Abstract
The aim of the study is to investigate the applicability of a newly developed corneal topographer in assessing tear film stability in Indian subjects. A prospective comparative study of 25 Indian subjects with dry eyes attending a tertiary eye care clinic in South India and 25 normal control subjects was conducted. The diagnosis of dry eye was made based on ocular surface disease index questionnaire. Non-invasive tear film break-up time (NI-TBUT) was measured using a new method based on a corneal topographer equipped with modified scan software. The correlations between the NI-TBUT and the traditional fluorescein tear film break-up time (F-TBUT), Schirmer I test values were determined. A total of 50 patients (100 eyes) were included. The values of NI-TBUT were significantly lower than the values of F-TBUT in both the cases (NI-TBUT 5.78 ± 0.8 s and F-TBUT 7.56 ± 0.5 s; p < 0.02) and controls (NI-TBUT 11.66 ± 1 s and F-TBUT 12.92 ± 1.2 s; p < 0.01). NI-TBUT values were significantly lower than the corresponding F-TBUT values in the varying grades of dry eyes. The mean NI-TBUT values in mild dry eyes was 6.42 ± 0.2 s, moderate dry eyes was 4.70 ± 0.3 s and in severe dry eyes was 2.32 ± 1.2 s. There was a significant difference in the NI-TBUT values for cases and controls (p < 0.001). There was a good correlation seen between the NI-TBUT values and the F-TBUT values, Schirmer I values and the ODSI scores. NI-TBUT was found to have a sensitivity of 86.1 % and a specificity of 81.1 % when the cut-off value was kept at 6.2 s. We investigated the performance of a non-invasive technique for measuring tear film stability to aid in the diagnosis of dry eye disease. It is a useful non-invasive objective method for the detection of dry eye, and its varying grades and may be useful in monitoring the efficacy of therapies for dry eye.
Similar content being viewed by others
References
Lemp MA (2007) The definition and classification of dry eye disease: report of the Definition and Classification Subcommittee of the International Dry Eye WorkShop. Ocul Surf 5:75–135
Serin D, Karsloglu S, Kyan A, Alagöz G (2007) A simple approach to the repeatability of the Schirmer test without anesthesia: eyes open or closed. Cornea 26:903–906
Clinch TE, Benedetto DA, Felberg NT et al (1983) Schirmer’s test. A closer look. Arch Ophthalmol 101:1383–1386
Lemp MA, Hamill JR Jr (1973) Factors affecting tear film breakup in normal eyes. Arch Ophthalmol 89:103–105
Lee JH, Kee CW (1988) The significance of tear film break-up time in the diagnosis of dry eye syndrome. Korean J Ophthalmol 2:69–71
Mengher LS, Bron AJ, Tonge SR, Gilbert DJ (1985) Effect of fluorescein instillation on the pre-corneal tear film stability. Curr Eye Res 4:9–12
Cho P, Brown B, Lau C (1996) Effect of fluorescein on the tear stability of Hong Kong-Chinese. Optom Vis Sci 73:1–7
Mengher LS, Bron AJ, Tonge SR, Gilbert DJ (1985) A non-invasive instrument for clinical assessment of the pre-corneal tear film stability. Curr Eye Res 4:1–7
Mengher LS, Pandher KS, Bron AJ (1986) Non-invasive tear film break-up time: sensitivity and specificity. Acta Ophthalmol (Copenh) 64:441–444
Hirji N, Patel S, Callander M (1989) Human tear film pre rupture phase time (TP RPT) a non invasive technique for evaluating the pre corneal tear film using a novel keratometer mire. Ophthal Physiol Opt 9:139–142
Cho P, Ho KY, Huang YC, Chui HY, Kwan MC (2004) Comparison of noninvasive tear break-up time measurements from black and white background instruments. Optom Vis Sci 81:436–441
Yokoi N, Komuro A (2004) Non-invasive methods of assessing the tear film. Exp Eye Res 78:399–407
Guillon JP (1998) Non-invasive tearscope plus routine for contact lens fitting. Cont Lens Ant Eye 21:S31–S40
Nichols JJ, Nichols KK, Puent B et al (2002) Evaluation of tear film interference patterns and measures of tear break-up time. Optom Vis Sci 79:363–369
