Chest imaging with flat-panel detector at low and standard doses: comparison with storage phosphor technology in normal patients

  • Chest
  • Published:
European Radiology Aims and scope Submit manuscript

Abstract.

The purpose of this study was to evaluate the performance of standard-dose and low-dose cesium iodide (CsI)-doted amorphous silicon (a-Si) flat-panel detector technology (FDT) as compared with storage-phosphor technology (SPT) in the depiction of relevant anatomical structures in chest radiography. In 75 patients referred for thoracic CT, digital chest radiographs were randomly obtained with either SPT at a standard dose (speed class S400, n=25), standard-dose FDT (S400, n=25) or FDT at a low dose (S800, n=25). Five radiologists evaluated the visibility of eight pulmonary and mediastinal anatomical structures using a five-point rating scale. To determine statistically significant differences between the three groups, the Mann-Whitney U-test was employed. No statistically significant differences were found in the depiction of eight criteria between SPT and standard-dose or low-dose FDT chest radiographs. The performance of FDT S400 was equal to SPT for most criteria and better for retrocardiac structures and soft tissue. FDT S800 was inferior to both SPT and FDT S400. Standard-dose FDT is equivalent to SPT in the depiction of relevant anatomical structures of the chest. Our results also indicate that a dose reduction of 50% with FDT may result in small but not significant decrease of image quality.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Electronic Publication

Rights and permissions

Reprints and permissions

About this article

Cite this article

Herrmann, K., Bonél, H., Stäbler, A. et al. Chest imaging with flat-panel detector at low and standard doses: comparison with storage phosphor technology in normal patients. Eur Radiol 12, 385–390 (2002). https://doi.org/10.1007/s00330-001-1166-4

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00330-001-1166-4

Navigation