Refraction artifact on abdominal sonogram

  • Review Article–Gastroenterology
  • Published:
Journal of Medical Ultrasonics Aims and scope Submit manuscript

Abstract

Ultrasonography (US) is the first-line diagnostic tool for observing the whole abdomen. Unfortunately, a wide spectrum of refraction-related artefactual images is very frequently encountered in routine US examinations. In addition, most practitioners currently perform abdominal US examinations without sufficient knowledge of refraction artifacts (RAs). This review article was designed to present many representative RA images seen in the clinical setting, with a brief explanation of the mechanism of these images, in certain cases through an analyzed and reconstructed method using computer simulation that supports clinical observations. RAs are encountered not only with B-mode US but also with Doppler US, contrast-enhanced US, and shear wave elastography. RAs change their appearance according to the situation, but they always have a significant effect on detailed interpretation of abdominal US images. Correct diagnosis of abdominal US relies on a deep understanding of each characteristic artifactual finding, which necessitates knowledge of basic US physics. When analyzing mass lesions, computer simulation analysis helps to reveal the global images of RAs around a lesion.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8
Fig. 9
Fig. 10

Similar content being viewed by others

References

  1. Quien MM, Saric M. Ultrasound imaging artifacts: how to recognize them and how to avoid them. Echocardiography. 2018;35:1388–401.

    Article  PubMed  Google Scholar 

  2. Konno K, Ishida H, Komatsuda T, et al. Image distortion in US-guided liver tumor puncture with curved linear array. Eur Radiol. 2003;13:1291–6.

    Article  PubMed  Google Scholar 

  3. Sato M, Ishida H, Konno K, et al. Analysis of refractive artifacts by reconstructed three-dimensional ultrasound imaging. J Med Ultrason. 2001;2006(33):11–6.

    Google Scholar 

  4. Sato M, Ishida H, Konno K, et al. Analysis of posterior echoes using reconstructed vertical ultrasound images. J Med Ultrason. 2001;2006(33):85–90.

    Google Scholar 

  5. Konno K, Ishida H, Sato M, et al. Liver tumors in fatty liver: difficulty in ultrasonographic interpretation. Abdom Imaging. 2001;26:487–91.

    Article  CAS  PubMed  Google Scholar 

  6. Ishida H, Konno K, Ishida J, et al. Splenic lymphoma: differentiation from splenic cyst with ultrasonography. Abdom Imaging. 2001;26:529–32.

    Article  CAS  PubMed  Google Scholar 

  7. Franquet T, Bescos JM, Barberena J, et al. Acoustic artifacts and reverberation shadows in gallbladder sonograms: their cause and clinical implications. Gastrointest Radiol. 1990;15:223–8.

    Article  CAS  PubMed  Google Scholar 

  8. Naganuma H, Ishida H, Funaoka M, et al. Mobile echoes in liver cysts: a form of range-ambiguity artifact. J Clin Ultrasound. 2010;38:475–9.

    Article  PubMed  Google Scholar 

  9. Naganuma H, Ishida H, Nagai H, et al. Range-ambiguity artifact in abdominal ultrasound. J Med Ultrason. 2001;2019(46):317–24.

    Google Scholar 

  10. Dawkins A, George N, Ganesh H, et al. Radiologist and sonographer interpretation discrepancies for biliary sonographic findings. Our experience. Ultrasound Q. 2017;33:261–4.

    Article  PubMed  Google Scholar 

  11. Moore CL, Copel JA. Point-of-care ultrasonography. N Engl J Med. 2011;364:749–57.

    Article  CAS  PubMed  Google Scholar 

  12. Naganuma H, Ishida H, Kuroda H, et al. Hereditary hemorrhagic telangiectasia: how to efficiently detect hepatic abnormalities using ultrasonography. J Med Ultrasonics. 2020;47:421–33.

    Article  Google Scholar 

  13. Zagzebski JA. Physics of diagnostic ultrasound. In: Zagzebski JA, editor. Essentials of ultrasound physics. St Louis: Mosby; 1996. p. 1–45.

    Google Scholar 

  14. Hedrick WR, Hykes DL, Starchman DE. Basic ultrasound physics. In: Hedrick WR, Hykes DL, Starchman DE, editors. Ultrasound Physics and Instrumentation. St Louis: Mosby; 2005. p. 1–22.

    Google Scholar 

  15. Ishida H, Yagisawa H, Naganuma H, et al. Rotation of the diaphragmatic echo behind the liver tumor. Clinical significance and computer analysis. Eur J Ultrasound. 1996;3:267–75.

