Non-invasive assessment of tear film stability with a novel corneal topographer in Indian subjects

  • Original Paper
  • Published:
International Ophthalmology Aims and scope Submit manuscript

Abstract

The aim of the study is to investigate the applicability of a newly developed corneal topographer in assessing tear film stability in Indian subjects. A prospective comparative study of 25 Indian subjects with dry eyes attending a tertiary eye care clinic in South India and 25 normal control subjects was conducted. The diagnosis of dry eye was made based on ocular surface disease index questionnaire. Non-invasive tear film break-up time (NI-TBUT) was measured using a new method based on a corneal topographer equipped with modified scan software. The correlations between the NI-TBUT and the traditional fluorescein tear film break-up time (F-TBUT), Schirmer I test values were determined. A total of 50 patients (100 eyes) were included. The values of NI-TBUT were significantly lower than the values of F-TBUT in both the cases (NI-TBUT 5.78 ± 0.8 s and F-TBUT 7.56 ± 0.5 s; p < 0.02) and controls (NI-TBUT 11.66 ± 1 s and F-TBUT 12.92 ± 1.2 s; p < 0.01). NI-TBUT values were significantly lower than the corresponding F-TBUT values in the varying grades of dry eyes. The mean NI-TBUT values in mild dry eyes was 6.42 ± 0.2 s, moderate dry eyes was 4.70 ± 0.3 s and in severe dry eyes was 2.32 ± 1.2 s. There was a significant difference in the NI-TBUT values for cases and controls (p < 0.001). There was a good correlation seen between the NI-TBUT values and the F-TBUT values, Schirmer I values and the ODSI scores. NI-TBUT was found to have a sensitivity of 86.1 % and a specificity of 81.1 % when the cut-off value was kept at 6.2 s. We investigated the performance of a non-invasive technique for measuring tear film stability to aid in the diagnosis of dry eye disease. It is a useful non-invasive objective method for the detection of dry eye, and its varying grades and may be useful in monitoring the efficacy of therapies for dry eye.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Graph 1
Graph 2
Graph 3

Similar content being viewed by others

References

  1. Lemp MA (2007) The definition and classification of dry eye disease: report of the Definition and Classification Subcommittee of the International Dry Eye WorkShop. Ocul Surf 5:75–135

    Article  Google Scholar 

  2. Serin D, Karsloglu S, Kyan A, Alagöz G (2007) A simple approach to the repeatability of the Schirmer test without anesthesia: eyes open or closed. Cornea 26:903–906

    Article  PubMed  Google Scholar 

  3. Clinch TE, Benedetto DA, Felberg NT et al (1983) Schirmer’s test. A closer look. Arch Ophthalmol 101:1383–1386

    Article  CAS  PubMed  Google Scholar 

  4. Lemp MA, Hamill JR Jr (1973) Factors affecting tear film breakup in normal eyes. Arch Ophthalmol 89:103–105

    Article  CAS  PubMed  Google Scholar 

  5. Lee JH, Kee CW (1988) The significance of tear film break-up time in the diagnosis of dry eye syndrome. Korean J Ophthalmol 2:69–71

    Article  CAS  PubMed  Google Scholar 

  6. Mengher LS, Bron AJ, Tonge SR, Gilbert DJ (1985) Effect of fluorescein instillation on the pre-corneal tear film stability. Curr Eye Res 4:9–12

    Article  CAS  PubMed  Google Scholar 

  7. Cho P, Brown B, Lau C (1996) Effect of fluorescein on the tear stability of Hong Kong-Chinese. Optom Vis Sci 73:1–7

    Article  CAS  PubMed  Google Scholar 

  8. Mengher LS, Bron AJ, Tonge SR, Gilbert DJ (1985) A non-invasive instrument for clinical assessment of the pre-corneal tear film stability. Curr Eye Res 4:1–7

    Article  CAS  PubMed  Google Scholar 

  9. Mengher LS, Pandher KS, Bron AJ (1986) Non-invasive tear film break-up time: sensitivity and specificity. Acta Ophthalmol (Copenh) 64:441–444

    Article  CAS  Google Scholar 

  10. Hirji N, Patel S, Callander M (1989) Human tear film pre rupture phase time (TP RPT) a non invasive technique for evaluating the pre corneal tear film using a novel keratometer mire. Ophthal Physiol Opt 9:139–142

    Article  CAS  Google Scholar 

  11. Cho P, Ho KY, Huang YC, Chui HY, Kwan MC (2004) Comparison of noninvasive tear break-up time measurements from black and white background instruments. Optom Vis Sci 81:436–441

    Article  PubMed  Google Scholar 

  12. Yokoi N, Komuro A (2004) Non-invasive methods of assessing the tear film. Exp Eye Res 78:399–407

    Article  CAS  PubMed  Google Scholar 

  13. Guillon JP (1998) Non-invasive tearscope plus routine for contact lens fitting. Cont Lens Ant Eye 21:S31–S40

    Article  Google Scholar 

  14. Nichols JJ, Nichols KK, Puent B et al (2002) Evaluation of tear film interference patterns and measures of tear break-up time. Optom Vis Sci 79:363–369

    Article  PubMed  Google Scholar 

  15. Wang J, Fonn D, Simpson TL, Jones L (2003) Precorneal and pre-and post lens tear film thickness measured indirectly with optical coherence tomography. Invest Ophthalmol Vis Sci 44(6):2524–2528

