Influence of intrastromal corneal ring segment implantation on corneal biomechanical parameters in keratoconic eyes

  • Clinical Investigation
  • Published:
Japanese Journal of Ophthalmology Aims and scope Submit manuscript

Abstract

Purpose

To assess the influence of intrastromal corneal ring segment (ICRS) implantation on corneal biomechanics in patients with keratoconus.

Methods

Patients with keratoconus who underwent femtosecond laser-assisted ICRS (Keraring) implantation were evaluated with the ocular response analyzer (ORA). Metrics of corneal biomechanical parameters, including corneal hysteresis (CH) and corneal resistance factor (CRF) were measured both before and after surgery. The postoperative measurements recorded at the 1st, 3rd and 6th months, and 1st and 2nd years after surgery were analyzed.

Results

The mean CRF was significantly lower in the early postoperative period (1st and 3rd month); however, the difference between the mean preoperative and late postoperative (6th months, 1st and 2nd years) CRF values was not significant. There were no significant differences between the mean preoperative and postoperative CH values.

Conclusions

In the early postoperative period a significant temporary decrease in CRF was observed after the femtosecond laser-assisted ICRS implantation; however, this surgery did not alter the CH significantly in the same time interval. Both CH and CRF did not show significant alteration from preoperative values in the late postoperative period.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1

Similar content being viewed by others

References

  1. Rabinowitz YS. Keratoconus. Surv Ophthalmol. 1998;42:297–319.

    Article  CAS  PubMed  Google Scholar 

  2. Lass JH, Lembach RG, Park SB, Hom DL, Fritz ME, Svilar GM, et al. Clinical management of keratoconus: a multi-center analysis. Ophthalmology. 1990;97:433–45.

    Article  CAS  PubMed  Google Scholar 

  3. Crews MJ, Driebe WT, Stern GA. The clinical management of keratoconus. A 6 year retrospective study. CLAO J. 1994;20:194–7.

    CAS  PubMed  Google Scholar 

  4. Javadi MA, Feizi S, Yazdani S, Mirbabaee F. Deep anterior lamellar keratoplasty versus penetrating keratoplasty for keratoconus: a clinical trial. Cornea. 2010;29:365–71.

    Article  PubMed  Google Scholar 

  5. Colin J, Cochener B, Savary G, Malet F. Correcting keratoconus with intracorneal rings. J Cataract Refract Surg. 2000;26:1117–22.

    Article  CAS  PubMed  Google Scholar 

  6. Burris TE, Ayer CT, Evensen DA, Davenport JM. Effects of intrastromal corneal ring size and thickness on corneal flattening in human eyes. Refract Corneal Surg. 1991;7:46–50.

    CAS  PubMed  Google Scholar 

  7. Alió JL, Shabayek MH, Belda JI, Correas P, Feijoo ED. Analysis of results related to good and bad outcomes of Intacs implantation for keratoconus correction. J Cataract Refract Surg. 2006;32:756–61.

    Article  PubMed  Google Scholar 

  8. Kwitko S, Severo NS. Ferrara intracorneal ring segments for keratoconus. J Cataract Refract Surg. 2004;30:812–20.

    Article  PubMed  Google Scholar 

  9. Coskunseven E, Kymionis GD, Tsiklis NS, Atun S, Arslan E, Jankoy MR, et al. One-year results of intrastromal corneal ring segment implantation (KeraRing) using femtosecond laser in patients with keratoconus. Am J Ophthalmol. 2008;145:775–9.

    Article  PubMed  Google Scholar 

  10. Luce DA. Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J Cataract Refract Surg. 2005;31:156–62.

    Article  PubMed  Google Scholar 

  11. Shah S, Laiquzzaman M, Bhojwani R, Mantry S, Cunliffe I. Assessment of the biomechanical properties of the cornea with the ocular response analyzer in normal and keratoconic eyes. Invest Ophthalmol Vis Sci. 2007;48:3026–31.

    Article  PubMed  Google Scholar 

  12. Kirwan C, O’Keefe M. Corneal hysteresis using the Reichert ocular response analyser: findings pre- and post-LASIK and LASEK. Acta Ophthalmol. 2008;86:215–8.

    Article  PubMed  Google Scholar 

  13. Kucumen RB, Yenerel NM, Gorgun E, Kulacoglu DN, Oncel B, Kohen MC, et al. Corneal biomechanical properties and intraocular pressure changes after phacoemulsification and intraocular lens implantation. J Cataract Refract Surg. 2008;34:2096–8.

    Article  PubMed  Google Scholar 

  14. Pepose JS, Feigenbaum SK, Qazi MA, Sanderson JP, Roberts CJ. Changes in corneal biomechanics and intraocular pressure following LASIK using static, dynamic, and noncontact tonometry. Am J Ophthalmol. 2007;143:39–47.

    Article  PubMed  Google Scholar 

  15. Yenerel NM, Kucumen RB, Gorgun E. Changes in corneal biomechanics in patients with keratoconus after penetrating keratoplasty. Cornea. 2010;29:1247–51.

    Article  PubMed  Google Scholar 

  16. Kucumen RB et al. In vivo confocal microscopic findings of intrastromal ring segment implanted corneas. Poster presentation at ESCRS 15th Winter Meeting. 2011.

  17. Dauwe C, Touboul D, Roberts CJ, Mahmoud AM, Kérautret J, Fournier P, et al. Biomechanical and morphological corneal response to placement of intrastromal corneal ring segments for keratoconus. J Cataract Refract Surg. 2009;35:1761–7.

    Article  PubMed  Google Scholar 

Download references

Acknowledgments

The authors would like to thank Prof. Dr. Ferda Ciftci for her contribution to the critical review of the article.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Ebru Gorgun.

About this article

Cite this article

Gorgun, E., Kucumen, R.B. & Yenerel, N.M. Influence of intrastromal corneal ring segment implantation on corneal biomechanical parameters in keratoconic eyes. Jpn J Ophthalmol 55, 467–471 (2011). https://doi.org/10.1007/s10384-011-0057-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10384-011-0057-8

Keywords

Navigation