Abstract
Gravimetrically prepared mono-elemental reference solutions having a well-known mass fraction of approximately 1 g/kg (or a mass concentration of 1 g/L) define the very basis of virtually all measurements in inorganic analysis. Serving as the starting materials of all standard/calibration solutions, they link virtually all measurements of inorganic analytes (regardless of the method applied) to the purity of the solid materials (high-purity metals or salts) they were prepared from. In case these solid materials are characterized comprehensively with respect to their purity, this link also establishes direct metrological traceability to The International System of Units (SI). This, in turn, ensures the comparability of all results on the highest level achievable. Several national metrology institutes (NMIs) and designated institutes (DIs) have been working for nearly two decades in close cooperation with commercial producers on making an increasing number of traceable reference solutions available. Besides the comprehensive characterization of the solid starting materials, dissolving them both loss-free and completely under strict gravimetric control is a challenging problem in the case of several elements like molybdenum and rhodium. Within the framework of the European Metrology Research Programme (EMRP), in the Joint Research Project (JRP) called SIB09 Primary standards for challenging elements, reference solutions of molybdenum and rhodium were prepared directly from the respective metals with a relative expanded uncertainty associated with the mass fraction of U rel(w) < 0.05 %. To achieve this, a microwave-assisted digestion procedure for Rh and a hotplate digestion procedure for Mo were developed along with highly accurate and precise inductively coupled plasma optical emission spectrometry (ICP OES) and multicollector inductively coupled plasma mass spectrometry (MC-ICP-MS) methods required to assist with the preparation and as dissemination tools.
Similar content being viewed by others
Introduction
The most elegant and conclusive way to achieve comparable measurement results or to render such results suitable for acceptance in law suits is to establish so-called metrological traceability [1]. The basic concept of metrological traceability is the establishment of an unbroken chain of calibrations, each contributing to the measurement uncertainty, to relate a measurement result to a reference, which at best is the realization of the SI unit [2, 3]. Large areas of analytical chemistry are concerned with the measurement of the contents of, e.g., precious, poisonous, or essential elements in a wide variety of sample matrices, with applications in environmental protection, health care, industrial production, etc., whenever limits or contents are legally and/or, often as a consequence, economically important or, e.g., when the health of human beings is at stake. In elemental analysis, worldwide attempts are underway to establish a calibration hierarchy for mono-elemental solutions [4–9] because these solutions with their well-known element contents are required for every kind of inorganic analysis, no matter what analytical method is applied. In the end, virtually all methods from AAS, to inductively coupled plasma optical emission spectrometry (ICP OES) and ICP MS, and to isotope dilution techniques rely on an elemental calibration solution. These usually commercially available solutions need to have a well-known content to yield comparable and accurate results. A prerequisite for the traceability of these solutions to the SI is the availability of so-called primary solutions prepared according to the highest metrological standards resulting in the lowest possible uncertainty associated with their element contents. Therefore, the primary solutions have to be prepared gravimetrically and loss-free directly, e.g., from pure metals with exactly known purity (including well-known metallic and non-metallic impurities) [8] and every weight measured on the way from the starting material to the final solution needs to be air buoyancy-corrected. These primary solutions are then used to link up the commercial solutions to ensure the comparability of “everyday” results determined in laboratories all over the world.
Rhodium and molybdenum are relatively precious and a lot of important technical applications rely on well-known contents. However, currently available primary solutions (like NIST SRM 3144 and 3134, respectively) are either prepared from salts or metal powders with an unknown and hard to clean surface (in the case of Rh [10]) or the matrices of the primary solution and the commercially available solutions do not match (in the case of Mo). Preparing a primary solution from a salt can result in a mass fraction that deviates from the theoretically calculated value, if the stoichiometry is not verified. Additionally, the relative expanded uncertainty of the mass fraction of those solutions is larger than the target uncertainty of 0.1 % for primary solutions. This uncertainty is vital for transfer standards, as a target uncertainty of 0.3 % for the calibration of commercial solutions should be obtained. Thus, primary solutions of molybdenum and rhodium were prepared in the framework of the European Metrology Research Programme (EMRP), within the Joint Research Project (JRP) called SIB09 Primary standards for challenging elements. To accomplish this, at the same time, contamination- and loss-free as well as quantitative procedures to bring these pure solid metals directly into solution had to be developed; in the case of rhodium, for the first time starting from the metal itself with a defined surface that was easy to clean. Additionally, the solutions match the matrix of the commercially available products, which eliminates matrix effects during the calibration of commercial solutions, resulting at least in a lower measurement uncertainty.
