Abstract.
Our objective was to compare a water-excitation (WE) 3D fast low-angle shot (FLASH) MR sequence for faster imaging of articular cartilage defects of the knee to a conventional fat-saturated (FS) 3D FLASH MR sequence. This prospective study included 16 knees of 16 patients with suspected cartilage lesions. The MR imaging in transverse and sagittal planes included (a) FS 3D FLASH (TR/TE: 45 ms/11 ms, scan time 8 min, flip angle 50°), and (b) WE 3D FLASH (TR/TE: 28 ms/11 ms, scan time 4 min 58 s, flip angle 40°). For each sequence signal-to-noise ratios (SNR) and contrast-to-noise ratios (CNR) were quantified. The detected cartilage lesions were evaluated using a semi-quantitative four-scale scoring system (grades 0–III). The data were compared between the sequences using the paired Student's t-test. No statistically significant differences between the sequences were found for SNR, CNR, and cartilage defect grading (p=0.14–0.8). The WE 3D FLASH MR imaging seems to be promising for fast imaging of articular cartilage lesions of the knee.