Introduction

Achieving glycemic targets to prevent diabetic complications remains challenging [1]. This is clearly true for adolescents and young adults as indicated in several diabetes registries [2, 3]. Insulin delivery administrated by pumps (continuous subcutaneous insulin infusion, CSII) has several advantages such as flexibility and reduction of severe hypoglycemia [4,5,6]. Use of this therapeutic option increased dramatically in the last decade [7]. A greater superiority of CSII over multiple daily injections (MDI) regimen was notably found in those subjects with the highest HbA1C levels at baseline [8]. Sensor-augmented pumps (SAP) with the addition of interstitial continuous glucose monitoring (CGM) gave additional insights in terms of adjustment to metabolic needs either as a stand-alone device [9, 10] or with predictive low glucose management (PLGM) as shown in randomized clinical trials (10). In addition, CGM has also help to define new glycemic targets for prevention of diabetic complications and hypoglycemia [11]. Hybrid Closed Loop (HCL) systems for people with type 1 diabetes have further improved diabetes control when compared to SAP [12]. Most of the studies that evaluate CSII, SAP or HCL therapies are trials that included selected populations of subjects with high levels of motivation that may not reflect the routine use of these therapies. In addition, modalities of reimbursement may vary across countries with involvement of private insurance payers which may increase social disparity. As with all new technology, it becomes increasingly important to translate in real-world situations the results from controlled studies with the required time to provide adequate education and follow-up. Modalities of CSII initiation differ according to centers and national regulations. In France, reimbursement is limited to inpatients from initiating centers [13]. Since the outbreak of the COVID-19, access to clinics dedicated to diabetes care has been limited during lockdown and activities were largely transitioned to a telehealth/telemedicine model of care. CSII initiation in outpatient settings actively challenges current health care organizations. Here, we report the glycemic and safety outcomes of outpatient pump initiation with telemonitoring that resulted in reduced HbA1C levels and increased times in range values.

Research design and methods

Study design and participants

The study was a single center open-label analysis of clinical outcomes after CSII initiation during free-living conditions after attending DIAB-eCARE (www.diab-ecare.fr) a T1D integrated practice unit of the University Hospital of Lyon (France) between September 2020 and May 2021. At enrollment, participants or parents gave written consent to participate and baseline data were collected. All device training was performed by a diabetes nurse educator during a single visit in addition to the settings of telehealth parameters. The procedure to initiate insulin pump therapy in outpatient settings was authorized and funded by the French Health Authority in contrast to current regulatory procedures [14]. Three categories of participants were identified: first pump users (n = 26) with stand-alone CGM, current CSII users changing their device for SAP with stand-alone CGM (n = 51) or SAP and predictive low glucose suspend (PLGS) capacities using Tandem t:slim X2™ insulin pump with Basal-IQ™ and Dexcom G6® CGM (n = 8) and subjects with Hybrid Closed Loop (HCL) technology (n = 36) using the Medtronic 780G ™ pump and Smartguard 3® with automated correction bolus. All subjects received fast-acting insulin analogs with either insulin Aspart or Lispro according to their previous therapy. Participants had willingness and ability to adhere to the observatory and shared data to health care providers with their smartphone and/or with Diabnext uploader or through CareLink™ Personal (Medtronic) when appropriate. Participants were asked to download their pump daily the first week then twice-weekly thereafter. A diabetes nurse educator monitored glucose and insulin data remotely at least once weekly to allow for real-time assessment of diabetes control and safety. The diabetes team members were alerted via text messages in case of adverse hyperglycemic or hypoglycemic events. Age, sex, duration of diabetes, body mass index (BMI), HbA1C values were recorded as well as glucose metrics with an observation frame of two weeks.

Closed loop system

Participants (n = 36) used the AHCL system in real-life conditions with the MiniMed 780G insulin pump, the guardian CGM components (Guardian sensor 3 and Guardian link transmitter) and a specific blood glucose meter (Accu-Chek® Guide). A minimum of three blood glucose tests per day was recommended for calibration of the sensor in accordance with the labeling of the CGM device. After a wash-out period of 1–2 weeks, insulin-to-carbohydrate ratios were set for each subject by investigators and auto-mode was activated with an initial target glucose set point of 6 mmol/L (110 mg/dl) and an active insulin time of 2 h. These set points were chosen to obtain improved glycemic control without increasing the risk for hypoglycemia. A temporary set point of 8.3 mmol/L (150 mg/dl) was used for exercise. In case of excessive low glucose levels, the glucose set point was increased to 6.7 mmol/L (120 mg/dL) with longer active insulin times to obtain less aggressive dosing.

