Longitudinal stent deformation elongating to left main trunk

  • Case Report
  • Published:
Cardiovascular Intervention and Therapeutics Aims and scope Submit manuscript

Abstract

A 67-year-old woman was admitted with an acute coronary syndrome. Angiographic examination revealed diffuse severe stenosis of the left circumflex artery. A Pressure Wire Certus (St. Jude Medical, St. Paul, MN, USA) was passed to the second post-lateral (PL) branch. To prevent side-branch occlusion, a SION wire (Asahi Intecc Co., Aichi, Japan) was left in the first PL branch, and a SION blue wire (Asahi Intecc) was placed in the second obtuse marginal branch. We implanted an everolimus-eluting stent (PROMUS Element 2.5 × 24 mm, Boston Scientific, Natick, MA, USA) in the culprit lesion. After retrieving the protection wire in first PL branch with resistance, we performed post-dilatation. However, the intravascular ultrasound images showed that the proximal portion of the implanted stent had elongated approximately 2 mm to the left main trunk (LMT), although the position of the distal edge of the stent was unchanged. We decided to additionally place a stent from the ostium of the LMT to the proximal left anterior descending coronary artery, and a biolimus-eluting stent (NOBORI 3.0 × 18 mm, Terumo Co., Tokyo, Japan) was implanted successfully. Longitudinal stent elongation might be caused by the small number of links between the hoops of a stent, originally intended to improve deliverability.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

References

  1. Leibundgut G, Gick M, Toma A, Valina C, Löffelhardt N, Büttner HJ, Neumann FJ. Longitudinal compression of the platinum-chromium everolimus-eluting stent during coronary implantation: predisposing mechanical properties, incidence and predictors in a large patient cohort. Catheter Cardiovasc Interv. 2012;. doi:10.1002/ccd.24472.

    PubMed  Google Scholar 

  2. Ormiston JA, Webber B, Webster MW. Stent longitudinal integrity bench insights into a clinical problem. JACC Cardiovasc Interv. 2011;4(12):1310–7. doi:10.1016/j.jcin.2011.11.002.

    Article  PubMed  Google Scholar 

  3. Kozuma K, Kimura T, Kadota K, Suwa S, Kimura K, Iwabuchi M, Kawai K, Miyazawa A, Kawamura M, Nakao K, Asano R, Yamamoto T, Suzuki N, Aoki J, Kyono H, Nakazawa G, Tanabe K, Morino Y, Igarashi K. Angiographic findings of everolimus-eluting as compared to sirolimus-eluting stents: angiographic sub-study from the Randomized Evaluation of Sirolimus-eluting versus Everolimus-eluting stent Trial (RESET). Cardiovasc Interv Ther. 2013;28(4):344–51 (Epub 2013 Apr 23).

    Article  CAS  PubMed  Google Scholar 

  4. Williams PD, Mamas MA, Morgan KP, El-Omar M, Clarke B, Bainbridge A, Fath-Ordoubadi F, Fraser DG. Longitudinal stent deformation: a retrospective analysis of frequency and mechanisms. EuroIntervention. 2012;8(2):267–74.

    Article  PubMed  Google Scholar 

  5. Robinson AD, Schreiber TL, Sobh MA, Grines CL. Deformation, longitudinal shortening, and accordion of an ion stent. J Interv Cardiol. 2011;24(6):493–5.

    Article  PubMed  Google Scholar 

  6. Kereiakes DJ, Popma JJ, Cannon LA, Kandzari DE, Kimmelstiel CD, Meredith IT, Teirstein PS, Verheye S, Allocco DJ, Dawkins KD, Stone GW. Longitudinal stent deformation: quantitative coronary angiographic analysis from the PERSEUS and PLATINUM randomised controlled clinical trials. EuroIntervention. 2012;8(2):187–95.

    Article  PubMed  Google Scholar 

  7. Goto Y, Kawasaki T, Koga H. A rare complication with remarkable stent shortening and successful recovery from the trouble. Cardiovasc Interv Ther. 2012;27(2):114–7.

    Article  PubMed  Google Scholar 

  8. Mamas MA, Williams PD. Longitudinal stent deformation: insights on mechanisms, treatments and outcomes from the Food and Drug Administration Manufacturer and User Facility Device Experience database. EuroIntervention. 2012. pii:20111129-01.

Download references

Conflict of interest

None.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Kensuke Matsushita.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Matsushita, K., Arakawa, K., Hibi, K. et al. Longitudinal stent deformation elongating to left main trunk. Cardiovasc Interv and Ther 29, 247–251 (2014). https://doi.org/10.1007/s12928-013-0220-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s12928-013-0220-x

Keywords

Navigation