Angiographic Guidewire with Measuring Markers: Design and Clinical Experience

  • CLINICAL INVESTIGATION
  • Published:
CardioVascular and Interventional Radiology Aims and scope Submit manuscript

Abstract

Purpose

We have developed an angiographic guidewire with measuring markers to determine accurately how far a guidewire is inserted within a catheter. We investigated whether use of this guidewire reduces the risk of vascular injury and the fluoroscopic time during guidewire manipulations.

Methods

Four markers were put on the surface of the guidewire at 80, 100, 110, and 120 cm from the tip. The actual lengths of 54 catheters from seven manufacturers were measured and compared with the nominal lengths. Sixty consecutive patients who underwent angiography were randomized into two groups: in one group guidewires with surface markers were used (marker group) and in the other group, conventional guidewires (control group). For each guidewire insertion, the fluoroscopic time before the guidewire was pushed forward into the vessel lumen was recorded. The number of occasions on which unintentionally the guidewire had already been pushed out of the catheter at the start of fluoroscopy was also evaluated.

Results

The actual lengths of all catheters were greater than the nominal lengths by 1.0–11.0 cm. Mean fluoroscopic time for each guidewire insertion was 3.3 sec in the marker group and 5.7 sec in the control group (p < 0.05). Guidewires were unintentionally pushed out of the catheters without fluoroscopy three times (3.6%), in each case in the control group.

Conclusion

The guidewire with measuring markers is effective for enhancing safety and in reducing fluoroscopic radiation during angiographic procedures. It is recommended that operators be aware that actual lengths of catheters may vary significantly from the nominal lengths listed; they should be aware of this with any guidewire, but particularly with the angiographic measuring guidewire.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Hessel SJ, Adams DF, Abrams HL (1981) Complications of angiography. Radiology 138:273–281

    PubMed  CAS  Google Scholar 

  2. Chung JW, Park JH, Han JK, et al. (1994) Perforation of the gastroduodenal artery induced by steerable guidewires in two cases: Treatment of hemorrhage by embolization. Cardiovasc Intervent Radiol 17:41–43

    Article  PubMed  CAS  Google Scholar 

  3. Cope C, Baum S (1997) Catheter, methods, and injections for superselective catheterization. Vascular and interventional radiology. In: Abram’s angiography, 4th edn. Little, Brown, Boston, pp 155–175

  4. Wagner LK, Eifel PJ, Geise RA (1994) Potential biological effects following high-x-ray dose interventional procedures. J Vasc Interv Radiol 5:71–84

    Article  PubMed  CAS  Google Scholar 

  5. Koenig TR, Wolff D, Mettler FA, et al. (2001) Skin injuries from fluoroscopically guided procedures. 1. Characteristics of radiation injury. AJR Am J Roentgenol 177:3–11

    PubMed  CAS  Google Scholar 

  6. Koenig TR, Mettler FA, Wagner LK (2001) Skin injuries from fluoroscopically guided procedures. 2. Review of 73 cases and recommendations for minimizing dose delivered to patient. AJR Am J Roentgenol 177:13–20

    PubMed  CAS  Google Scholar 

  7. Valentin J (ed) (2000) Avoidance of radiation injuries from medical interventional procedures. Ann ICRP 30(2):7–67

    PubMed  CAS  Google Scholar 

  8. Haskal ZJ (2004) Cataracts in interventional radiology: an occupational hazard? Abstracts of the annual meeting of the Society for Interventional Radiology

    Google Scholar 

  9. Brenner DJ, Doll R, Goodhead DT, et al. (2003) Cancer risks attributable to low doses of ionizing radiation: Assessing what we really know. Proc Natl Acad Sci USA 100:13761–13766

    Article  PubMed  CAS  Google Scholar 

  10. Gonzalez AB, Darby S (2004) Risk of cancer from diagnostic X-rays: Estimates for the UK and 14 other countries. Lancet 363:345–351

    Article  Google Scholar 

Download references

Acknowledgment

The guidewires with measuring markers were manufactured for the trial by Terumo Corporation (Tokyo, Japan).

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Seiji Kamei.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Kamei, S., Ishiguchi, T., Murata, K. et al. Angiographic Guidewire with Measuring Markers: Design and Clinical Experience. Cardiovasc Intervent Radiol 29, 981–985 (2006). https://doi.org/10.1007/s00270-005-0294-7

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00270-005-0294-7

Keywords

Navigation