Study objectives

The primary objective was to evaluate the mean change in HbA1c levels between initial visit and after 3 months of follow-up based on treatment effect. Secondary objectives were percentage of glucose values in target range of 3.9–10 mmol/L (70–180 mg/dL), glycemic variability (mean amplitude of glycemic excursions of sensor glucose values). Additional objectives were to confirm safety by comparing the percentage of glucose values in the percentage of time spent ≤ 3.0 mmol/L (54 mg/dL) and above 250 mg/dl.

Statistical methods

We report mean (SEM) for normally distributed variables and median (IQR) for skewed variables. Categorical variables were summarized by frequencies. Comparisons between baseline characteristics and outcomes after 3 months at home were performed using two-sided paired Student's t-test for normally distributed outcomes or a Wilcoxon signed-rank test for skewed outcomes. The overall differences in percentage TIR between periods were compared by paired t-test and a significance level of 0.025 (one-sided). Comparisons were performed with Chi-square correlations between delta Final-Initial HbA1C values and initial HbA1C levels with Pearson test according to Gaussian distribution. Due to the limited number of subjects with a PLGM system, we pooled these individuals with those with stand-alone CGM in the SAP group for comparison with the HCL group using unpaired t-test or Mann–Whitney when appropriate. All statistical analyses were performed using GraphPad Prism 7.05 software (San Diego, CA) and Medistica Pvalue.io available on https://www.pvalue.io/fr.

Results

Patient characteristics and CGM use

A total of 121 subjects (71 females) with mean age 32.0 [20.2; 41.4] years including 26 subjects aged < 18 years, were enrolled. The median length of diagnosis was 18.0 [9.00; 28.0] years. Baseline demographics of the three categories of participants are described in Table 1. With respect to technology use, 77 subjects were on SAP with stand-alone CGM, eight participants had a SAP + LGSM system and 36 subjects used a HCL system. As shown in Table 1, no difference in age, BMI, gender, baseline HbA1C levels could be noticed between groups. No difference in age, BMI, duration of disease, median initial HbA1C values and percentage of glucose values in TIR was found between groups.

Table 1 Characteristics of outpatients with type 1 diabetes who initiated pump therapy (initiation, n = 26) or changed their pump model (switch, n = 95). Initiation was performed exclusively with sensor-augmented pumps (SAP) and already pump users had either SAP or HCL system. Variables are indicated as median value [25; 75 QI]

Evolution of HbA1C

All groups improved their diabetes control (Fig. 1) but at different levels. A significant reduction was found both in those subjects on MDI who initiated pump therapy with 7.66 ± 0.05% (60.2 ± 1.9 mmol/mol) vs. 8.27 ± 0.26% (66.76 ± 2.8 mmol/mol), p < 0.0001, but also in CSII users who changed their pump model with stand-alone CGM or SAP with 7.5 ± 0.05% (60.54 ± 1.12 mmol/mol) vs. 7.88 ± 0.11% (62.4 ± 0.98 mmol/mol), p < 0.001. During HCL, mean ± SEM HbA1C levels were significantly reduced with 6.84 ± 0.05% (51.24 ± 0.61 mmol/mol) vs. 7.66 ± 0.1% (60.69 ± 1.2 mmol/mol), p < 0.0001 and were lower than subjects with SAP (p < 0.0001) as shown in Table 2. At 3 months, we observed 23/85 (27%) in the SAP group and 33/36 (91%) subjects in the HCL group within the recommended HbA1C levels according to age. Differences between final and initial HbA1C levels were inversely correlated to initial HbA1C levels (Fig. 2) for subjects in the HCL group (R coefficient [95% CI] = − 0,8687 [− 0,9308 to − 0,7581], p = 0.0001) and also for participants who initiated SAP therapy (R coefficient = − 0,5131 [− 0,7564 to − 0,1452], p = 0.0073) and for previously pump users on SAP with R coefficient of − 0,3081 [− 0,5303 to − 0,04631] (p = 0.0186). The results were similar in adolescents (n = 26, age 14 [12;16] years) with a mean reduction of HbA1C levels of − 0.6 ± 0.9% from 8.15% [7.7;8.52] to 7.75% [7.2;8.12], p < 0.0001) including six subjects with HCL and a reduction of − 0.8 ± 0.3% of HbA1C levels.

