. The diffusion time is \(\tau _{D}=\frac{d_{h}^{2}}{D^{S}_{L}}\), where d\(_{h}=7\) nm is the hydrodynamic diameter of BSA [4], and \(D^{S}_{L}\) is the longtime self-diffusion coefficient of the BSA molecule, which is estimated by the generalized Stokes–Einstein relationship [51] as
with \(k_{B}\) and \(T\approx\) 25 \(^{\circ }\)C being the Boltzmann constant and temperature, respectively.
3.4 Particle-image velocimetry and data analysis method
Velocity fields at the center plane of the “syringe-on-chip” flow channels were measured by a \(\mu\)-PIV system (TSI Instruments Ltd, USA). It consists of a pulsed Nd:YAG laser emitting at a wavelength of 532 nm, a Nikon TE2000-E fluorescent microscope, and an 8 Hz CCD camera with a resolution of 1280 \(\times\) 1024 \(\times\) 12 bits. A Nikon objective lens with magnification \(M=10X\) and numerical aperture \(NA=0.3\) was used. In Fig. 3, a schematic diagram of the “syringe-on-chip” device coupled with the PIV system is shown.
The BSA solutions were seeded with 0.01 wt% epi-fluorescent particles (diameter \(d_{p}=1.0\,\upmu\)m, \(Ex_{\text {max}}/Em_{\text {max}}=542/612\), Duke Scientific Co.). For the above setup, the estimated measurement depth is 26 \(\upmu\)m, equivalent to 52% of the channel depth [38]. All velocity measurements were taken after the pressure drop reached a steady value. Velocity fields were obtained by means of a Nyquist grid engine and a cross-correlation PIV algorithm supplied by TSI. All the \(\mu\)-PIV measurements presented in this work were performed at the middleplane of flow channel, determined by positioning the focus plane at half of the depth between the top and bottom surfaces of the microchannel, using a high-precision y-axis linear motor with step resolution 50 nm. The error in the velocity measurements is about 12–16% below the true value. This systematic error could be reduced where a higher magnification objective lens with a larger NA used. The details of the \(\mu\)-PIV techniques can be found in [39, 52]. All the velocity fields presented in this work were obtained by averaging 50 consecutive instantaneous velocity fields, corresponding to an overall measurement time of 12.5 s. From the measured two-dimensional planar velocity field \({\underline{v}}= {\underline{v}}(x,z)\), the largest eigenvalue \(\lambda _{1}\) of the strain rate tensor \(\underline{{\underline{D}}}=\frac{1}{2}\left( \nabla {\underline{v}}+\nabla {\underline{v}}^{T}\right)\) is then evaluated to construct the local extensional rate field \({\dot{\varepsilon }}\left( x,z\right)\). The uncertainty about the PIV velocity measurements is about 12–16%. More information on the computation of \({\dot{\varepsilon }}_{1}\left( x,z\right)\) can be found in [33].
4 Results
4.1 BSA solutions under steady shear flow
The shear viscosity of the BSA solutions is plotted as a function of the shear rate 200 s\(^{-1} \le {\dot{\gamma }} \le 2\times 10^{4}\) s\(^{-1}\) in Fig. 4a. The viscosity curve of a weakly shear thinning, semi-dilute polyacrylamide (PAAm) solution [30] is also shown for a comparison. The rheograms of BSA appeared to be essentially flat, which is attributed to Pe being below unity throughout the imposed range of \({\dot{\gamma }}\) (see Fig. 4b). A recent work [12] demonstrated that concentrated lysozyme solutions at 5 \(^{\circ }\)C undergo shear thinning over a similar range of \({\dot{\gamma }}\) than what studied here. Such nonlinear effect was attributed to short-range attractive interactions, which lead to the formation of intermediate structures with a size larger than that of the monomer. The absence of shear thinning effect suggests that no intermediate structures are formed in the BSA solutions studied in this work. Unlike the BSA conditions studied here, mAb solutions are known to display shear-thinning effect even at \(\mathrm{Pe}<<1\) [11, 18, 47]. In a recent work, the Rheo-chip technique adopted here has been used for studying the formulation dependence of \(\eta _{0}\) of concentrated solutions of two mAbs [56]. Moreover, in a future work the relationship between the protein–protein interactions and the shear thinning behavior of concentrated solutions of the same two mAbs will be investigated in detail.