Wang J, Fonn D, Simpson TL, Jones L (2003) Precorneal and pre-and post lens tear film thickness measured indirectly with optical coherence tomography. Invest Ophthalmol Vis Sci 44(6):2524–2528
Goto E, Tseng SC (2003) Kinetic analysis of tear interference images in aqueous tear deficiency dry eye before and after punctal occlusion. Invest Ophthalmol Vis Sci 44:1897–1905
Goto E, Tseng SC (2003) Differentiation of lipid tear deficiency dry eye by kinetic analysis of tear interference images. Arch Ophthalmol 121:173–180
Paiva CS, Lindsey J, Pflugfelder SC (2003) Assessing the severity of keratitis sicca with videokeratoscopic indices. Ophthalmology 110:1102–1109
Liu Z, Pflugfelder SC (1999) Corneal surface regularity and the effect of artificial tears in aqueous tear deficiency. Ophthalmology 106:939–943
Wilson SE, Klyce SD (1991) Advances in the analysis of corneal topography. Surv Ophthalmol 35:269–277
Goto T, Zheng X, Klyce SD et al (2003) A new method for tear film stability analysis using videokeratography. Am J Ophthalmol 135:607–612
Goto T, Zheng Z, Okamoto S, Ohashi Y (2004) Tear film stability analysis system: introducing a new application for videokeratography. Cornea 23:65–70
Kojima T, Ishida R, Dogru M et al (2004) A new noninvasive tear stability analysis system for the assessment of dry eyes. Invest Ophthalmol Vis Sci 45:1369–1374
Gumus K, Crockett CH, Rao K, Yeu E, Weikert MP, Shirayama M, Pflugfelder S (2011) Noninvasive assessment of tear stability with the tear stability analysis system in tear dysfunction patients. Invest Ophthal V Sci 52(1):456–461
Hong J, Sun X, Wei A, Cui X, Li Y, Qian T (2013) Assessment of tear film stability in dry eye with a newly developed keratograph. Cornea 32(5):716–721
Schiffman RM, Dale Christianson M, Jacobse G et al (2000) Reliability and validity of the ocular surface disease index. Arch Ophthalmol 118:615–621
Nichols KK, Nichols JJ, Mitchell GL (2004) The lack of association between signs and symptoms in patients with dry eye disease. Cornea 23:762–770
Nichols KK, Mitchell GL, Zadnik K (2004) The repeatability of clinical measurements of dry eye. Cornea 23:272–285
Sweeney DF, Millar TJ, Raju SR (2013) Tear film stability: a review. Exp Eye Res 117:28–38
Goto E, Ishida R, Kaido M et al (2006) Optical aberrations and visual disturbances associated with dry eye. Ocul Surf 4:207–213
Lin YY, Carrel H, Wang IJ et al (2005) Effect of tear film break-up on higher order aberrations of the anterior cornea in normal, dry, and post-LASIK eyes. J Refract Surg 21:S525–S529
Cho P, Douthwaite W (1995) The relation between invasive and noninvasive tear break-up time. Optom Vis Sci 72:17–22
Cho, P (1993) Reliability of a portable noninvasive tear break-up time test on Hong Kong-Chinese. Optom Vis Sci 70(12):1049–1054
Szczesna-Iskander DH, Iskander DR (2012) Future directions in non-invasive measurements of tear film surface kinetics. Optom Vis Sci 89(5):749–759
Best N, Drury L, Wolffsohn JS (2012) Clinical evaluation of the Oculus Keratograph. Cont Lens Anter Eye 35:171–174
Wiedemann D Development and realization of a new non-invasive method for tear film assessment using a corneal topographer. Available at: http://www.oculus.de/en/downloads/dyn/sonstige/sonstige/oculus_keratograph_ sonderdruck_kontaktlinse_en.pdf. Accessed November 17, 2011
Jones SM, Nischal KK (2013) The non-invasive tear film break-up time in normal children. Br J Ophthalmol bjophthalmol
Amaechi OU, Osunwoke CM (2004) The relation between invasive and non-invasive tear break-up time in young adults. J Nigerian Optom Assoc 11(1)
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Bhandari, V., Reddy, J.K., Relekar, K. et al. Non-invasive assessment of tear film stability with a novel corneal topographer in Indian subjects. Int Ophthalmol 36, 781–790 (2016). https://doi.org/10.1007/s10792-016-0186-7
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10792-016-0186-7