Abstract
Ultrasonography (US) is the first-line diagnostic tool for observing the whole abdomen. Unfortunately, a wide spectrum of refraction-related artefactual images is very frequently encountered in routine US examinations. In addition, most practitioners currently perform abdominal US examinations without sufficient knowledge of refraction artifacts (RAs). This review article was designed to present many representative RA images seen in the clinical setting, with a brief explanation of the mechanism of these images, in certain cases through an analyzed and reconstructed method using computer simulation that supports clinical observations. RAs are encountered not only with B-mode US but also with Doppler US, contrast-enhanced US, and shear wave elastography. RAs change their appearance according to the situation, but they always have a significant effect on detailed interpretation of abdominal US images. Correct diagnosis of abdominal US relies on a deep understanding of each characteristic artifactual finding, which necessitates knowledge of basic US physics. When analyzing mass lesions, computer simulation analysis helps to reveal the global images of RAs around a lesion.
Similar content being viewed by others
References
Quien MM, Saric M. Ultrasound imaging artifacts: how to recognize them and how to avoid them. Echocardiography. 2018;35:1388–401.
Konno K, Ishida H, Komatsuda T, et al. Image distortion in US-guided liver tumor puncture with curved linear array. Eur Radiol. 2003;13:1291–6.
Sato M, Ishida H, Konno K, et al. Analysis of refractive artifacts by reconstructed three-dimensional ultrasound imaging. J Med Ultrason. 2001;2006(33):11–6.
Sato M, Ishida H, Konno K, et al. Analysis of posterior echoes using reconstructed vertical ultrasound images. J Med Ultrason. 2001;2006(33):85–90.
Konno K, Ishida H, Sato M, et al. Liver tumors in fatty liver: difficulty in ultrasonographic interpretation. Abdom Imaging. 2001;26:487–91.
Ishida H, Konno K, Ishida J, et al. Splenic lymphoma: differentiation from splenic cyst with ultrasonography. Abdom Imaging. 2001;26:529–32.
Franquet T, Bescos JM, Barberena J, et al. Acoustic artifacts and reverberation shadows in gallbladder sonograms: their cause and clinical implications. Gastrointest Radiol. 1990;15:223–8.
Naganuma H, Ishida H, Funaoka M, et al. Mobile echoes in liver cysts: a form of range-ambiguity artifact. J Clin Ultrasound. 2010;38:475–9.
Naganuma H, Ishida H, Nagai H, et al. Range-ambiguity artifact in abdominal ultrasound. J Med Ultrason. 2001;2019(46):317–24.
Dawkins A, George N, Ganesh H, et al. Radiologist and sonographer interpretation discrepancies for biliary sonographic findings. Our experience. Ultrasound Q. 2017;33:261–4.
Moore CL, Copel JA. Point-of-care ultrasonography. N Engl J Med. 2011;364:749–57.
Naganuma H, Ishida H, Kuroda H, et al. Hereditary hemorrhagic telangiectasia: how to efficiently detect hepatic abnormalities using ultrasonography. J Med Ultrasonics. 2020;47:421–33.
Zagzebski JA. Physics of diagnostic ultrasound. In: Zagzebski JA, editor. Essentials of ultrasound physics. St Louis: Mosby; 1996. p. 1–45.
Hedrick WR, Hykes DL, Starchman DE. Basic ultrasound physics. In: Hedrick WR, Hykes DL, Starchman DE, editors. Ultrasound Physics and Instrumentation. St Louis: Mosby; 2005. p. 1–22.
Ishida H, Yagisawa H, Naganuma H, et al. Rotation of the diaphragmatic echo behind the liver tumor. Clinical significance and computer analysis. Eur J Ultrasound. 1996;3:267–75.
Ziskin MC, LaFollette PS, Blathras K, et al. Effect of scan format on refraction artifacts. Ultrasound Med Biol. 1990;16:183–91.
Steel R, Poepping TL, Thompson RS, et al. Origins of the edge shadowing artifact in medical ultrasound imaging. Ultrasound Med Biol. 2004;30:1153–62.
Sauerbrei EE, Nguyen KT, Nolan RL. The liver. In: Sauerbrei EE, Nguyen KT, Nolan RL, editors. Abdominal sonography. New York: Raven Press; 1992. p. 73–110.
Makuuchi M, Miyagawa S. Liver intraoperative ultrasound. In: Solbiati L, Rizzatto G, editors. Ultrasound of superficial structures. New York: Churchill Livingstone; 1995. p. 393–402.
Goss SA, Johnston RL, Dunn F. Comprehensive compilation of empirical ultrasonic properties of mammalian tissues. J Acoust Soc Am. 1978;64:423–57.
Uno A, Ishida H, Komatsuda T, et al. Computer analysis of lateral shadowing of capsulated liver tumor. Jpn J Med Ultrasonics. 2004;31suppl:S180.
Kim KW, Choi BI, Yoo SY, et al. Real-time compound ultrasonography: pictorial review of technology and the preliminary experience in clinical application of the abdomen. Abdom Imaging. 2004;29:491–7.
Ishida H, Yagisawa H, Morikawa P, et al. Signe du Drapeau: nouvel aspect échographique de la cirrhose macronodulaire. JEMU. 1988;9:133–8.