4.2 Velocity and local extensional rate fields in “syringe on chip”
In Fig. 5a, the PIV-measured, planar velocity field of the Re \(=0.09\), Pe \(=0.24\) flow of the 300 g/L BSA solution through the 26-G microchannel is shown, to demonstrate its Newtonian-like behavior. To further assess the linearity of the flow field, the \({\dot{\varepsilon }}\left( x,z\right)\) field of the same BSA solution is plotted in Fig. 5b. The maximum measured extensional rate is localized near the contraction throat and resulted to be \(\approx\) 1500 s\(^{-1}\). Such strain rate is one order of magnitude above the threshold value for BSA unfolding (\(10^{2}\) s\(^{-1}\)) reported by Bekard et al. [5], and about 10 times below the strain rate at which extensional flow-driven aggregation of BSA in multi-pass extensional flow occurs [13]. The maximum measured value of the extensional rate is located at the point \(\left( x,z\right) \approx \left( 0,0\right)\). This is a further indication of the absence nonlinear flow effects, like steady viscoelastic flows or lip vortices [50], which are typically revealed by the maximum strain rate point being located at \(x=0\) and \(\left| \frac{z}{w_{d}}\right| >0.5\), that is, outside the lips of the contraction [6, 31, 32]. Therefore, these results indicate that the BSA molecules did not unfold under the existing flow conditions or, even if any unfolding event occurred, it was not released a sufficient amount of elastic energy to trigger the onset of non-Newtonian flow effects.
In Fig. 6, the dimensionless axial velocity profiles \(\frac{v\left( x=0\right) }{<v_{u}>}\), where \(<v_{u}>=\frac{Q}{hw_{u}}\) is the mean flow velocity in the upstream channel, are shown for 0.2 mL/h \(\le Q \le\) 1 mL/h. The velocity profiles overlap within the range of measured Re, which again indicates that the flow field is always Newtonian-like within the range of explored flow rates.
4.3 Pressure measurements through “syringe on chip”
In Fig. 7, the pressure measurements through the 26-G microcontraction are plotted as a function of Re for the buffer as well as for the 100 g/L and the 200 g/L BSA solutions. The experimental data are compared with the corresponding steady-state pressure drop contributions \(\varDelta P_{\text {shear}}\), which are estimated as
where the pressure gradients \(\frac{dP}{dz}\left( \alpha _{i}\right)\) are computed according to the analytical prediction for Newtonian flows through rectangular ducts given to White [64],
where \(\alpha _{1}=\frac{w_{u}}{h}\), \(\alpha _{2}=\frac{w_{d}}{h}\), and \(K=\Sigma _{n=1,3,5\ldots }^{+\infty }\frac{\tanh \left( \frac{n\pi }{2\alpha _{i}}\right) }{n^{5}}\). Such prediction ignores entry and exit effects and assumes that the flow instantaneously reaches the fully developed profile in both the upstream and downstream regions. The total pressure drop across the microchannels is above the purely shear contribution for all the BSA solutions. Such discrepancy is attributed to extensional effects across the sudden contraction, which are in turn related to the extensional viscosity of the model BSA solutions [21, 23, 33]. Due to the fact that the flow geometry used here is a sudden contraction, the difference between the total pressure drop and the purely shear contribution will be larger than in the case of Newtonian flows through hyperbolic contractions [41]. The extra pressure drop \(\varDelta P'=\frac{\varDelta P-\varDelta P_{\text {shear}}}{\varDelta P_{\text {shear}}}\) for the BSA solutions across the 26G and 27G flow geometries resulted to be within the range \(1.5 \le \varDelta P' \le 3\) throughout the explored range of Re (see Fig. 8). The \(\varDelta P'\) of the more viscous 300 g/L solution through the 26-G channel resulted larger than that of the other two solutions. Moreover, the \(\varDelta P'\) of the 100 and 200 g/L solutions through the 27-G channel are above the corresponding values for the 26-G channel because the Hencky strain is larger for the 27G than for the 26G geometry. Note that the simulations carried out by Oliveira et al. [40] for planar contraction-expansion flows of Newtonian fluids show that the Couette correction (analogous to the \(\varDelta P'\) defined here) increases sharply with Re for Re > 1. This is due to the fact that, differently than in the work by Oliveira et al., the flow geometries adopted here do not include an expansion part, where inertia-driven flow instabilities, which also cost nonlinearity in the pressure drop-flow rate curve, are typically observed.
