Abstract
The aim of the study was to test the reliability of intracerebral haemorrhage (ICH) detection with C-arm-mounted flat-detector computed tomography (FD-CT) in the angio suite as compared to multislice CT (MSCT). In this study 44 patients with 45 ICH were included. All patients were investigated with MSCT and FD-CT during angiographic evaluation. As a control group we included 16 patients without ICH. In each haematoma we assessed volumetric data of the ICH and counted the numbers of ICH-positive slices. Using interobserver ratings, we additionally investigated the potential of FD-CT to serve as a diagnostic tool to detect ICH. In FD-CT three haematomas were not detected because of motion and beam-hardening artefacts in the region close to the skull base. The r value for the degree of interobserver agreement for the number of slices was 0.95 for MSCT and 0.94 for FD-CT. Measurements of the area and the calculated volume of the ICH showed high inter- and intraobserver agreement. Our results indicate that FD-CT is a helpful tool in the daily emergency management of ICH patients as detection of ICH was found to be nearly as reliable as in MSCT. Limitations of this technology are motion and beam-hardening artefacts that may mask small haematomas located in the posterior fossa or the skull base.
Similar content being viewed by others
References
Sahni R, Weinberger J (2007) Management of intracerebral hemorrhage. Vasc Health Risk Manag 3(5):701–709
Hänggi D, Steiger HJ (2008) Spontaneous intracerebral haemorrhage in adults: a literature overview. Acta Neurochir 150:371–379
Brisman JL, Niimi Y, Song JK, Berenstein A (2005) Aneurysmal rupture during coiling: low incidence and good outcomes at a single large volume centre. Neurosurgery 57:1103–1109
Carvi y Nievas M, Haas E, Höllerhage HG (2007) Severe intracranial bleedings during endovascular procedures: outcome of surgically treated patients. Neurol Res 29(1):81–90
Kalender WA, Kyriakou Y (2007) Flat-detector computed tomography (FD-CT). Eur Radiol 17:2767–2779
Schumacher M, Kutluk K, Ott D (1989) Digital rotational radiography in neuroradiology. Am J Neuroradiol 10:644–649
Kalender WA (2003) [The use of flat-panel detectors for CT imaging]. Radiologe 43:379–387
Loose R, Wucherer M, Brunner T, Adamus R, Simmler R (2005) Visualization of 3D low contrast objects by CT cone-beam reconstruction of a rotational angiography with a dynamic solid body detector. Roefo S1:PO 160
Heran NS, Song JK, Namba K, Smith W, Niimi Y, Berenstein A (2006) The utility of DynaCT in neuroendovascular procedures. Am J Neuroradiol 27:330–332
Arakawa H, Marks MP, Do HM, Bouley DM, Strobel N, Moore T, Fahrig R (2008) Experimental study of intracranial hematoma detection with flat panel detector C-arm CT. Am J Neuroradiol 29(4):766–772
Kwak R, Kadoya S, Suzuke T (1983) Factors affecting the prognosis of thalamic hemorrhage. Stroke 14:493–500
Richter G, Engelhorn T, Struffert T, Doelken M, Ganslandt O, Hornegger J, Kalender WA, Doerfler A (2007) Flat panel detector angiographic CT for stent-assisted coil embolization of broad-based cerebral aneurysms. Am J Neuroradiol 28(19):1902–1908
Engelhorn T, Rennert J, Richter G, Struffert T, Ganslandt O, Doerfler A (2007) Myelography using flat panel volumetric computed tomography: a comparative study in patients with lumbar spinal stenosis. Spine 32(18):E523–E527
Doelken M, Struffert T, Richter G, Engelhorn T, Nimsky C, Ganslandt O, Hammen T, Doerfler A (2008) Flat-panel detector volumetric CT for visualization of subarachnoid hemorrhage and ventricles: preliminary results compared to conventional CT. Neuroradiology 50(6):517–523
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Struffert, T., Richter, G., Engelhorn, T. et al. Visualisation of intracerebral haemorrhage with flat-detector CT compared to multislice CT: results in 44 cases. Eur Radiol 19, 619–625 (2009). https://doi.org/10.1007/s00330-008-1183-7