Fig. 1
figure 1

Evolution of HbA1C levels at baseline (M0) and after 3 months (M3) for subjects with HCL (A) or SAP (B) systems with HCL (A) or SAP (B) systems

Table 2 Evolution of HbA1C levels percentages of time in range (TIR) 3.9–10 mmol/L (70–180 mg/dl) and coefficient of variation of glucose values between baseline at M0 and after 3 months of insulin pump therapy with sensor-augmented pumps (SAP) or Hybrid Closed Loop (HCL) systems
Fig. 2
figure 2

Correlations between initial HbA1C levels and delta values between initial and final HbA1C levels from subjects changing their pump for HCL (closed circles) or SAP (open squares) and those who initiated SAP (open circles). The broken line corresponds to the linear correlation for subjects with HCL

Meeting glycemic targets

A shown in Fig. 3, TIR increased both subjects with SAP or HCL systems. With respect to previous insulin experience, mean ± SEM improvement in TIR 3.9-10 mmol/L (70–180 mg/dL) was similar between subjects who initiated or changed their pump with an open-loop SAP system (57.64 ± 3.36% vs. 53.46 ± 1.8%, p = 0.27). TIR was higher when HCL was used as compared to baseline (75.57 ± 1.2% vs. 49.8 ± 1.76%, p < 0.0001). The proportion of participants with TIR ≥ 70% during follow-up was 14/85 (16.5%) in the SAP group vs. 30/36 (83%) in the HCL group. No significant difference was noticed at 3 months vs baseline in percentage of time < 3 mmol/L (54 mg/dl) for HCL (0.78 ± 0.16% vs. 0.7 ± 0.2%, p = 0.46), first time SAP users (0.9 ± 0.32% vs. 0.75 ± 0.28%, p = 0.84) or experienced users with SAP (0.88 ± 0.17% vs. 1.18 ± 0.27%, p = 0.44). At 3 months, the percentage of time between 3 and 3.9 mmol/L (54–70 mg/dl) was lower during HCL vs. SAP with 2.08 ± 0.26% and 3.36 ± 0.28%, respectively, (p < 0.001). HCL reduced significantly the time spent > 250 mg/dl in comparison to baseline (4 ± 0.49% vs. 14.5 ± 1.74%, p < 0.0001). This reduction was also seen in SAP group (17.25 ± 1.63% vs. 19.3 ± 1.5%, p < 0.001) but was limited. A significant reduction was observed in the number of participants who spent < 25% of time > 10.0 mmol/L (180 mg/dL) after 3 months for subjects with HCL (5/36 vs. 25/36, p < 0.0001) but not for SAP (11/85 vs. 15/85, p = 0.52). A significant reduction in % CV was noticed in the HCL group with 34.1 ± 0.63% vs. 37.7 ± 1.17% (p < 0.05), but not for subjects who initiated pump therapy or changed pump model in the SAP group.

Fig. 3
figure 3

Evolution of % time in range at baseline (M0) and after 3 months (M3) of subjects with HCL (A) or SAP (B) systems

Discussion

We provide the evidence that insulin pump initiation in outpatients is safe and efficient both for the initiation or change of pump model. This organization was designed to meet user’s desire for convenience and improvement in diabetes management. This required the support of a dedicated health care team and telehealth monitoring. Metabolic improvement was heterogeneous but was maximal with Hybrid Closed Loop systems.

Although several guidelines have been published [16, 17], variation in insulin pump initiation regarding education and training modalities can be observed according to center’s practice and national policies. Initiation of insulin pump therapy in an ambulatory setting is not new and has been used successfully in some centers. This requires flexible outpatient care program and 24-h health provider availability for assistance. Outpatient initiation of pump therapy has several advantages since patient's life can more quickly approach to normal with subjects able to work normally, go to school, and remain in their familiar surroundings. Several advantages for hospitalization have been advocated including stopping the long-acting insulin dose the night before pump initiation and progressive acquisition of knowledge with formal pump training sessions spanned during several days. Inpatient initiation is required in France for more than 20 years to get full coverage of patients by health insurance and to provide revenues for both the initiating center and health care providers who ensure the on-call and supply of patients at home [13]. This organization significantly increases the economic burden and has been seriously challenged by the COVID-19 pandemic with the dramatic reduction of hospitalization and educational capacities. The development of telemedicine and availability of connected devices allowed us to propose an alternative model for outpatients with an experimentation taken in charge by the payers [14].

