Radiation reduction with prospective ECG-triggering acquisition using 64-multidetector computed tomographic angiography

  • Original Paper
  • Published:
The International Journal of Cardiovascular Imaging Aims and scope Submit manuscript

Abstract

Current 64-multidetector Computed Tomographic scanners (MDCT) utilize retrospective overlapping helical acquisition (RS-OHA) which imparts a higher than desired radiation dose. Although the radiation burden of computed tomographic angiography (CTA) can be efficiently reduced by dose modulation and limiting field of view, a further decrease in radiation without compromising diagnostic image quality would be indeed very desirable. An alternative imaging mode is the axial prospective ECG-triggering acquisition (prospective gating). This study was done to compare the effective radiation dose and the image quality with two techniques to reduce radiation doses with CTA studies utilizing 64-MDCT scanners. The study included 149 consecutive patients (48 females and 101 males) 64-MDCT (mean age = 67 ± 11 years, 72.2% male). Patients underwent CT coronary angiography using one of three algorithms: retrospective triggering with dose modulation; prospective triggering with padding (step and shoot acquisition with additional adjacent phases); and prospective triggering without padding (single phase acquisition only). Based on body habitus, two different voltages were utilized: 100 kVp (<85 kg) or 120 kVp (>85 kg). Radiation doses and image quality (signal to noise ratio) was measured for each patient, and compared between different acquisition protocols. The signal to-noise ratio of the ascending aorta (SNR-AA) was calculated from the mean pixel values of the contrast-filled left ventricular chamber divided by the standard deviation of these pixel values. Use of 100 kVp reduced radiation dose 41.5% using prospective triggering and 39.6% using retrospective imaging as compared to 120 kVp (P < 0.001). Use of prospective imaging reduced radiation exposure by 82.6% as compared to retrospective imaging (P < 0.001). Using both prospective imaging and 100 kVp without padding (single phase data, no other phases obtained), radiation dose was reduced by 90% (P < 0.001). In terms of image quality, the coefficient of variation of ascending aortic contrast enhancement between kVp of 120 and kVp of 100 was 6% (1.05, 95 CI 0.93–1.17), and 7.8% (0.9, 95% CI 0.7–1.2) at the pulmonary artery. The prospective ECG-Triggered acquisition and 100 kVp images were of diagnostic quality, allowing adequate assessment in all patients. CTA using PA and 100 kVp reduced the radiation dose by up to 90% without compromising the image quality.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. Budoff MJ, Achenbach S, Duerinckx A (2003) Clinical utility of computed tomography and magnetic resonance techniques for noninvasive coronary angiography. J Am Coll Cardiol 42(11):1867–1878

    Article  PubMed  Google Scholar 

  2. Cademartiri F, Malagutti P, Belgrano M et al (2005) Non-invasive coronary angiography with 64-slice computed tomography. Minerva Cardioangiol 53(5):465–472 Review

    PubMed  CAS  Google Scholar 

  3. Budoff MJ, Gopal A, Gopalakrishnan D (2006) Cardiac computed tomography: diagnostic utility and integration in clinical practice. Clin Card 29(9 Suppl 1):I4–I14

    Google Scholar 

  4. Flohr TG, Stierstorfer K, Ulzheimer S, Bruder H, Primak AN, McCollough CH (2005) Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot. Med Phys 32(8):2536–2547

    Article  PubMed  CAS  Google Scholar 

  5. Hausleiter J, Meyer T, Hadamitzky M et al (2006) Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates. Circulation 113(10):1305–1310

    Article  PubMed  Google Scholar 

  6. Morin RL, Gerber TC, McCollough CH (2003) Radiation dose in computed tomography of the heart. Circulation 107(6):917–922

    Article  PubMed  Google Scholar 

  7. United Nations (1993) Sources and effects of ionizing radiation, United Nations Scientific Committee on the effects of atomic radiation. Report to the general assembly with scientific annexes. United Nations Sales Publication (E.94.IX.2.)

