Abstract
Objectives
The aim of the patient out-of-plane shield is to reduce the patient radiation dose. Its effect on tube current modulation was evaluated with the out-of-plane shield visible in the localizer but absent in the scan range in chest CT with different CT scanners.
Methods
An anthropomorphic phantom was scanned with six different CT scanners from three different vendors. The chest was first scanned without any shielding, and then with the out-of-plane shield within the localizer but outside the imaged volume. All pitch values of each scanner were used. The tube current values with and without the out-of-plane shield were collected and used to evaluate the effect of overscanning and tube current modulation (TCM) on patient radiation dose.
Results
The highest increase in cumulative mA was 217%, when the pitch was 1.531. The tube current value increased already 8.9 cm before the end of the scanned anatomy and the difference between the tube current of the last slices (with and without the out-of-plane shield in the localizer) was 976%.
Conclusion
Applying an out-of-plane shield outside the scanned volume but visible in the localizer images may increase the patient dose considerably if the scanner’s TCM function is based only on localizer images.
Clinical relevance statement
The use of an out-of-plane shield in CT may strongly increase the tube current modulation and thus provide the patient with a higher radiation dose.
Key Points
• Applying an out-of-plane shield outside the scanned volume but visible in the localizer images may increase patient radiation dose considerably.
• The effect is visible with scanners that use solely localizer-based tube current modulation.
• Features like overscanning may be difficult for the user to notice when planning the scanning, and yet they may affect tube current modulation and through it to patient dose.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
The number of examinations and the amount of cumulative patient radiation dose from computed tomography studies (CT) is increasing in Organisation for Economic Co-operation and Development (OECD) countries, due to their versatility in assessing various pathologies. CT exams provided about 50% of the collective effective dose from medical imaging in Europe and over 60% in the USA during the past decade [1,2,3]. Since CT examinations cause a considerable amount of medical radiation exposure, it is important to find methods to optimize the application.
The routine use of patient in- or out-of-plane shielding during x-ray-based imaging has received a lot of focus in recent years and continues to be a source of intensive debate. Concerns related to the use of contact shielding have led to a variety of publications with e.g., International Commission on Radiological Protection (ICRP) and International Atomic Energy Agency (IAEA) endorsing its use [4, 5], while others like the European as well as the American Association of Physicist in Medicine (AAPM) guidelines discourage its use [6, 7]. This mismatch between guidelines can also be seen in practice, because according to the survey by Johnston et al, there are differences in the use of patient contact shielding in clinical CT practice [8]. Patient contact shield, as defined by the British Institute of Radiology (BIR), means radiation protection applied directly to a patient undergoing diagnostic X-ray procedures, but it excludes shielding built into the imaging equipment or in the room design [9]. There are two types of contact shielding: bismuth shields placed on the patient and inside the imaged volume (in-plane shields), and lead shields wrapped around the patient, outside the imaged volume (out-of-plane shields). In this study, we focused on out-of-plane shields.
Helical scanning has a requirement to overscan the imaged volume to provide enough data to reconstruct the first and last image. The amount of overscanning can extend a considerable non-intuitive distance beyond the image volume as the overscanning in multidetector helical CT examinations can be from half to even two x-ray tube rotations [10,11,12,13,14,15]. The amount of overscanning depends on the manufacturer and scanner. In some scanners, there is a mechanical method to reduce the effect of overscanning (so-called adaptive or dynamic collimation). There is very little information of the exact amount of overscanning in the literature [10]. The length of overscanning is not visible before or during the scanning, thus making it impossible to take into account when placing an out-of-plane shield. The placement of an out-of-plane shield beyond the imaged volume is therefore not a simple and error-free task.
Tube current modulation (TCM) is a main technique to optimize the patient’s radiation dose in CT while keeping the image quality sufficient. Scanner modulates the tube current output in the x-, y-, and z-directions to maintain the predetermined image noise level over the complete image volume. TCM function uses one or two localizer images to measure the x-ray attenuation due to the patient anatomy (the localizer-based method, used by GE Healthcare and Canon) [10,11,12,13,14, 16]. Another method uses the same technique but it also measures the attenuation during scanning, updating the information from the previous 180° rotation and making corrections to the tube current accordingly (“near real-time modulation”). Siemens Healthineers and Philips Healthcare [16] use the latter method. Our hypothesis was that tube current increase when approaching the diaphgram is stronger using an out-of-plane shield than without.