Continuous subcutaneous insulin infusion (CSII) using an external pump is an intensive diabetes therapy recognized to improve metabolic control and to reduce both glycemic instability and frequency of severe hypoglycemia as shown in a meta-analysis of several studies [5]. CSII initiation requires to set basal rates, individualized insulin-to-carbohydrate ratios and sensitivity factors, a sum of educational and clinical factors that are more challenging to succeed in naïve individuals treated by MDI. CSII is also cost-effective vs. MDI across several settings for patients who have poor glycemic control and/or problematic hypoglycemia [18]. However, when mean HbA1C levels were analyzed in large clinic registries according to CGM and type of insulin administration, pump and injection users had similar results using CGM for glucose monitoring with the highest HbA1C levels in adolescent and young adults irrespective of device use [2]. While insulin pump and CGM use increased between 2010 and 2012 and 2016–2018, only a minority of adults and youth with T1D in the USA achieve ADA goals for HbA1c. This was also observed in our population with only 18/121 (14.8%) participants who met initially the recommended HbA1C targets of ≤ 7% for adults [11] or 7.5% for those before 18-year-old [15]. Addition of CGM for people managing their diabetes with an insulin pump has clear benefits as shown in the SWITCH crossover trial [9]. The ability of control devices to provide automatic flow of data to allow caregivers to remotely monitor insulin and glucose data provides additional efficiency as shown previously (19).

Development of insulin pump therapy has several strategic issues and management of diabetes is moving rapidly into the area of sensor-augmented therapy which now encompasses closed loop approaches. As insulin pump therapy is a large contributor to diabetes health care expenditures among insulin treated patients [20], it is urgent to optimize the metabolic results. Although many studies have demonstrated the cost-effectiveness of CSII over MDI including less acute events and hospitalizations, reduction of costs is a necessity with regard to new indications and solutions. This requires appropriate education and follow-up including different devices as well as data transfer and management. It is interesting that we observed significant improvements in all conditions, i.e., naïve vs experienced subjects. We observed significant improvements in HbA1C values with a larger but still insufficient proportion of subjects meeting glycemic targets in those who changed their pump model. As expected, we were able to confirm the superiority of HCL vs SAP systems in the capacity to reach target glucose values. Beside the addition of automated insulin administration, it must be emphasized that HCL requires announced meals with carbohydrate counting which is not the case for SAP. Possible changes in diet profiles may occur according to pump systems but this has not yet been examined.

Changes in model of care cannot come at a cost of greater hypoglycemia. The percentage of time < 3 mmol/L (54 mg/dl) and < 3.9 mmol/L (70 mg/dl) was far below the recommendations, including additional benefits in the HCL group. The improvements observed in our study for patients with previous pump experience clearly underline the added value of carbohydrate counting [21] and telehealth monitoring as previously noticed by our group [19]. This also suggests that conditions for access and follow-up are also important factors for success. It cannot be excluded that our population was selected and strongly motivated to improve their glucose control due the simplified procedures and therefore may not reflect the general population of subjects with diabetes. Initiation of pump therapy in outpatients requires that health care professionals determine which patients are likely to benefit the most from CSII. This implies pre-pump start preparation and planning.

Insulin pump initiation in outpatient settings does not alter the performance of HCL system. Our results after 3 months in the real life confirm the superiority of this automatized insulin administration and the capacity to reach the recommended glucose values. Our data also provide confidence in the system’s performance across different age groups including adolescents with similar results to randomized studies [22]. The capacity to provide continuous patient education with telemedicine both for pump parameters adjustments and carb counting issues are important factors for success in our outpatient model of care. In few subjects with SAP, telehealth monitoring was not possible due to technical reasons or was refused. This may explain in part the heterogeneity of the results in this group. The ability for caregivers to view patient data without manual uploads is still an unmet challenge due to the absence of automatic flow of data for all devices. The pace of development in diabetes technology is extremely rapid. It is mandatory that data could be automatically shared with secured cloud-based data management systems. During this process, the most important component remains the patient. It is therefore crucial to maintained dedicated teams with specific skills to assist the patient in device/ program selection and to support its use through ongoing education and training with both presential and virtual diabetes care.

In conclusions, we provide evidence for the safe combination of outpatient initiation of insulin pump with adapted education programs and telehealth monitoring performed by a dedicated healthcare team. This organization is perfectly adapted for the initiation of closed loop systems.