Previous works [1, 15] do not consider the contribution to pressure drop due to extensional forces [21, 23] in modeling the injection force of dense protein solutions through syringes. More specifically, the authors assume that the total injection force is composed by only two components: an hydrodynamic contribution, which is based on purely shear forces, and a friction force which takes into account the interaction between plunger and syringe barrel. On the other hand, the results presented here suggest that the overall injection force of biopharmaceuticals through syringes also contains a contribution due to purely extensional forces, which cannot be neglected in order for accurate predictions to be computed.
5 Conclusions
Using the microfluidic Rheo-Chip platform, the flow of concentrated (100 g/L \(\le c_{2} \le\) 300 g/L) BSA solutions under repulsive conditions was studied through shear, and contraction flows mimicking the typical geometry of syringes equipped with 26-G and 27-G needles. The shear viscosity retained a Newtonian behavior across the range of investigated shear rates (\(10\, \text {s}^{-1} \le {\dot{\gamma }}, \dot{\eta _{\text {ext.}}} \le 10^{4}\, \text {s}^{-1}\)), which was attributed to the Péclet number being below unity. Similarly, the flow of the 300 g/L BSA solution through the 26-G “syringe-on-chip” microdevices appeared to be Newtonian-like up to Pe \(=0.24\), which was likely because the BSA macromolecule retains its stability under the flow conditions imposed here [13].
Importantly, this work shows that the extensional viscosity of concentrated biopharmaceutical solutions also needs to be considered when one calculates the total force of syringe injection tests, because the measured pressure drop of the BSA solutions across the syringe-like microdevice resulted to be well above the purely shear contribution. This implies that the extensional viscosity plays an important role in determining the overall pressure drop of the complex flows underwent by protein solutions through needles, even if it is Newtonian. The extensional contributions are typically neglected in the formulation of predictions of injection forces [1, 7, 15, 48] and should then be considered if more realistic calculations have to be implemented.
Additionally, because the typical shear rate experienced by blood ranges from 10 \(\text {s}^{-1}\) in the ascending aorta [16] up to 100 \(\text {s}^{-1}\) in capillaries [25], while its pH \(\approx\) 7.4 [55] is close to what probed here. Therefore, this work also indicates that human serum albumin maintains its native conformation under physiological flow conditions.
References
Allmendinger A, Fischer S, Huwyler J, Mahler HC, Schwarb E, Zarraga IE, Mueller R (2014) Rheological characterization and injection forces of concentrated protein formulations: an alternative predictive model for non-newtonian solutions. Eur J Pharm Biopharm 87(2):318–328
Amin S, Barnett GV, Pathak JA, Roberts CJ, Sarangapani PS (2014) Protein aggregation, particle formation, characterization & rheology. Curr Opin Colloid Interface Sci 19(5):438–449
Arratia PE, Thomas C, Diorio J, Gollub JP (2006) Elastic instabilities of polymer solutions in cross-channel flow. Phys Rev Let 96(14):144502
Axelsson I (1978) Characterization of proteins and other macromolecules by agarose gel chromatography. J Chromatogr A 152(1):21–32
Bekard IB, Asimakis P, Teoh CL, Ryan T, Howlett GJ, Bertolini J, Dunstan DE (2012) Bovine serum albumin unfolds in Couette flow. Soft Matter 8(2):385–389
Binding D (1988) An approximate analysis for contraction and converging flows. J Non-Newton Fluid Mech 27(2):173–189