Materials and methods
An anthropomorphic phantom and clinical helical thorax protocols were used to estimate the effect of patient contact shielding on TCM, with an out-of-plane shield (Scanflex Medical AB, 0.5 mm lead) placed to protect the abdomen and pelvis (Fig. 1). The phantom (Kyoto Kagaku, model PBU 60, Kyoto Kagaku Co Ltd) was scanned with six different scanners from three different vendors. The phantom corresponds to the average size of a small adult, 50 kg and 165 cm, and is made of tissue equivalent plastic. The CT scanners used were Optima 660 and Revolution (GE Healthcare); Aquilion One, Aquilion Prime, and Aquilion Prime SP (Canon Medical Systems Co, formerly known as Toshiba Medical Systems Co); and Siemens Somatom go.Up (Siemens Healthineers) (Table 1).
For every scanner, the available tube current range to be used in TCM was extended from the clinical protocol settings to begin from 10 mA and reach the maximum value allowed by the scanner, to better visualize the effect of shielding. Scans were repeated twice with all pitch values of each scanner (Table 1): first without any shielding wrapped around the phantom, and then with the contact shield visible in the localizer images but not included in the scanned volume. The beam collimation used in clinical protocols varied 2.24–8 cm. It was constant per scanner.
The phantom was scanned head first with standard clinical chest protocol from T1 to T12 vertebra, total length of scanned volume being 32 cm. The scan range was same for all scans. The shield was placed 1 cm outside the end of the scanned volume and wrapped around the phantom, beginning at the top edge of the 2nd lumbar vertebra and covering the abdomen and pelvis (Fig. 1).
An in-house developed MatLab program (The MathWorks, Inc.) was used to collect the average mA value of each slice to identify the location of the possible mA increase due to the out-of-plane shield visible in the localizer images. The cumulative tube current value of each scan was also calculated to determine the overall effect of the shielding.
When comparing the scans, the maximum limit of 20% mA difference between representative individual slices was set as acceptable. The value is selected according to AAPM report concerning the acceptability and reliability of computed tomography exposure reproducibility, stating that the maximum approvable variation from the mean value of measurements taken is 20% [17].
Results
For most scanners, we observed an increase in cumulative tube current when we used an out-of-plane shielding. In most cases, the effect was notable in scanners using only localizer-based TCM (GE and Canon scanners), but could not be seen in Siemens Go.Up, and was minimal with Canon Prime SP scanner when the pitch 1.388 was used. Visible shielding in localizer images appears to affect the mA modulation function of a scan, as does the pitch, also (Fig. 2).
The earliest tube current increase was detected in Canon Prime, beginning 9.7 cm before the end of the scanned volume with 1.388 pitch (Fig. 2b, Fig. 3). In that incidence, the tube current difference of the last slices of the volume was 323%, and in the cumulative mA a difference of 53% (Table 2). The most notable difference with the cumulative mA value was, however, with another scanner. Anatomically, the second longest difference in single slices mA values was with GE Optima, being 8.9 cm with 1.531 pitch, and producing a 217% increase in cumulative tube current used in imagining the volume (Fig. 2a). This began 0.8 cm later in z-direction, but was remarkably steeper, as scanner imaged against its upper limit of TCM (Table 2). The tube current difference of the last slices was 976%.
Discussion
We observed that using out-of-plane shielding (placing the contact shield within the localizer images but outside the scanned volume) is likely to increase the tube current in helical scanning with scanners using only localizer-based TCM method (in this study GE Healthcare and Canon Medical Systems Co). The effect seems to relate to pitch values since with larger pitch values used the image reconstruction algorithm requires more information from the area outside the primary interest anatomical volume, and the TCM algorithm tends to increase the tube current early to maintain sufficient image quality. With a scanner that applies a near real-time TCM (in this study Siemens Healthineers), the tube current increase was not detected, even if the out-of-plane shield was used and visible in the localizer image.