    Article  Google Scholar 

  16. Ziskin MC, LaFollette PS, Blathras K, et al. Effect of scan format on refraction artifacts. Ultrasound Med Biol. 1990;16:183–91.

    Article  CAS  PubMed  Google Scholar 

  17. Steel R, Poepping TL, Thompson RS, et al. Origins of the edge shadowing artifact in medical ultrasound imaging. Ultrasound Med Biol. 2004;30:1153–62.

    Article  PubMed  Google Scholar 

  18. Sauerbrei EE, Nguyen KT, Nolan RL. The liver. In: Sauerbrei EE, Nguyen KT, Nolan RL, editors. Abdominal sonography. New York: Raven Press; 1992. p. 73–110.

    Google Scholar 

  19. Makuuchi M, Miyagawa S. Liver intraoperative ultrasound. In: Solbiati L, Rizzatto G, editors. Ultrasound of superficial structures. New York: Churchill Livingstone; 1995. p. 393–402.

    Google Scholar 

  20. Goss SA, Johnston RL, Dunn F. Comprehensive compilation of empirical ultrasonic properties of mammalian tissues. J Acoust Soc Am. 1978;64:423–57.

    Article  CAS  PubMed  Google Scholar 

  21. Uno A, Ishida H, Komatsuda T, et al. Computer analysis of lateral shadowing of capsulated liver tumor. Jpn J Med Ultrasonics. 2004;31suppl:S180.

    Google Scholar 

  22. Kim KW, Choi BI, Yoo SY, et al. Real-time compound ultrasonography: pictorial review of technology and the preliminary experience in clinical application of the abdomen. Abdom Imaging. 2004;29:491–7.

    CAS  PubMed  Google Scholar 

  23. Ishida H, Yagisawa H, Morikawa P, et al. Signe du Drapeau: nouvel aspect échographique de la cirrhose macronodulaire. JEMU. 1988;9:133–8.

    Google Scholar 

  24. Claudon M, Dietrich CF, Choi BI, et al. Guidelines and good clinical practice recommendations for contrast enhanced ultrasound (CEUS) in the liver–update 2012: a WFUMB-EFSUMB initiative in cooperation with representatives of AFSUMB, AIUM, ASUM. FLAUS and ICUS Ultraschall Med. 2013;34:11–29.

    Article  CAS  PubMed  Google Scholar 

  25. Dietrich CF, Ignee A, Greis C, et al. Artifacts and pitfalls in contrast-enhanced ultrasound of the liver. Ultraschall Med. 2014;35:108–28.

    Article  CAS  PubMed  Google Scholar 

  26. Fetzer DT, Rafailidis V, Peterson C, et al. Artifacts in contrast-enhanced ultrasound: a pictorial essay. Abdom Radiol. 2018;43:977–97.

    Article  Google Scholar 

  27. Ishida H, Yagisawa H, Morikawa P, et al. Raindrops sign-a sonographic sign suggesting Liver cirrhosis with multiple focal fatty nodules. Jpn J Med Ultrasonics. 1988;15:S123–4.

    Google Scholar 

  28. Kamiyama N, Sumino Y, Maruyama K, et al. Discussion of mechanism behind “bamboo blind” sign observed in fatty liver parenchyma. Jpn J Med Ultrasonics. 2016;43:655–62.

    Article  Google Scholar 

  29. Hadzik R, Bombiński P, Brzewski M. Double aorta artifact in sonography—a diagnostic challenge. J Ultrason. 2017;17:36–40.

    Article  PubMed  PubMed Central  Google Scholar 

  30. Konno K, Ishida H, Uno A, et al. Accuracy of liver puncture under US guidance: re-evaluation by microcomputer simulation model. J Clin Ultrasound. 1997;25:127–32.

    Article  CAS  PubMed  Google Scholar 

  31. Midia M, Odedra D, Shuster A, et al. Predictors of bleeding complications following percutaneous image-guided liver biopsy: a scoping review. Diagn Interv Radiol. 2019;25:71–80.

    Article  PubMed  PubMed Central  Google Scholar 

  32. Rockey DC, Caldwell SH, Goodman ZD, et al. Liver biopsy. Hepatology. 2009;49:1017–44.

    Article  PubMed  Google Scholar 

  33. Hedrick WR, Hykes DL, Starchman DE. Doppler imaging. In: Hedrick WR, Hykes DL, Starchman DE, editors. Ultrasound physics and instrumentation. St Louis: Mosby; 2005. p. 239–58.