Both ICP OES and multicollector inductively coupled plasma mass spectrometry (MC-ICP-MS) are commonly recognized as achieving highly precise and accurate measurement results in elemental analysis, especially when sophisticated measurement regimes like bracketing combined with an internal standard (IS) are applied [11–14, 7, 15]. Therefore, these techniques were applied not only to develop methods for the linkup of commercial solutions to the primary solutions but also to validate the repeatability of the dissolution and storage procedures as well as to ensure the completeness of the dissolution and the absence of losses.
Experimental
Molybdenum solution
Molybdenum rods (3 mm diameter, 99.95 %, Alfa Aesar GmbH, Germany, lot F05F20) were cut into pieces of approximately 1 g (14 mm) each. The surface of the pieces was cleaned using an etching procedure, in accordance with the recipe for the determination of impurities in the solid metal [8]. Briefly, the molybdenum rods were placed into a mixture of 4 mL nitric acid (w(HNO3) = 0.65 g/g, sub-boiled) and 1 mL hydrofluoric acid (w(HF) = 0.40 g/g, ultrapur, Merck KGaA, Germany) for 10 s, then they were rinsed with water (σ < 0.066 μS/cm, Milli-Q Element A10, EMD Millipore Corporation, USA), etched with hydrochloric acid (w(HCl) = 0.32 g/g, ultrapur, Merck KGaA, Germany), again cleaned with water, and then treated with ethanol (p. A., Merck KGaA, Germany) in an ultrasonic bath. The last two steps were repeated three times. After weighing 2 g (air buoyancy-corrected mass) of molybdenum (2 pieces) directly into a pre-cleaned and blank-checked DURAN® bottle (Schott AG, Germany), 30 g of nitric acid (w(HNO3) = 0.30 g/g) were added to dissolve the sample within 6 h using a heating block on a hotplate (120 °C hotplate temperature). Afterwards, the solution was evaporated to dryness and the sample was allowed to cool down to room temperature. Then, 50 g of ammonium hydroxide (w(NH3) = 0.25 g/g, ultrapur, Merck KGaA, Germany) were added to the residue and heated again on the hotplate (parameters as before) for 3 h to ensure complete dissolution. The resulting clear solution was evaporated to dryness and water was added to the residue to adjust a molybdenum mass fraction of w(Mo) ≈ 4 g/kg. This stock solution was diluted gravimetrically to yield several 500 g aliquots of the desired w(Mo) = 1 g/kg solution.
Rhodium solution
Rhodium foil (thickness 0.025 mm, 99.8 %, Alfa Aesar GmbH, Germany, lot MM48317) was cut into pieces of approximately 100 mg each. The pieces were cleaned using an ultrasonic bath for 1 min each utilizing acetone (SupraSolv®, Merck KGaA, Germany), nitric acid (n(HNO3)/m = 0.15 mol/kg), water (σ < 0.066 μS/cm, Milli-Q Element A10, EMD Millipore Corporation, USA), and again acetone, respectively. After drying in a clean environment at room temperature, one piece each was weighed directly into four thoroughly pre-cleaned and blank-checked 35 mL quartz vessels (MLS GmbH, Germany). About 15 mL hydrochloric acid (w(HCl) = 0.32 g/g, ultrapur, Merck KGaA, Germany) and 3 mL hydrogen peroxide (w(H2O2) = 0.31 g/g, ultrapur, Merck KGaA, Germany) were then added to each vessel. After attaching PTFE caps featuring pressure release holes, the vessels were placed in a microwave autoclave (turboWAVE, MLS GmbH, Germany). At room temperature, a base pressure of 50 bar was applied to the autoclave using argon (5.0, Linde, Germany). Following this, the samples were heated, microwave-assisted, to 260 °C (≤105 bar) within 15 min. After holding this temperature for another 50 min, the samples were allowed to cool down to 35 °C within 30 min. Finally, the remaining pressure was released at a rate of 5 bar/min. The vessels were then placed on a hotplate and the rhodium solution was evaporated to near dryness within 6 h applying a hotplate surface temperature of 130 °C. One milliliter sub-boiled nitric acid (prepared from w(HNO3) = 0.65 g/g, EMSURE® ISO using a Cetac OmniPure, Teledyne Technologies Inc., USA) was added. After evaporation to near dryness, the residues were redissolved in sub-boiled nitric acid (w(HNO3) = 0.025 g/g) and homogenization was achieved by placing the vessels in an ultrasonic bath heated to 70 °C for about 4 h. Finally, the solutions were transferred into a thoroughly pre-cleaned and blank-checked 500-mL PFA bottle (AHF analysentechnik, Germany). The air buoyancy-corrected mass of the PFA bottle had been determined beforehand. A suitable amount of nitric acid (w(HNO3) = 0.025 g/g) was added to result in approximately 500 g of a solution with a rhodium mass fraction of w(Rh) ≈ 1.5 g/kg. Several of these stock solutions were prepared and diluted gravimetrically to yield a corresponding number of 500 g aliquots of the desired 1 g/kg solution.