In similar research setting as ours, Begano et al observed that shielding increased the effective dose to mother as well as to the fetus, even with a scanner that corrects the TCM values based on the localizer images with the attenuation measurement of the previous half round of the x-ray tube [18]. However, in their research, the increase was due to the scattered radiation. Overscanning is a relevant factor when optimizing patient dose in CT. It differs prominently between vendors. With some scanners, the manufacturer-specific details of overscanning are difficult to find. The required overscanning can be even two rotations [10,11,12,13,14,15]. Thus, depending on beam collimation and pitch, the amount of overscanning can extend a considerable non-intuitive distance beyond the examination volume in the z-direction, amounting to or in some cases even exceeding 16 cm. We were not able to find from the literature the amount of overscanning for the Siemens scanner used in this study. When contacting the manufacturer, we received a reply stating that this information is not available. The latest found reference for the Siemens scanners is van der Molen and Geleijins’ study [15]. However, our results do give an impression that the amount of overscanning they have measured is not applicable in Siemens Somatom Go.Up scanner used in our study. With dose measurements conducted with a fixed tube current, scattered radiation follows an exponential relationship as a function of the distance from the edge of the scanned area, as also does the dose saved by patient contact shielding [19]. As Yu et al present, the dose saved with a use of a margin may be extremely small compared with the overall dose from the examination [20]. For scanners that only use a localizer-based TCM, our study indicates that it might be necessary to place the contact shield out of the overscanned area, which may extend even as far as a distance twice the width of the used beam collimation multiplied by the pitch from the primary scan plane edge. The dose reduction, if gained at all from the use of patient contact shield, may not outweigh the potential risk of artifacts. The TCM is only available in helical scanning. One might wish to use axial imaging to be able to apply an out-of-plane shield in some examinations. As Weber et al state, the use of axial acquisition allows the precise positioning of the out-of-plane shield. However, in clinical setting, the benefit of helical scanning is faster scan time, and using axial acquisition may result in motion artifacts [19].
In Fig. 2a, the tube current values are very similar between pitches 0.984 and 1.375. We were not able to identify the reason for that.
The difference in results between two Canon scanners (Prime and Prime SP) is notable. Canon Prime is an older model, and one can clearly see in the results the combined effect of the out-of-field shield, tube current modulation, pitch, and overscanning. With the newer model Prime SP, the tube current difference with and without shield is more obscure. It may be that the image reconstruction algorithm is of a new version, and better observes and utilizes the information of the active collimation. The slight difference in Prime SP results with pitches 0.637 and 0.813, and with pitch 1.388, may be due to the x-ray tube placement in respect to couch.
If adaptive or dynamic shielding is not implemented in the scanner, an out-of-plane shielding could reduce radiation exposure due to overscanning. Unfortunately, the TCM algorithm uses data from the localizer outside the imaged volume that increases the mA although this may not be necessary for constant image quality. The impact of this mA increase on a constant image quality was not examined in this study. The clinical protocols use a lower maximum mA level for TCM. Therefore, the overscanning effect on TCM might in clinical scanning be lower as in some cases the upper mA limit for TCM was reached (see Fig. 2a, pitch 1.531).
There are limitations in this study. We could have measured the dose more exactly with thermo- or radiophotoluminescent dosimeters, instead of using the tube current value indicated by the scanner. However, in this context, tube current is directly proportional to the dose, and thus applicable when evaluating the effect of an out-of-plane shield to patient dose. Using such dosimeters would also have enabled us to measure the effect of adaptive or dynamic collimation, which we now were not able to do. There may be some fluctuation with the placement of the contact shield due to the anatomical shape of the phantom, but the effect on the results is assumed to be small. The beam collimation used in clinical protocols varied 2.24–8 cm. Since we compared the scanner properties for each scanner, not between scanners, we considered it not essential. The differences in the TCM due to the x-ray tube placement in respect to couch, in the beginning of the scan, would have been less meaningful if we had repeated the scans multiple times.
Some user manuals provided the scanner-specific amount to overscan, but for few scanners the information was not found (Table 1). Since we did not measure the exact dimension of overscanning, it has only been reported here as an additional information. However, the results we observed are unambiguous. While the results of this study are limited to phantom scans using chest protocols, they are likely applicable to scans of other anatomy, too.
In the light of our results, we recommend not to apply an out-of-plane shield for helical CT thorax scans if using a scanner that applies only the localizer-based TCM.