Burckbuchler V, Mekhloufi G, Giteau AP, Grossiord J, Huille S, Agnely F (2010) Rheological and syringeability properties of highly concentrated human polyclonal immunoglobulin solutions. Eur J Pharm Biopharm 76(3):351–356
Castellanos MM, Pathak JA, Colby RH (2014) Both protein adsorption and aggregation contribute to shear yielding and viscosity increase in protein solutions. Soft Matter 10(1):122–131
Churchich JE (1962) Polarization of fluorescence studies of reduced bovine serum albumin. Arch Biochem Biophys 97(3):574–577
De Gennes PG (1979) Scaling concepts in polymer physics. Cornell University Press
Dear BJ, Hung JJ, Truskett TM, Johnston KP (2017) Contrasting the influence of cationic amino acids on the viscosity and stability of a highly concentrated monoclonal antibody. Pharm Res 34(1):193–207
Dharmaraj VL, Godfrin PD, Liu Y, Hudson SD (2016) Rheology of clustering protein solutions. Biomicrofluidics 10(4):043509
Dobson J, Kumar A, Willis LF, Tuma R, Higazi DR, Turner R, Lowe DC, Ashcroft AE, Radford SE, Kapur N, Brockwell DJ (2017) Inducing protein aggregation by extensional flow. Proc Natl Acad Sci 114(18):4673–4678. https://doi.org/10.1073/pnas.1702724114
Fasano M, Curry S, Terreno E, Galliano M, Fanali G, Narciso P, Notari S, Ascenzi P (2005) The extraordinary ligand binding properties of human serum albumin. IUBMB Life 57(12):787–796
Fischer I, Schmidt A, Bryant A, Besheer A (2015) Calculation of injection forces for highly concentrated protein solutions. Int J Pharm 493(1):70–74. https://doi.org/10.1016/j.ijpharm.2015.07.054
Gabe IT, Gault JH, Ross JRJ, Mason DT, Mills CJ, Schillingford JP, Braunwald E (1969) Measurement of instantaneous blood flow velocity and pressure in conscious man with a catheter-tip velocity probe. Circulation 40(5):603–614
Garidel P, Kuhn AB, Schäfer LV, Karow-Zwick AR, Blech M (2017) High-concentration protein formulations: how high is high? Eur J Pharm Biopharm 119:353–360
Godfrin PD, Zarraga IE, Zarzar J, Porcar L, Falus P, Wagner NJ, Liu Y (2016) Effect of hierarchical cluster formation on the viscosity of concentrated monoclonal antibody formulations studied by neutron scattering. J Phys Chem B 120(2):278–291
Gulati S, Liepmann D, Muller SJ (2008a) Elastic secondary flows of semidilute DNA solutions in abrupt 90 microbends. Phys Rev E 78(3):036314
Gulati S, Muller SJ, Liepmann D (2008b) Direct measurements of viscoelastic flows of DNA in a 2: 1 abrupt planar micro-contraction. J Non-Newton Fluid Mech 155(1):51–66
Haward SJ (2014) Characterization of hyaluronic acid and synovial fluid in stagnation point elongational flow. Biopolymers 101(3):287–305
Haward SJ, McKinley GH (2012) Stagnation point flow of wormlike micellar solutions in a microfluidic cross-slot device: effects of surfactant concentration and ionic environment. Phys Rev E 85(3):031502
Haward SJ, Odell JA, Berry M, Hall T (2011) Extensional rheology of human saliva. Rheol Acta 50(11–12):869–879
Haward SJ, McKinley GH, Shen AQ (2016) Elastic instabilities in planar elongational flow of monodisperse polymer solutions. Sci Rep 6:1–18
Ivanov K, Kalinina M, Levkovich YI (1981) Blood flow velocity in capillaries of brain and muscles and its physiological significance. Microvascu Res 22(2):143–155
Jaishankar A, McKinley GH (2013) Power-law rheology in the bulk and at the interface: quasi-properties and fractional constitutive equations. Proc R Soc A 469:20120284
Jaspe J, Hagen SJ (2006) Do protein molecules unfold in a simple shear flow? Biophys J 91(9):3415–3424