  8. Budoff MJ, Achenbach S, Blumenthal RS, Carr JJ, Goldin JG, Greenland P, Guerci AD, Lima JAC, Rader DJ, Rubin GD, Shaw LJ, Wiegers SE (2006) Assessment of coronary artery disease by cardiac computed tomography, a scientific statement from the American Heart Association Committee on cardiovascular imaging and intervention, Council on Cardiovascular Radiology and Intervention, and Committee on Cardiac Imaging, Council on Clinical Cardiology. Circulation 114(16):1761–1791

    Article  PubMed  Google Scholar 

  9. Gopal A, Budoff MJ (2007) A new method to reduce radiation exposure during multi-row detector cardiac computed tomographic angiography. Int J Cardiol. doi:10.1016/j.ijcard.2007.08.072

  10. McCollough CH, Bruesewitz MR, Kofler JM Jr (2006) CT dose reduction and dose management tools: overview of available options. Radiographics 26(2):503–512

    Article  PubMed  Google Scholar 

  11. Althen JN (2005) Automatic tube-current modulation in CT–a comparison between different solutions. Radiat Prot Dosim 114(1–3):308–312

    Article  Google Scholar 

  12. Tack D, De Maertelaer V, Gevenois PA (2003) Dose reduction in multidetector CT using attenuation-based online tube current modulation. AJR Am J Roentgenol 181(2):331–334

    PubMed  Google Scholar 

  13. Rizzo S, Kalra M, Schmidt B et al (2006) Comparison of angular and combined automatic tube current modulation techniques with constant tube current CT of the abdomen and pelvis. AJR Am J Roentgenol 186(3):673–679

    Article  PubMed  Google Scholar 

  14. Greess H, Wolf H, Baum U et al (2000) Dose reduction in computed tomography by attenuation based on-line modulation of tube current: evaluation of six anatomical regions. Eur Radiol 10(2):391–394

    Article  PubMed  CAS  Google Scholar 

  15. Hundt W, Rust F, Stabler A, Wolff H, Suess C, Reiser M (2005) Dose reduction in multislice computed tomography. J Comput Assist Tomogr 29(1):140–147

    Article  PubMed  Google Scholar 

  16. Jakobs TF, Becker CR, Ohnesorge B et al (2002) Multislice helical CT of the heart with retrospective EKG gating: reduction of radiation exposure by EKG-controlled tube current modulation. Eur Radiol 12(5):1081–1086

    Article  PubMed  Google Scholar 

  17. Abada HT, Larchez C, Daoud B, Sigal-Cinqualbre A, Paul JF (2006) MDCT of the coronary arteries: feasibility of low-dose CT with EKG-pulsed tube current modulation to reduce radiation dose. AJR Am J Roentgenol 186(6 Suppl 2):S387–S390

    Article  PubMed  Google Scholar 

  18. Mahnken AH, Bruners P, Muhlenbruch G et al (2007) Low tube voltage improves computed tomography imaging of delayed myocardial contrast enhancement in an experimental acute myocardial infarction model. Invest Radiol 42(2):123–129

    Article  PubMed  Google Scholar 

  19. Hamberg LM, Rhea JT, Hunter GJ, Thrall JH (2003) Multi-detector row CT: radiation dose characteristics. Radiology 226(3):762–772

    Article  PubMed  Google Scholar 

  20. Schussler JM, Grayburn PA (2007) Non-invasive coronary angiography using multislice computed tomography. Heart 93(3):290–297 Epub 2005 PMID: 16387814

    Article  PubMed  Google Scholar 

  21. Shuman WP, Branch KR, May JM, Mitsumori LM, Lockhart DW, Dubinsky TJ, Warren BH, Caldwell JH (2008) Prospective versus retrospective ECG gating for 64-detector CT of the coronary arteries: comparison of image quality and patient radiation dose. Radiology 248(2):431–437 Epub 2008 Jun 13

    Article  PubMed  Google Scholar 

  22. Hirai N, Horiguchi J, Fujioka C, Kiguchi M, Yamamoto H, Matsuura N, Kitagawa T, Teragawa H, Kohno N, Ito K (2008) Prospective versus retrospective ECG-gated 64-detector coronary CT angiography: assessment of image quality, stenosis, and radiation dose. Radiology 248(2):424–430 Epub 2008 Jun 23

    Article  PubMed  Google Scholar 

  23. Earls JP, Berman EL, Urban BA, Curry CA, Lane JL, Jennings RS, McCulloch CC, Hsieh J, Londt JH (2008) Prospectively gated transverse coronary CT angiography versus retrospectively gated helical technique: improved image quality and reduced radiation dose. Radiology 246(3):742–753 Epub 2008 Jan 14

