We have developed an angiographic guidewire with measuring markers to determine accurately how far a guidewire is inserted within a catheter. We investigated whether use of this guidewire reduces the risk of vascular injury and the fluoroscopic time during guidewire manipulations.
Methods
Four markers were put on the surface of the guidewire at 80, 100, 110, and 120 cm from the tip. The actual lengths of 54 catheters from seven manufacturers were measured and compared with the nominal lengths. Sixty consecutive patients who underwent angiography were randomized into two groups: in one group guidewires with surface markers were used (marker group) and in the other group, conventional guidewires (control group). For each guidewire insertion, the fluoroscopic time before the guidewire was pushed forward into the vessel lumen was recorded. The number of occasions on which unintentionally the guidewire had already been pushed out of the catheter at the start of fluoroscopy was also evaluated.
Results
The actual lengths of all catheters were greater than the nominal lengths by 1.0–11.0 cm. Mean fluoroscopic time for each guidewire insertion was 3.3 sec in the marker group and 5.7 sec in the control group (p < 0.05). Guidewires were unintentionally pushed out of the catheters without fluoroscopy three times (3.6%), in each case in the control group.
Conclusion
The guidewire with measuring markers is effective for enhancing safety and in reducing fluoroscopic radiation during angiographic procedures. It is recommended that operators be aware that actual lengths of catheters may vary significantly from the nominal lengths listed; they should be aware of this with any guidewire, but particularly with the angiographic measuring guidewire.
Similar content being viewed by others
References
Hessel SJ, Adams DF, Abrams HL (1981) Complications of angiography. Radiology 138:273–281
Chung JW, Park JH, Han JK, et al. (1994) Perforation of the gastroduodenal artery induced by steerable guidewires in two cases: Treatment of hemorrhage by embolization. Cardiovasc Intervent Radiol 17:41–43
Cope C, Baum S (1997) Catheter, methods, and injections for superselective catheterization. Vascular and interventional radiology. In: Abram’s angiography, 4th edn. Little, Brown, Boston, pp 155–175
Wagner LK, Eifel PJ, Geise RA (1994) Potential biological effects following high-x-ray dose interventional procedures. J Vasc Interv Radiol 5:71–84
Koenig TR, Wolff D, Mettler FA, et al. (2001) Skin injuries from fluoroscopically guided procedures. 1. Characteristics of radiation injury. AJR Am J Roentgenol 177:3–11
Koenig TR, Mettler FA, Wagner LK (2001) Skin injuries from fluoroscopically guided procedures. 2. Review of 73 cases and recommendations for minimizing dose delivered to patient. AJR Am J Roentgenol 177:13–20
Valentin J (ed) (2000) Avoidance of radiation injuries from medical interventional procedures. Ann ICRP 30(2):7–67
Haskal ZJ (2004) Cataracts in interventional radiology: an occupational hazard? Abstracts of the annual meeting of the Society for Interventional Radiology
Brenner DJ, Doll R, Goodhead DT, et al. (2003) Cancer risks attributable to low doses of ionizing radiation: Assessing what we really know. Proc Natl Acad Sci USA 100:13761–13766
Gonzalez AB, Darby S (2004) Risk of cancer from diagnostic X-rays: Estimates for the UK and 14 other countries. Lancet 363:345–351
Acknowledgment
The guidewires with measuring markers were manufactured for the trial by Terumo Corporation (Tokyo, Japan).
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Kamei, S., Ishiguchi, T., Murata, K. et al. Angiographic Guidewire with Measuring Markers: Design and Clinical Experience. Cardiovasc Intervent Radiol 29, 981–985 (2006). https://doi.org/10.1007/s00270-005-0294-7
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00270-005-0294-7