Mass fractions of molybdenum and rhodium and their associated uncertainties
All mass fractions were calculated from the evaporation correction/sampling scheme applied to the solutions (Fig. 1). Provided a suitable bottle material and an appropriate solvent and, thus, a stable chemical form of the element was chosen in preliminary experiments, the mass fraction of the element in the solution is a function of time and changes due to evaporation losses during storage. To correct for these losses, not only the mass of the remaining solution has to be determined prior to taking an aliquot from the bottle but also the remaining mass immediately after the sampling (and prior to the next storage stage). First of all, the empty bottle was weighed on a balance (B) with a mass resolution of Δm = 0.01 mg (MC210S, Sartorius AG, Germany) and at the same time (1) on the balance (A) intended for the weighing of the final solution (Δm = 0.1 mg, H315, Mettler Instrumente AG, Switzerland). Additionally, the ambient conditions (temperature ϑ 1, relative humidity φ 1, and pressure p 1 of the air) were determined using a corresponding gauge (testo 650 with sensors 0638.1645 and 0636.9741, Testo GmbH & Co, Germany) to correct for the air buoyancy. This was accomplished by applying an air buoyancy correction factor K to all readings mʹ from the balances to calculate the actual mass m:
The air buoyancy correction factor of a certain object weighed on a balance is a function of the object’s density ρ i , the density of the traceable weights used to calibrate the balance ρ cal, and of the air density ρ air, which, in turn, is a function of the ambient conditions [16–18]:
Next, the appropriate amount of Mo was placed in the bottle and the latter was weighed again on balance B at a time (2). This yielded the actual mass m z,2 of Mo (since pure metallic Mo with a purity of w pur was used, the stoichiometric factor, f M , becomes 1). For more details, refer to Fig. 1:
Subsequently, the Mo was dissolved (see above) and the solution was adjusted to yield the intended Mo mass fraction by adding the appropriate amount of solvent. Before withdrawing a sub-sample from the bottle at a time (3), the bottle was weighed again on balance A to be able to calculate the recent mass fraction w z,3 of Mo corrected for evaporation losses:
To enable the calculation of future mass fractions after the next storage time(s), the bottle was also weighed after the first withdrawal to determine the amount of Mo (solution) remaining in the bottle:
These alternating storage periods and withdrawals can be described in general:
i th | weighing with | i | ∈ | N *\{1,2} | = | {3,4,5,…} |
k th | withdrawal with | k | ∈ | N * | = | {1,2,3,…} |
and w z,i with | i | = | 2 × k + 1 | = | {3,5,7,…} |
The Rh mass fraction was calculated accordingly. In this work, several primary Mo and Rh solutions with SI-traceable mass fractions of w(E) ≈ 1 g/kg (approximately 1 g/L) and an associated relative expanded uncertainty of U rel(w) ≤ 0.05 % were prepared. The uncertainties according to the Guide to the Expression of Uncertainty in Measurement (GUM) [19] were calculated using the software GUM Workbench Pro 2.4.1.392 (Metrodata GmbH, Germany) considering [20].