    Article  PubMed  Google Scholar 

  24. Brodoefel H, Reimann A, Burgstahler C et al (2008) Noninvasive coronary angiography using 64-slice spiral computed tomography in an unselected patient collective: effect of heart rate, heart rate variability and coronary calcifications on image quality and diagnostic accuracy. Eur J Radiol 66:134–141

    Article  PubMed  CAS  Google Scholar 

  25. European Commission (2000) European guidelines for quality criteria for CT. European Commission, Luxembourg

  26. International Commission on Radiological Protection (1990–1991) Recommendations of the International Commission on radiological protection, 60. Annals of the ICRP 21:1–3. Pergamon Press; ICRP Publication 60, Oxford

  27. McCollough CH (2003) Patient dose in cardiac computed tomography. Herz 28(1):1–6

    Article  PubMed  Google Scholar 

  28. Dowe D (2007) Low-dose ‘quadruple rule-out’ portends era of half-body CTA Diagnostic imaging pp 22–24

  29. Budoff MJ, Dowe D, Jollis JG, Gitter M, Sutherland J, Halamert E, Scherer M, Bellinger R, Martin A, Benton R, Delago A, Min JK (2008) Diagnostic performance of 64-detector row coronary computed tomographic angiography of individuals undergoing invasive coronary prospective multicenter ACCURACY (Assessment by coronary computed individuals without known coronary artery disease: results from the tomographic angiography for evaluation of coronary Artery stenosis in angiography) trial. J Am Coll Cardiol 52(21):1724–1732

    Article  PubMed  Google Scholar 

  30. Perisinakis K, Theocharopoulos N, Karkavitsas N, Damilakis J (2002) Patient effective radiation dose and associated risk from transmission scans using 153Gd line sources in cardiac spect studies. Health Phys 83(1):66–74 PMID: 12075685

    Article  PubMed  CAS  Google Scholar 

  31. Leschka S, Stolzmann P, Schmid FT, Scheffel H, Stinn B, Marincek B, Alkadhi H, Wildermuth S (2008) Low kilovoltage cardiac dual-source CT: attenuation, noise, and radiation dose. Eur Radiol 18:1809–1817

    Article  PubMed  Google Scholar 

  32. Gutstein A, Dey D, Cheng V et al (2008) Algorithm for radiation dose reduction with helical dual source coronary computed tomography angiography in clinical practice. J Cardiovasc Comp Tomogr 2:311–322

    Article  Google Scholar 

  33. Stolzmann P, Leschka S, Scheffel H et al (2008) Dual-source CT in step-and-shoot mode: noninvasive coronary angiography with low radiation dose. Radiology 249:71–80

    Article  PubMed  Google Scholar 

  34. Husmann L, Valenta I, Gaemperli O, Adda O, Treyer V, Wyss CA, Veit-Haibach P, Tatsugami F, von Schulthess GK, Kaufmann PA (2008) Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating. Eur Heart J 29(2):191–197

    Article  PubMed  Google Scholar 

  35. Flohr TG, Schoepf UJ, Kuettner A, Halliburton S, Bruder H, Suess C, Schmidt B, Hofmann L, Yucel EK, Schaller S (2003) Ohnesorge BM: advances in cardiac imaging with 16-section CT systems. Acad Radiol 10:386–401

    Article  PubMed  Google Scholar 

  36. Gutstein A, Wolak A, Lee C et al (2008) Predicting success of prospective and retrospective gating with dual-source coronary computed tomography angiography: development of selection criteria and initial experience. J Cardiovasc Comp Tomogr 2:81–90

    Google Scholar 

  37. Kalra MK, Brady TJ (2008) Current status and future directions in technical developments of cardiac computed tomography. J Cardiovasc Comp Tomogr 2:71–80

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Matthew J. Budoff.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Gopal, A., Mao, S.S., Karlsberg, D. et al. Radiation reduction with prospective ECG-triggering acquisition using 64-multidetector computed tomographic angiography. Int J Cardiovasc Imaging 25, 405–416 (2009). https://doi.org/10.1007/s10554-008-9396-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10554-008-9396-z

Keywords

Navigation