Jorgensen L, Hostrup S, Moeller EH, Grohganz H (2009) Recent trends in stabilising peptides and proteins in pharmaceutical formulation-considerations in the choice of excipients. Expert Opin Drug Deliv 6(11):1219–1230
Kanai S, Liu J, Patapoff TW, Shire SJ (2008) Reversible self-association of a concentrated monoclonal antibody solution mediated by fab-fab interaction that impacts solution viscosity. J Pharm Sci 97(10):4219–4227
Lanzaro A, Yuan XF (2011) Effects of contraction ratio on non-linear dynamics of semi-dilute, highly polydisperse paam solutions in microfluidics. J Non-Newton Fluid Mech 166(17):1064–1075
Lanzaro A, Yuan XF (2014) A quantitative analysis of spatial extensional rate distribution in nonlinear viscoelastic flows. J Non-Newton Fluid Mech 207:32–41
Lanzaro A, Li Z, Yuan XF (2015) Quantitative characterization of high molecular weight polymer solutions in microfluidic hyperbolic contraction flow. Microfluid Nanofluid 18(5–6):819–828
Lanzaro A, Corbett D, Yuan XF (2017) Non-linear dynamics of semi-dilute paam solutions in a microfluidic 3D cross-slot flow geometry. J Non-Newton Fluid Mech 242:57–65
Larson RG (1999) The structure and rheology of complex fluids, vol 150. Oxford University Press, New York
Li Z, Yuan XF, Haward SJ, Odell JA, Yeates S (2011) Non-linear dynamics of semi-dilute polydisperse polymer solutions in microfluidics: a study of a benchmark flow problem. J Non-Newton Fluid Mech 166(16):951–963
Liu J, Nguyen MD, Andya JD, Shire SJ (2005) Reversible self-association increases the viscosity of a concentrated monoclonal antibody in aqueous solution. J Pharm Sci 94(9):1928–1940
Mathaes R, Koulov A, Joerg S, Mahler HC (2016) Subcutaneous injection volume of biopharmaceuticals: pushing the boundaries. J Pharm Sci 105(8):2255–2259
Meinhart C, Wereley S, Gray M (2000) Volume illumination for two-dimensional particle image velocimetry. Measur Sci Technol 11(6):809–14
Meinhart CD, Wereley ST, Santiago JG (1999) PIV measurements of a microchannel flow. Exp Fluids 27(5):414–419
Oliveira MS, Rodd LE, McKinley GH, Alves MA (2008) Simulations of extensional flow in microrheometric devices. Microfluid Nanofluid 5(6):809
Oliveira MSN, Alves MA, Pinho FT, McKinley GH (2007) Viscous flow through microfabricated hyperbolic contractions. Exp Fluids 43(2–3):437–451
Pan W, Filobelo L, Pham ND, Galkin O, Uzunova VV, Vekilov PG (2009) Viscoelasticity in homogeneous protein solutions. Phys Rev Lett 102(5):058101
Patel AR, Kerwin BA, Kanapuram SR (2009) Viscoelastic characterization of high concentration antibody formulations using quartz crystal microbalance with dissipation monitoring. J Pharm Sci 98(9):3108–3116
Pathak JA, Hudson SD (2006) Rheo-optics of equilibrium polymer solutions: wormlike micelles in elongational flow in a microfluidic cross-slot. Macromolecules 39(25):8782–8792
Perkins TT, Smith DE, Chu S (1997) Single polymer dynamics in an elongational flow. Science 276(5321):2016–2021
Pipe CJ, Majmudar TS, McKinley GH (2008) High shear rate viscometry. Rheol Acta 47(5–6):621–642
Ramallo N, Paudel S, Schmit J (2019) Cluster formation and entanglement in the rheology of antibody solutions. J Phys Chem B 123(18):3916–3923
Rathore N, Pranay P, Bernacki J, Eu B, Ji W, Walls E (2012) Characterization of protein rheology and delivery forces for combination products. J Pharm Sci 101(12):4472–4480
Rodd L, Cooper-White J, Boger D, McKinley G (2007) Role of the elasticity number in the entry flow of dilute polymer solutions in micro-fabricated contraction geometries. J Non-Newton Fluid Mech 143(2):170–191
Rodd LE, Scott TP, Boger DV, Cooper-White JJ, McKinley GH (2005) The inertio-elastic planar entry flow of low-viscosity elastic fluids in micro-fabricated geometries. J Non-Newton Fluid Mech 129(1):1–22
