Abstract
The aim of the present study was to compare the cyclic fatigue resistances of HyFlex EDM (HEDM), WaveOne Gold (WOG), Reciproc Blue (RB), and 2Shape (TS) NiTi systems having different metallurgic properties. HEDM, WOG, RB, and TS instruments were rotated in artificial canals which were made of stainless steel with an inner diameter of 1.5 mm, 45°, and 90° angles of curvatures and a radius of curvature of 5 mm until fracture occurred, and the time to fracture (TTF) was recorded in seconds. Also, number of cycles to fracture was calculated (NCF). Data were statistically analyzed using Kruskal–Wallis and t test. The statistical significance level was set at P < 0.05. When comparing the TTF of all the instruments tested in the artificial canal with 45° curvature, RB had statistically the highest cyclic fatigue resistance followed by HEDM and TS (P < 0.05), while the WOG showed the lowest cyclic fatigue resistance (P < 0.05). In artificial canal with 90° curvature, RB showed a significantly higher cyclic fatigue resistance than the other tested files (P < 0.05), while there was no significant difference among the TS, WOG and HEDM groups (P > 0.05). When the NCF values were taken into consideration, HEDM reported a significantly higher cyclic fatigue resistance than TS in both canal curvatures analyzed (P < 0.05). Within the limitations of the present study, RB NiTi files showed statistically higher cyclic fatigue resistance in artificial canals with 45° and 90° than the other NiTi files tested.
Similar content being viewed by others
References
Walia H, Brantley WA, Gerstein H. An initial investigation of the bending and torsional properties of Nitinol root canal files. J Endod. 1988;14:346–51.
Paqué F, Ganahl D, Peters OA. Effects of root canal preparation on apical geometry assessed by micro–computed tomography. J Endod. 2009;35:1056–9.
Plotino G, Grande NM, Cordaro M, Testarelli L, Gambarini G. A review of cyclic fatigue testing of nickel-titanium rotary instruments. J Endod. 2009;35:1469–76.
Sattapan B, Nervo GJ, Palamara JE, Messer HH. Defects in rotary nickel-titanium files after clinical use. J Endod. 2000;26:161–5.
Testarelli L, Plotino G, Al-Sudani D, Vincenzi V, Giansiracusa A, Grande NM, et al. Bending properties of a new nickel-titanium alloy with a lower percent by weight of nickel. J Endod. 2011;37:1293–5.
Plotino G, Grande NM, Cotti E, Testarelli L, Gambarini G. Blue treatment enhances cyclic fatigue resistance of vortex nickel-titanium rotary files. J Endod. 2014;40:1451–3.
Peters O, Gluskin A, Weiss R, Han J. An in vitro assessment of the physical properties of novel Hyflex nickel–titanium rotary instruments. Int Endod J. 2012;45:1027–34.
Özyürek T. Cyclic fatigue resistance of Reciproc, WaveOne, and WaveOne gold nickel-titanium instruments. J Endod. 2016;42:1536–9.
Plotino G, Grande NM, Bellido MM, Testarelli L, Gambarini G. Influence of temperature on cyclic fatigue resistance of ProTaper gold and ProTaper Universal rotary files. J Endod. 2017;43:200–2.
De-Deus G, Silva EJNL, Vieira VTL, Belladonna FG, Elias CN, Plotino G, et al. Blue thermomechanical treatment optimizes fatigue resistance and flexibility of the reciproc files. J Endod. 2017;43:462–6.
Haapasalo M, Shen Y. Evolution of nickel–titanium instruments: from past to future. Endod Top. 2013;29:3–17.
Pedullà E, Savio FL, Boninelli S, Plotino G, Grande NM, La Rosa G, et al. Torsional and cyclic fatigue resistance of a new nickel-titanium instrument manufactured by electrical discharge machining. J Endod. 2016;42:156–9.
Gündoğar M, Özyürek T. Cyclic fatigue resistance of OneShape, HyFlex EDM, WaveOne gold, and Reciproc blue nickel-titanium instruments. J Endod. 2017;43:1192–6.
Saleh AM, Gilani PV, Tavanafar S, Schäfer E. Shaping ability of 4 different single-file systems in simulated S-shaped canals. J Endod. 2015;41:548–52.
Topçuoğlu H, Topçuoğlu G, Aktı A. Comparative evaluation of cyclic fatigue resistance of D-RaCe and ProTaper retreatment instruments in curved artificial canals. Int Endod J. 2016;49:604–9.
Plotino G, Costanzo A, Grande NM, Petrovic R, Testarelli L, Gambarini G. Experimental evaluation on the influence of autoclave sterilization on the cyclic fatigue of new nickel-titanium rotary instruments. J Endod. 2012;38:222–5.
Pedullà E, Grande NM, Plotino G, Gambarini G, Rapisarda E. Influence of continuous or reciprocating motion on cyclic fatigue resistance of 4 different nickel-titanium rotary instruments. J Endod. 2013;39:258–61.
Plotino G, Ahmed HMA, Grande NM, Cohen S, Bukiet F. Current assessment of reciprocation in endodontic preparation: a comprehensive review—part II: properties and effectiveness. J Endod. 2015;41:1939–50.
Grande NM, Ahmed HMA, Cohen S, Bukiet F, Plotino G. Current assessment of reciprocation in endodontic preparation: a comprehensive review—part I: historical perspectives and current applications. J Endod. 2015;41:1778–83.
Grande NM, Plotino G, Ahmed HMA, Cohen S, Bukiet F. The reciprocating movement in endodontics. Endod Prac. 2016;9:37–42.
Elnaghy A, Elsaka S. Effect of sodium hypochlorite and saline on cyclic fatigue resistance of WaveOne gold and Reciproc reciprocating instruments. Int Endod J. 2017;50:991–8.
Topçuoğlu H, Düzgün S, Aktı A, Topçuoğlu G. Laboratory comparison of cyclic fatigue resistance of WaveOne Gold, Reciproc and WaveOne files in canals with a double curvature. Int Endod J. 2017;50:713–7.
Cheung G, Zhang E, Zheng Y. A numerical method for predicting the bending fatigue life of NiTi and stainless steel root canal instruments. Int Endod J. 2011;44:357–61.
Grande NM, Plotino G, Pecci R, Bedini R, Somma F. Cyclic fatigue resistance and three-dimensional analysis of instruments from two nickel-titanium rotary systems. Int Endod J. 2006;39:755–63.
Plotino G, Grande NM, Melo MC, Bahia MG, Testarelli L, Gambarini G. Cyclic fatigue of NiTi rotary instruments in a simulated apical abrupt curvature. Int Endod J. 2010;43:226–30.
Plotino G, Grande NM, Testarelli L, Gambarini G. Cyclic fatigue of Reciproc and WaveOne reciprocating instruments. Int Endod J. 2012;45:614–8.
Pruett JP, Clement DJ, Carnes DL. Cyclic fatigue testing of nickel-titanium endodontic instruments. J Endod. 1997;23:77–85.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors deny any conflicts of interest related to this study.
Rights and permissions
About this article
Cite this article
Özyürek, T., Gündoğar, M., Uslu, G. et al. Cyclic fatigue resistances of Hyflex EDM, WaveOne gold, Reciproc blue and 2shape NiTi rotary files in different artificial canals. Odontology 106, 408–413 (2018). https://doi.org/10.1007/s10266-018-0340-y
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10266-018-0340-y