Biomechanische Eigenschaften der Interferenzverschraubung beim Ersatz des vorderen Kreuzbandes mit Patellar- und Hamstring-Transplantaten

Eine experimentelle Studie mittels Röntgenstereometrieanalyse (RSA)

Biomechanical properties of patellar and hamstring graft fixation for anterior cruciate ligament reconstruction using an interference screw. Experimental study with roentgen stereometric analysis (RSA)

  • Originalien
  • Published:
Der Orthopäde Aims and scope Submit manuscript

Zusammenfassung

Mit Hilfe der Röntgenstereometrieanalyse (RSA) sollten bei Ersatzplastiken des vorderen Kreuzbandes (VKB) Relativbewegungen zwischen Transplantat und Bohrtunnel unter Last erfasst werden. Zielsetzung war die Bestimmung der Steifigkeit der Patellarsehneninterferenzverschraubung und der Vergleich mit der direkten Schraubenfixation von Hamstring-Transplantaten.

Die Untersuchungen erfolgten an frischen Präparaten vom Schwein. In 10 Tibiae wurde ein 10×25-mm-Bone-patellar-tendon-bone-(BPTB-)Transplantat entweder mit einer 7×25-mm Titan- oder einer 7×25-mm-Polylactidschraube in einem 10-mm-Bohrloch fixiert. In der Vergleichsgruppe erfolgte die direkte Verschraubung eines 4fachen Hamstring-Transplantats mit einer 7×25-mm-Polylactidschraube in einem 8-mm-Bohrkanal. Knochenblock, Sehne, Schraube und Tibia wurden mit RSA-Markierungen versehen. Die Zugbelastung wurde stufenweise bis zum Versagen des Systems gesteigert. Mittels der RSA wurden auf jeder Belastungsstufe die Transplantatbewegungen im Bohrkanal bestimmt. Die RSA-Messgenauigkeit betrug 0,05 mm.

Bei der direkten Verschraubung der 4fachen Hamstring-Sehne fand sich eine signifikant niedrigere maximale Haltekraft als bei der BPTB-Verschraubung (492±30 vs. 658±98 N). Die lineare Steifigkeit der Fixation war ebenfalls deutlich geringer (403±141 vs. 3500±1300 N/mm). Zwischen Polylactid- und Titanschraube fanden sich bei der BPTB-Verschraubung keine signifikanten Unterschiede. Bei den verschraubten Hamstring-Transplantaten zeigte sich bereits nach durchschnittlich 82±35 N Zuglast ein Nachgeben der Fixation (Elastizitätsgrenze). Die Elastizitätsgrenze der BPTB-Verschraubung lag mit durchschnittlich 428±135 N deutlich höher. Bei submaximaler Belastung mit 3/4 der maximalen Haltekraft fand sich beim Patellarsehnendrittel eine mittlere Migration in Zugrichtung von 0,36±0,25 vs. 2,58±1,08 mm beim Hamstring-Transplantat.

Die direkte Verschraubung der 4fachen Hamstring-Sehne zeigte im Vergleich zur Interferenzverschraubung der Patellarsehne eine geringe lineare Steifigkeit und eine niedrigere Elastizitätsgrenze. Hierdurch kann es bereits bei physiologischer Last zu einem Rutschen des Sehnentransplantats in Zugrichtung kommen.

Abstract

This experimental roentgen stereometric analysis (RSA) study was performed to measure micromotions between the graft and tunnel under submaximal load in anterior cruciate ligament (ACL) reconstruction. The aim was to evaluate the maximum load at failure, linear stiffness, and slippage of bone-patellar-tendon-bone (BPTB) grafts fixed with interference screws compared to quadrupled hamstring grafts fixed with interference screws.

We used 15 porcine tibia specimens for the study. In the BPTB group, the 10×25-mm bone plugs were fixed in a tunnel with 10 mm in diameter using a 7×25-mm titanium interference screw (n=5) or a biodegradable screw (n=5). The five hamstring transplants were folded to a four-stranded graft and armed with a baseball stitch suture. The sutured side was fixed with a 7×25-mm biodegradable polylactide screw in an 8-mm tunnel. The tibial bones, tendon grafts, and interference screws were marked with tantalum beads. The grafts were mounted to a custom made load frame and loaded parallel to the axis under RSA control increasing the force in steps of 50 N. Micromotions between bone plug, screw, and tibia were measured with RSA.

Accuracy of RSA for the in vitro study was evaluated as 0.05 mm. Hamstring grafts failed at significantly lower maximum loads (492±30 N) than BPTB grafts (658±98 N). Linear stiffness of the hamstring graft fixations was eight times lower compared to the BPTB grafts (403±141 N/mm vs 3500±1300 N/mm). There was no significant difference between the biodegradable and titanium screws in the BPTB group. Slippage of the graft started at 82±35 N load in the hamstring group and at 428±135 N in the BPTB group. Slippage of the graft at 75% of the maximum pull-out strength was measured as 0.36±0.25 mm in the BPTB and 2.58±1.08 mm in the hamstring group.

The interference screw fixation of a quadrupled hamstring graft showed a lower linear stiffness and an earlier slippage compared to a patellar tendon bone plug. Slippage of the hamstring grafts at submaximal loads may result in fixation failure during rehabilitation.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Adam, F., Pape, D., Steimer, O. et al. Biomechanische Eigenschaften der Interferenzverschraubung beim Ersatz des vorderen Kreuzbandes mit Patellar- und Hamstring-Transplantaten . Orthopäde 30, 649–657 (2001). https://doi.org/10.1007/s001320170053

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/s001320170053

Navigation