Abbreviations
- AAPM:
-
American Association of Physicist in Medicine
- BIR:
-
British Institute of Radiology
- IAEA:
-
International Atomic Energy Agency
- ICRP:
-
International Commission on Radiological Protection
- OECD:
-
Organisation for Economic Co-operation and Development
- TCM:
-
Tube current modulation
References
European Commission (2015) Radiation Protection No 180. Medical radiation exposure of the European population, Part 1/2. Publications Office of the European Union, Luxembourg. https://doi.org/10.2833/708119
United Nations (2011) Summary of low-dose radiation effects on health. United Nations Scientific Committee on the Effects of Atomic Radiation, sources and effects of ionizing radiation, Vienna. Available via https://www.unscear.org/unscear/uploads/documents/unscear-reports/UNSCEAR_2010_Report.pdf. Accessed 11 Dec 2021
Mettler FA Jr, Mahesh M, Bhargavan-Chatfield M et al (2020) Patient exposure from radiologic and nuclear medicine procedures in the United States: procedure volume and effective dose for the period 2006–2016. Radiology 295:418–427
International Commission on Radiation Protection (2013) ICRP Publication 121: radiological protection in paediatric diagnostic and interventional radiology. Ann. ICRP 42(2)
International Atomic Energy Agency (2018) Radiation protection and safety in medical uses of ionizing radiation. International Atomic Energy Agency, Vienna. Available via https://www-pub.iaea.org/MTCD/Publications/PDF/PUB1775_web.pdf. Accessed 8 Jan 2022
Hiles P, Gilligan P, Damilakis J et al (2022) European consensus on patient contact shielding. Phys Med 96:198–203
American Association of Physicist in Medicine (2019) AAPM position statement on the use of patient gonadal and fetal shielding PS8-A. Available via https://www.aapm.org/org/policies/details.asp?type=PP&id=2552. Accessed 8 Mar 2023
Johnston J, Killion JB, Vealé B, Comello R (2011) U.S. technologists’ radiation exposure perceptions and practices. Radiol Technol 82:311–320
British Institute of Radiology (2020) Guidance on using shielding on patients for diagnostic radiology applications. London. Available via https://www.bir.org.uk/media/414334/final_patient_shielding_guidance.pdf. Accessed 15 Dec 2021
Rousselle I, Farah J, Dufay F et al (2019) Over-ranging in CT scanners: a comparison of equipment models and medical practices in France. EuroSafe Imaging, poster ESI-0078. https://doi.org/10.26044/esi2019/ESI-0078
Technical Revolution CT ES, Manual Reference (2017) Direction 5763055–1EN, Revision 2. General Electric Company, Waukesha
Operation manual Toshiba Scanner Aquilion One (2011), TSX-301A, Basic volume (2B201–415FI*V
Operation manual Toshiba Scanner Aquilion Prime (2012), TSX-302A, Basic volume (2B201–475EN*J)
Operation manual for Toshiba Scanner Aquilion Prime (2016), TSX-303A, Basic volume (2B201–718EN*F)
van der Molen AJ, Geleijins J (2007) Overranging in multidetector CT: quantification and relative contribution to dose - comparison of four 16-section CT scanners. Radiology 242:208–216
McCollough CH, Bruesewitz MR, Kofler JM Jr (2006) CT dose reduction and dose management tools: overview of available options. Radiographics 26:503–512
American Association of Physicist in Medicine (2019) AAPM Report 233 performance evaluation of computed tomography systems. Available via https://issuu.com/aapmdocs/docs/tg-233_final_8ec461f2715a5e?mode=embed&viewMode=singlePage&backgroundColor=eeeeee. Accessed 6 Mar 2023
Begano D, Söderberg M, Bolejko A (2020) To use or not to use patient shielding on pregnant women undergoing CT pulmonary angiography: a phantom study. Radiat Prot Dosimetry 189:458–465
Weber N, Monnin P, Elandoy C, Ding S (2015) A model-based approach of scatter dose contributions and efficiency of apron shielding for radiation protection in CT. Phys Med 31:889–896
Yu L, Bruesewitz MR, Vrieze TJ, McCollough CH (2019) Lead shielding in pediatric chest CT: effect of apron placement outside the scan volume on radiation dose reduction. AJR Am J Roentgenol 212:151–156
Acknowledgements
The authors wish to thank Adjunct Professor Jani Saunavaara for his overall support, and MSc Tero Patana for the assistance with the figures.
Funding
Open Access funding provided by University of Turku (UTU) including Turku University Central Hospital. This study was financially supported by the Radiology Society of Finland, and the State Research Funding of the Turku University Hospital.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Guarantor
The scientific guarantor of this publication is Jani Saunavaara, Adjunct Professor, Department of Clinical Medicine, Institute of Biomedicine in University of Turku.
Conflict of interest
The authors of this manuscript declare no relationships with any companies, whose products or services may be related to the subject matter of the article.
Statistics and biometry
No complex statistical methods were necessary for this paper.
Informed consent
Written informed consent was not required for this study because this is a phantom study.
Ethical approval
Institutional Review Board approval was not required because this is a phantom study.
Study subjects or cohorts overlap
No study subjects or cohorts have been previously reported.
Methodology
• Experimental
• Performed at one institution
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Larjava, H.R.S., Eneh, C.T.M. & Niiniviita, H.M. To shield or not to shield: shielding may have unintended effects on patient dose in CT. Eur Radiol 34, 2480–2486 (2024). https://doi.org/10.1007/s00330-023-10211-3
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00330-023-10211-3