    Article  PubMed  Google Scholar 

  16. Goto E, Tseng SC (2003) Kinetic analysis of tear interference images in aqueous tear deficiency dry eye before and after punctal occlusion. Invest Ophthalmol Vis Sci 44:1897–1905

    Article  PubMed  Google Scholar 

  17. Goto E, Tseng SC (2003) Differentiation of lipid tear deficiency dry eye by kinetic analysis of tear interference images. Arch Ophthalmol 121:173–180

    Article  PubMed  Google Scholar 

  18. Paiva CS, Lindsey J, Pflugfelder SC (2003) Assessing the severity of keratitis sicca with videokeratoscopic indices. Ophthalmology 110:1102–1109

    Article  PubMed  Google Scholar 

  19. Liu Z, Pflugfelder SC (1999) Corneal surface regularity and the effect of artificial tears in aqueous tear deficiency. Ophthalmology 106:939–943

    Article  CAS  PubMed  Google Scholar 

  20. Wilson SE, Klyce SD (1991) Advances in the analysis of corneal topography. Surv Ophthalmol 35:269–277

    Article  CAS  PubMed  Google Scholar 

  21. Goto T, Zheng X, Klyce SD et al (2003) A new method for tear film stability analysis using videokeratography. Am J Ophthalmol 135:607–612

    Article  PubMed  Google Scholar 

  22. Goto T, Zheng Z, Okamoto S, Ohashi Y (2004) Tear film stability analysis system: introducing a new application for videokeratography. Cornea 23:65–70

    Article  Google Scholar 

  23. Kojima T, Ishida R, Dogru M et al (2004) A new noninvasive tear stability analysis system for the assessment of dry eyes. Invest Ophthalmol Vis Sci 45:1369–1374

    Article  PubMed  Google Scholar 

  24. Gumus K, Crockett CH, Rao K, Yeu E, Weikert MP, Shirayama M, Pflugfelder S (2011) Noninvasive assessment of tear stability with the tear stability analysis system in tear dysfunction patients. Invest Ophthal V Sci 52(1):456–461

    Article  Google Scholar 

  25. Hong J, Sun X, Wei A, Cui X, Li Y, Qian T (2013) Assessment of tear film stability in dry eye with a newly developed keratograph. Cornea 32(5):716–721

    Article  PubMed  Google Scholar 

  26. Schiffman RM, Dale Christianson M, Jacobse G et al (2000) Reliability and validity of the ocular surface disease index. Arch Ophthalmol 118:615–621

    Article  CAS  PubMed  Google Scholar 

  27. Nichols KK, Nichols JJ, Mitchell GL (2004) The lack of association between signs and symptoms in patients with dry eye disease. Cornea 23:762–770

    Article  PubMed  Google Scholar 

  28. Nichols KK, Mitchell GL, Zadnik K (2004) The repeatability of clinical measurements of dry eye. Cornea 23:272–285

    Article  PubMed  Google Scholar 

  29. Sweeney DF, Millar TJ, Raju SR (2013) Tear film stability: a review. Exp Eye Res 117:28–38

    Article  CAS  PubMed  Google Scholar 

  30. Goto E, Ishida R, Kaido M et al (2006) Optical aberrations and visual disturbances associated with dry eye. Ocul Surf 4:207–213

    Article  PubMed  Google Scholar 

  31. Lin YY, Carrel H, Wang IJ et al (2005) Effect of tear film break-up on higher order aberrations of the anterior cornea in normal, dry, and post-LASIK eyes. J Refract Surg 21:S525–S529

    PubMed  Google Scholar 

  32. Cho P, Douthwaite W (1995) The relation between invasive and noninvasive tear break-up time. Optom Vis Sci 72:17–22

    Article  CAS  PubMed  Google Scholar 

  33. Cho, P (1993) Reliability of a portable noninvasive tear break-up time test on Hong Kong-Chinese. Optom Vis Sci 70(12):1049–1054

    Article  CAS  PubMed  Google Scholar 

  34. Szczesna-Iskander DH, Iskander DR (2012) Future directions in non-invasive measurements of tear film surface kinetics. Optom Vis Sci 89(5):749–759

    Article  PubMed  Google Scholar 

  35. Best N, Drury L, Wolffsohn JS (2012) Clinical evaluation of the Oculus Keratograph. Cont Lens Anter Eye 35:171–174

    Article  CAS  Google Scholar 

  36. Wiedemann D Development and realization of a new non-invasive method for tear film assessment using a corneal topographer. Available at: http://www.oculus.de/en/downloads/dyn/sonstige/sonstige/oculus_keratograph_ sonderdruck_kontaktlinse_en.pdf. Accessed November 17, 2011

  37. Jones SM, Nischal KK (2013) The non-invasive tear film break-up time in normal children. Br J Ophthalmol bjophthalmol

  38. Amaechi OU, Osunwoke CM (2004) The relation between invasive and non-invasive tear break-up time in young adults. J Nigerian Optom Assoc 11(1)

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Vipul Bhandari.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Bhandari, V., Reddy, J.K., Relekar, K. et al. Non-invasive assessment of tear film stability with a novel corneal topographer in Indian subjects. Int Ophthalmol 36, 781–790 (2016). https://doi.org/10.1007/s10792-016-0186-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10792-016-0186-7

Keywords

Navigation