Measurement method
To be able to compare two Mo or Rh solutions with each other (e.g., to link a secondary or transfer standard to the primary solution) on the highest metrological level (best precision and accuracy), a standard bracketing procedure with an IS was developed and implemented successfully using an ICP OES as well as an MC-ICP-MS.
Two blends (1) and (2) of a well-known standard solution (z) of the analyte element having an analyte mass fraction w z with the internal standard solution (y) and one blend (bx) of an unknown sample solution (x) with the same internal standard solution (y) were prepared (Fig. 2). Suitable masses m of the solutions were mixed to yield an intensity ratio close to unity in the sample blend (bx) (in order to minimize biases due to detector dead time, CCD readout rates, etc.) and intensity ratios in the standard blends (1) and (2) of ±5 % around this unity ratio (bracket). The mass fraction w x of the analyte in the sample was calculated according to Eq. (10). This equation is derived in the Appendix. R denotes the intensity ratio of the internal standard Y over the analyte element E.
Since it was not the intention to determine absolute analyte mass fractions but to determine how accurate and precise two (primary) solutions can be compared to each other, all three blends (bx), (1), and (2), were prepared from the same primary solution. The resulting mass fraction was then compared to the one calculated from the gravimetric preparation. In a second step, one primary solution was used to prepare the blends (1) and (2), while the sample blend (bx) was prepared from another primary solution. This way the uncertainty associated with the comparison of two (primary) solutions was estimated.
Bracketing measurements with inductively coupled plasma optical emission spectrometry
All measurements were performed using a commercial Spectro Ciros CCD (SPECTRO Analytical Instruments GmbH, Germany). Table 1 summarizes the main instrumental settings. Yttrium served as the internal standard in the case of both the Mo and the Rh measurements.
Within a 3-h sequence the sample blend (bx) and the blends (1) and (2) were measured in turn, starting and ending with blend (1), see Fig. 3. Twenty measurements of blend (bx) were bracketed with 11 measurements of blend (1) and ten of blend (2). To compare different solutions, one primary solution was used to prepare blends (1) and (2), while blend (bx) was prepared from a second primary solution. The relative expanded uncertainty U rel according to [19] associated with the comparison of the two primary solutions ranged between 0.1 % (Mo) and 0.2 % (Rh). The uncertainty calculations were carried out with the software GUM Workbench Pro.
Bracketing measurements with multi-collector inductively coupled plasma mass spectrometry
All measurements were performed using a commercial ThermoFisher Scientific Neptune double-focusing MC-ICP-MS equipped with 9 Faraday detectors coupled to 1011 Ω resistors. The mass spectrometer is equipped with an additional air laminar flow filter system (Spetec Laminar Flow Box FBS-V 75, class H14, efficiency ≥99.995 %) for the sample introduction area. It is located in a clean and air-conditioned laboratory (Δϑ = ±1 °C) accessible through an air lock. Table 2 summarizes the main instrumental settings.
Standard nickel cones and 20-mL PFA sample vials (both AHF analysentechnik, Germany) were used. Blanks were determined after an appropriate wash time and before every single measurement; they were subtracted from the subsequent measurement data accordingly.
Indium (In) served as the internal standard for the rhodium measurements. Therefore, the 103Rh and 115In beams were focused into two suitable Faraday cups (called L3 and H3). The measurements were performed without any changes of the magnetic and electrostatic field or acceleration voltage (so-called static mode).
In the case of molybdenum, yttrium was used as the internal standard. The 89Y and 98Mo beams were focused into another 2 Faraday cups (called L4 and H4). Again, the measurements were performed in static mode.
Using Eq. (10), relative expanded measurement uncertainties for the comparison of two Mo and two Rh solutions of U rel ≤ 0.1 %, respectively, were obtained.
Results and discussion
Three primary Mo stock solutions with w(Mo) ≈ 4 g/kg and three primary Rh stock solutions with w(Rh) ≈ 1.5 g/kg as well as several primary solutions with w(E) ≈ 1 g/kg were prepared according to the highest metrological standards to yield relative expanded measurement uncertainties associated with gravimetrically determined mass fractions w of U rel(w) ≤ 0.05 %. These uncertainties include the purities of the solid starting materials as well as all contributions due to preparation and storage. The purity of the starting materials is by far the largest contribution to the uncertainties, with a relative contribution close to 99 % and, hence, also the limiting contribution to the uncertainties associated with the mass fractions of the primary solutions (Fig. 4a). The weighing of the solid materials increases the total uncertainties only slightly (less than 1 %). The uncertainties associated with handling, storage, and air buoyancy can be neglected in practice, at least at the very beginning of the lifetime of such a solution. After a certain storage period, these uncertainties become more and more important.