Claudon M, Dietrich CF, Choi BI, et al. Guidelines and good clinical practice recommendations for contrast enhanced ultrasound (CEUS) in the liver–update 2012: a WFUMB-EFSUMB initiative in cooperation with representatives of AFSUMB, AIUM, ASUM. FLAUS and ICUS Ultraschall Med. 2013;34:11–29.
Dietrich CF, Ignee A, Greis C, et al. Artifacts and pitfalls in contrast-enhanced ultrasound of the liver. Ultraschall Med. 2014;35:108–28.
Fetzer DT, Rafailidis V, Peterson C, et al. Artifacts in contrast-enhanced ultrasound: a pictorial essay. Abdom Radiol. 2018;43:977–97.
Ishida H, Yagisawa H, Morikawa P, et al. Raindrops sign-a sonographic sign suggesting Liver cirrhosis with multiple focal fatty nodules. Jpn J Med Ultrasonics. 1988;15:S123–4.
Kamiyama N, Sumino Y, Maruyama K, et al. Discussion of mechanism behind “bamboo blind” sign observed in fatty liver parenchyma. Jpn J Med Ultrasonics. 2016;43:655–62.
Hadzik R, Bombiński P, Brzewski M. Double aorta artifact in sonography—a diagnostic challenge. J Ultrason. 2017;17:36–40.
Konno K, Ishida H, Uno A, et al. Accuracy of liver puncture under US guidance: re-evaluation by microcomputer simulation model. J Clin Ultrasound. 1997;25:127–32.
Midia M, Odedra D, Shuster A, et al. Predictors of bleeding complications following percutaneous image-guided liver biopsy: a scoping review. Diagn Interv Radiol. 2019;25:71–80.
Rockey DC, Caldwell SH, Goodman ZD, et al. Liver biopsy. Hepatology. 2009;49:1017–44.
Hedrick WR, Hykes DL, Starchman DE. Doppler imaging. In: Hedrick WR, Hykes DL, Starchman DE, editors. Ultrasound physics and instrumentation. St Louis: Mosby; 2005. p. 239–58.
Uno A, Ishida H, Konno K, et al. Post-tumoral distorted vascular images: diagnostic problem of sonogram. J Med Ultrasonics. 2001;28:89–96.
Uno A, Ishida H, Konno K, et al. Errors in measuring blood flow velocity behind hepatic mass lesions using color Doppler sonography. J Med Ultrasonics. 2003;30:133–40.
Ferraioli G, Wong VW, Castera L, et al. Liver ultrasound elastography: an update to the world federation for ultrasound in medicine and biology guidelines and recommendations. Ultrasound Med Biol. 2018;44:2419–40.
Dietrich CF, Bamber J, Berzigotti A, et al. EFSUMB guidelines and recommendations on the clinical use of liver ultrasound elastography, update 2017 (long version). Ultraschall Med. 2017;38:e16-47.
Jian ZC, Long JF, Liu YJ, et al. Diagnostic value of two dimensional shear wave elastography combined with texture analysis in early liver fibrosis. World J Clin Cases. 2019;26:1122–32.
Park HS, Choe WH, Han HS, et al. Assessing significant fibrosis using imaging-based elastography in chronic hepatitis B patients: pilot study. World J Gastroenterol. 2019;25:3256–67.
Abe T, Kuroda H, Fujiwara Y, et al. Expand accuracy of 2D shear wave elastography in the diagnosis of liver fibrosis in patients with chronic hepatitis C. J Clin Ultrasound. 2018;46:319–27.
Ferraioli G, Tinelli C, Bello BD, et al. Accuracy of real-time shear wave elastography for assessing liver fibrosis in chronic hepatitis C: a pilot study. Hepatology. 2012;56:2125–33.
Leung VY, Shen J, Wong VW, et al. Quantitative elastography of liver fibrosis and spleen stiffness in chronic hepatitis B carriers: comparison of shear-wave elastography and transient elastography with liver biopsy correlation. Radiology. 2013;269:910–8.
Hermann E, Lédinghen V, Cassinotto C, et al. Assessment of biopsy-proven liver fibrosis by two-dimensional shear wave elastography: an individual patient data-based meta-analysis. Hepatology. 2018;67:260–72.
Hu X, Huang X, Chen H, et al. Diagnostic effect of shear wave elastography imaging for differentiation of malignant liver lesions: a meta-analysis. BMC Gastroenterol. 2019;19:60.
Gerber L, Fitting D, Srikantharajah K, et al. Evaluation of 2D- shear wave elastography for characterisation of focal liver lesions. J Gastrointestin Liver Dis. 2017;26:283–90.
Yoon K, Jeong WK, Kim Y, et al. Two-dimensional shear wave elastography: interobserver agreement and factors related to interobserver discrepancies. PLoS ONE. 2017. https://doi.org/10.1371/journal.pone.0175747.
Ryu H, Ahn SJ, Yoon JH, et al. Reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique. Ultrasonography. 2019;38:246–54.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
We have no potential conflicts of interest.
Ethics approval
All procedures followed were in accordance with the ethical standards of the responsible Committee on Human Experimentation (Institutional and National) and with the Helsinki Declaration of 1975, as revised in 2008 (5).
Informed consent
Informed consent was obtained from all patients for being included in the study.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
About this article
Cite this article
Naganuma, H., Ishida, H., Uno, A. et al. Refraction artifact on abdominal sonogram. J Med Ultrasonics 48, 273–283 (2021). https://doi.org/10.1007/s10396-021-01097-z
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10396-021-01097-z