    Google Scholar 

  34. Uno A, Ishida H, Konno K, et al. Post-tumoral distorted vascular images: diagnostic problem of sonogram. J Med Ultrasonics. 2001;28:89–96.

    Article  Google Scholar 

  35. Uno A, Ishida H, Konno K, et al. Errors in measuring blood flow velocity behind hepatic mass lesions using color Doppler sonography. J Med Ultrasonics. 2003;30:133–40.

    Article  Google Scholar 

  36. Ferraioli G, Wong VW, Castera L, et al. Liver ultrasound elastography: an update to the world federation for ultrasound in medicine and biology guidelines and recommendations. Ultrasound Med Biol. 2018;44:2419–40.

    Article  PubMed  Google Scholar 

  37. Dietrich CF, Bamber J, Berzigotti A, et al. EFSUMB guidelines and recommendations on the clinical use of liver ultrasound elastography, update 2017 (long version). Ultraschall Med. 2017;38:e16-47.

    PubMed  Google Scholar 

  38. Jian ZC, Long JF, Liu YJ, et al. Diagnostic value of two dimensional shear wave elastography combined with texture analysis in early liver fibrosis. World J Clin Cases. 2019;26:1122–32.

    Article  Google Scholar 

  39. Park HS, Choe WH, Han HS, et al. Assessing significant fibrosis using imaging-based elastography in chronic hepatitis B patients: pilot study. World J Gastroenterol. 2019;25:3256–67.

    Article  PubMed  PubMed Central  Google Scholar 

  40. Abe T, Kuroda H, Fujiwara Y, et al. Expand accuracy of 2D shear wave elastography in the diagnosis of liver fibrosis in patients with chronic hepatitis C. J Clin Ultrasound. 2018;46:319–27.

    Article  PubMed  Google Scholar 

  41. Ferraioli G, Tinelli C, Bello BD, et al. Accuracy of real-time shear wave elastography for assessing liver fibrosis in chronic hepatitis C: a pilot study. Hepatology. 2012;56:2125–33.

    Article  PubMed  Google Scholar 

  42. Leung VY, Shen J, Wong VW, et al. Quantitative elastography of liver fibrosis and spleen stiffness in chronic hepatitis B carriers: comparison of shear-wave elastography and transient elastography with liver biopsy correlation. Radiology. 2013;269:910–8.

    Article  PubMed  Google Scholar 

  43. Hermann E, Lédinghen V, Cassinotto C, et al. Assessment of biopsy-proven liver fibrosis by two-dimensional shear wave elastography: an individual patient data-based meta-analysis. Hepatology. 2018;67:260–72.

    Article  CAS  Google Scholar 

  44. Hu X, Huang X, Chen H, et al. Diagnostic effect of shear wave elastography imaging for differentiation of malignant liver lesions: a meta-analysis. BMC Gastroenterol. 2019;19:60.

    Article  PubMed  PubMed Central  Google Scholar 

  45. Gerber L, Fitting D, Srikantharajah K, et al. Evaluation of 2D- shear wave elastography for characterisation of focal liver lesions. J Gastrointestin Liver Dis. 2017;26:283–90.

    Article  PubMed  Google Scholar 

  46. Yoon K, Jeong WK, Kim Y, et al. Two-dimensional shear wave elastography: interobserver agreement and factors related to interobserver discrepancies. PLoS ONE. 2017. https://doi.org/10.1371/journal.pone.0175747.

    Article  PubMed  PubMed Central  Google Scholar 

  47. Ryu H, Ahn SJ, Yoon JH, et al. Reproducibility of liver stiffness measurements made with two different 2-dimensional shear wave elastography systems using the comb-push technique. Ultrasonography. 2019;38:246–54.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Hiroko Naganuma.

Ethics declarations

Conflict of interest

We have no potential conflicts of interest.

Ethics approval

All procedures followed were in accordance with the ethical standards of the responsible Committee on Human Experimentation (Institutional and National) and with the Helsinki Declaration of 1975, as revised in 2008 (5).

Informed consent

Informed consent was obtained from all patients for being included in the study.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Naganuma, H., Ishida, H., Uno, A. et al. Refraction artifact on abdominal sonogram. J Med Ultrasonics 48, 273–283 (2021). https://doi.org/10.1007/s10396-021-01097-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10396-021-01097-z

Keywords

Navigation