The bracketing plus internal standard-based measurement procedures intended to be used to compare the primary solutions with each other and to relate the primary solutions to transfer or commercial solutions in order to establish metrological traceability of the commercial solution showed fit-for-purpose measurement uncertainties of U rel ≤ 0.2 % (ICP OES) and U rel ≤ 0.1 % (MC-ICP-MS). A comparison between the measurement results obtained by ICP OES and MC-ICP-MS shows that both measuring techniques are suitable to recover the mass fraction calculated from the gravimetric preparation (Fig. 5): for example, all results for the Mo solution (internally called Mo-1-15.1) are consistent within the limits of uncertainty. Naturally, the uncertainty associated with the Mo mass fraction calculated from the preparation (U rel = 0.05 %) is the limiting uncertainty because this mass fraction is an input quantity of the measured Mo mass fractions. While the MC-ICP-MS measurement adds only a small amount to this uncertainty, resulting in an uncertainty associated with the measured Mo mass fraction of U rel = 0.061 %, the ICP OES measurement adds a little more, so that uncertainties of U rel = 0.10 % were obtained. When looking at the different distributions of the main contributions to the uncertainties associated with measured Mo mass fractions (Fig. 4b, c), it becomes obvious that in the case of the MC-ICP-MS, the mass fraction of the primary solution used to prepare the calibration blends (1) and (2) is by far the largest contribution, even though the absolute uncertainty of this mass fraction is equal in both cases. The reason for this is the uncertainty associated with the intensity ratios, which is nearly 1 order of magnitude better (smaller) in the case of the MC-ICP-MS compared with the ICP OES. Another reason (which could open up the possibility for future improvements) is the much higher sensitivity of the MC-ICP-MS necessitating one more dilution step on the way to the preparation of the blends (1), (2), and (bx), resulting in larger masses with a smaller relative uncertainty. This finding means in turn that (because the Mo mass fraction w z is the limiting contribution and both the measurement itself and the preparation do not add that much) the MC-ICP-MS measurement procedure is capable of comparing two or more solutions to each other without limits caused by the measurement. Therefore, this method was applied for the validation of the dissolution procedure: several diluted primary solutions (w(E) ≈ 1 g/kg), prepared from different stock solutions, were measured against each other. The measured mass fractions were in agreement with the values calculated from the gravimetric preparation, validating the preparation procedure via the measurement procedure and vice versa. Hence, the basis for the preparation of commercial Mo and Rh 1 g/kg solutions with a conservatively estimated actual expanded uncertainty of U rel(w) ≤ 0.5 % has been established.
References
Güttler B, Richter W (2009) Traceability of chemical measurement results. Chimia 63:619–623
Bureau International des Poids et Mesures (2008) International vocabulary of metrology—basic and general concepts and associated terms (VIM). JCGM 200
Bureau International des Poids et Mesures (2006) The international system of units (SI), 8th edition
Kipphardt H, Matschat R, Rienitz O, Schiel D, Gernand W, Oeter D (2006) Traceability system for elemental analysis. Accred Qual Assur 10:633–639
Richter S, Sargent M, Schiel D, Kipphardt H (2013) Known purity—the fundament of element determination by atomic spectrometry. J Anal At Spectrom 28(10):1540–1543
Rienitz O, Schiel D, Görlitz V, Jährling R, Vogl J, Lara-Manzano JV, Zoń A, Fung W, Buzoianu M, Caciano de Sena R, Augusto dos Reis L, Valiente L, Yim Y, Hill S, Champion R, Fisicaro P, Bing W, Turk GC, Winchester MR, Saxby D, Merrick J, Hioki A, Miura T, Suzuki T, Linsky M, Barzev A, Máriássy M, Cankur O, Ari B, Tunç M, Konopelko LA, Kustikov YA, Bezruchko M (2012) CCQM-K87 “Mono-elemental calibration solutions”. Metrologia, Tech Suppl 49:08010