Roos M, Ott M, Hofmann M, Link S, Rossler E, Balbach J, Krushelnitsky A, Saalwachter K (2016) Coupling and decoupling of rotational and translational diffusion of proteins under crowding conditions. J Am Chem Soc 138(32):10365–10372
Santiago JG, Wereley ST, Meinhart CD, Beebe D, Adrian RJ (1998) A particle image velocimetry system for microfluidics. Exp Fluids 25(4):316–319
Sarangapani PS, Hudson SD, Migler KB, Pathak JA (2013) The limitations of an exclusively colloidal view of protein solution hydrodynamics and rheology. Biophys J 105(10):2418–2426
Schneider S, Nuschele S, Wixforth A, Gorzelanny C, Alexander-Katz A, Netz R, Schneider M (2007) Shear-induced unfolding triggers adhesion of von willebrand factor fibers. Proc Natl Acad Sci 104(19):7899–7903
Severinghaus J (1958) Oxyhemoglobin dissociation curve correction for temperature and pH variation in human blood. J Appl Physiol 12(3):485–486
Shah M, Corbett D, Lanzaro A, Roche A, Sibanda N, Davis P, Uddin S, van der Walle C, Curtis R, Pluen A (2020) Micro-and macro-viscosity relations in high concentration antibody solutions. Eur J Pharm Biopharm 153:211–221
Sharma V, Jaishankar A, Wang YC, McKinley GH (2011) Rheology of globular proteins: apparent yield stress, high shear rate viscosity and interfacial viscoelasticity of bovine serum albumin solutions. Soft Matter 7(11):5150–5160
Shire SJ, Shahrokh Z, Liu J (2004) Challenges in the development of high protein concentration formulations. J Pharma Sci 93(6):1390–1402
Sønderby P, Bukrinski JT, Hebditch M, Peters GH, Curtis RA, Harris P (2018) Self-interaction of human serum albumin: a formulation perspective. ACS Omega 3(11):16105–16117
Szymczak P, Cieplak M (2006) Stretching of proteins in a uniform flow. J Chem Phys 125(16):164903
Szymczak P, Cieplak M (2007) Proteins in a shear flow. J Chem Phys 127(15):10B618
Viola M, Sequeira J, Seiça R, Veiga F, Serra J, Santos AC, Ribeiro AJ (2018) Subcutaneous delivery of monoclonal antibodies: how do we get there? J Controll Release 286:301–314
Wang WH, Lilyestrom WG, Hu ZY, Scherer TM (2018) Cluster size and quinary structure determine the rheological effects of antibody self-association at high concentrations. J Phys Chem B 122(7):2138–2154
White FM, Corfield I (2006) Viscous fluid flow, vol 3. McGraw-Hill Higher Education, Boston
Yadav S, Sreedhara A, Kanai S, Liu J, Lien S, Lowman H, Kalonia DS, Shire SJ (2011) Establishing a link between amino acid sequences and self-associating and viscoelastic behavior of two closely related monoclonal antibodies. Pharm Res 28(7):1750–1764
Yentis SM, Hirsch NP, Ip J (2018) Anaesthesia and intensive care AZ E-book: an encyclopaedia of principles and practice. Elsevier Health Sciences
Yuan Xf (2016) Rheometry apparatus. US Patent 9,354,152
Zarraga IE, Taing R, Zarzar J, Luoma J, Hsiung J, Patel A, Lim FJ (2013) High shear rheology and anisotropy in concentrated solutions of monoclonal antibodies. J Pharm Sci 102(8):2538–2549
Zhao Y, Cheung P, Shen AQ (2014) Microfluidic flows of wormlike micellar solutions. Adv Colloid Interface Sci 211:34–46
Acknowledgements
This work was funded by the Biotechnology and Biological Sciences Research Council (BB/K011146). A. L. thanks Robin Curtis for useful discussions.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Lanzaro, A. A microfluidic approach to studying the injection flow of concentrated albumin solutions. SN Appl. Sci. 3, 783 (2021). https://doi.org/10.1007/s42452-021-04767-2
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s42452-021-04767-2