Jun W, Hongmei L, Liandi MA (2009) NIM’s role in developing national system of metrology in chemistry for food analysis, IMEKO World congress fundamental and applied metrology, 6–11, Lisbon, Portugal, 2583–2587
Kipphardt H, Czerwensky M, Matschat R (2005) ICP-MS analysis of high purity molybdenum used as SI-traceable standard of high metrological quality. J Anal At Spectrom 20:28–34
Felber H, Weber M, Rivier C (2002) Final report on key comparison CCQM-K8 of monoelemental calibration solutions. Metrologia,Tech Suppl 39:08002
Beck CM, Salit ML, Water RL Jr, Butler TA, Wood LJ (1993) Preparation and certification of a rhodium standard reference material solution. Anal Chem 85:2899–2902
Hauswaldt A-L, Rienitz O, Jährling R, Fischer N, Schiel D, Labarraque G, Magnusson B (2012) Uncertainty of standard addition experiments: a novel approach to include the uncertainty associated with the standard in the model equation. Accred Qual Assur 17:129–138
Serapinas P, Labarraque G, Charlet P, Ežerinskis Ž, Juzikienė V (2010) Method of standard additions for arsenic measurements in water by ICP sector field mass spectrometry at accuracy comparable to isotope dilution. J Anal At Spectrom 25:624–630
Salit ML, Turk GC (1998) A drift correction procedure. Anal Chem 70:3184–3190
Salit ML, Turk GC (2005) Traceability of single-element calibration solutions. Anal Chem 77:136 A–141 A
Salit ML, Turk GC, Lindstrom AP, Butler TA, Beck CM II, Norman B (2001) Single-element solution comparisons with a high-performance inductively coupled plasma optical emission spectrometric method. Anal Chem 73:4821–4829
Nater R, Reichmuth A, Schwartz R, Borys M, Zervos P (2009) Dictionary of weighing terms. Springer, Berlin
Spieweck F, Bettin H (1992) Review: solid and liquid density determination. Tech Mess 59:285–292
Bowman HA, Schoonover RM, Jones MW (1967) Procedure for high precision density determinations by hydrostatic weighing. J Res Natl Bur Stand C Eng Instrumen 71C:179–198
Bureau International des Poids et Mesures (2008) Evaluation of measurement data—guide to the expression of uncertainty in measurement, JCGM 100
ISO Guide 35:2006, Reference materials—general and statistical principles for certification
Acknowledgments
Financial Support by EMRP (the European Metrology Research Programme) is gratefully acknowledged (EMRP SIB09, “Primary Standards for Challenging Elements”). The EMRP is jointly funded by the EMRP participating countries within EURAMET and the European Union.
Author information
Authors and Affiliations
Corresponding author
Additional information
Published in the topical collection Reference Materials for Chemical Analysis with guest editors Hendrik Emons and Stephen A. Wise.
Appendix
Appendix
Bracketing plus internal standard method
In the case of a linear relationship of the measured signal intensity I of an element E or isotope I and the amount of substance n of this element or isotope carried into the measurement device (e.g., ICP OES), a straight line connecting the two standard solutions (bracket) can be assumed. Its parameters slope tanα and y-intercept R 0 can then be applied to calculate the mass fraction w x of the element in the original sample. If all solutions were additionally doped with an internal standard Y, having a mass fraction of w y, the straight line can be expressed via signal intensity and amount of substance ratios, respectively (Fig. 6).
In case the isotopic composition of the element of interest is the same in the sample and in the standard, the ratio of the molar masses cancels from the last equation.
In case a mass spectrometer is used and the element of interest has at least two isotopes, the amount of substance fractions x of the isotope of the analyte element measured in sample x and in standard z have to be taken into account.
Quantities, indices, and other symbols
Table 3 summarizes all quantities, indices, and additional symbols used throughout the document.
Rights and permissions
About this article
Cite this article
Kaltenbach, A., Noordmann, J., Görlitz, V. et al. Gravimetric preparation and characterization of primary reference solutions of molybdenum and rhodium. Anal Bioanal Chem 407, 3093–3102 (2015). https://doi.org/10.1007/s00216-014-8395-2
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00216-014-8395-2