Skip to main content
Log in

Design features and surgical use of a cannulated extrusion needle

  • Clinical Investigations
  • Published:
Graefe's Archive for Clinical and Experimental Ophthalmology Aims and scope Submit manuscript

Abstract

A cannulated extrusion needle was designed as a modification of the standard 20-gauge extrusion by including an inner 24-gauge flexible cannula that can be extended up to 18 mm beyond the metal tip. Indications for use of the instrument are reviewed along with selected case histories demonstrating its use. The instrument is used primarily for internal subretinal fluid drainage in eyes with complex retinal detachments caused by proliferative vitreoretinopathy. The flexible cannula is guided into the subretinal space through an open peripheral retinal break and the subretinal fluid is evacuated during a simultaneous fluid/air exchange. The positive pressure of the automated air pump creates a pressure gradient sufficient to achieve passive egress of subretinal fluid through the cannula. Active suction is avoided because of the risk of posterior retinal incarceration into the cannula tip. In addition, the cannula can be used to remove hemorrhage, gas, or silicone oil in the subretinal space.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Charles S (1980) Vitrectomy for retinal detachment. Trans Ophthalmol Soc UK 100: 542–549

    Google Scholar 

  2. Charles S (1987) Vitreous microsurgery. Williams and Wilkins, Baltimore, p 84

    Google Scholar 

  3. Charles S, Wang C (1981) A motorized gas injector for vitreous surgery. Arch Ophthalmol 99: 1398

    Google Scholar 

  4. Doft BH (1986) Intentional retinotomy for internal drainage of subretinal fluid. Arch Ophthalmol 104: 807–808

    Google Scholar 

  5. Flynn HW, Blumenkranz MS, Parel JM, Lee WG (1987) Cannulated subretinal fluid aspirator for vitreoretinal microsurgery. Am J Ophthalmol 103: 106–108

    Google Scholar 

  6. Flynn HW, Davis JL, Parel JM, Lee WG (1988) Applications of a cannulated extrusion needle during vitreoretinal microsurgery. Retina 8: 42–49

    Google Scholar 

  7. Flynn HW, Lee WG, Parel JM (1988) A simple extrusion needle with flexible cannula tip for vitreoretinal microsurgery. Am J Ophthalmol 105: 215–216

    Google Scholar 

  8. Glaser BM (1986)) Treatment of giant retinal tears combined with proliferative vitreoretinopathy. Ophthalmology 93: 1193–1197

    Google Scholar 

  9. Gonvers M (1984) Retinal perforation for internal drainage. Am J Ophthalmol 97: 786–787

    Google Scholar 

  10. Hanneken AM, Michels RG (1988) Vitrectomy and scleral buckling methods for proliferative vitreoretinopathy. Ophthalmology 95: 865–869

    Google Scholar 

  11. Landers MD, Stefansson E, Wolbarsh ML (1981) The optics of vitreous surgery. Am J Ophthalmol 91: 611–614

    Google Scholar 

  12. McCuen BW, Bessler M, Hickingbotham D, Isbey E (1981) Automated Fluid-gas exchange. Am J Ophthalmol 95: 717

    Google Scholar 

  13. McDonald HR, Lewis H, Aaberg TM, Abrams GW (1989) Complications of drainage retinotomies used during vitreous surgery for complicated retinal detachment. Ophthalmology 96: 358–363

    Google Scholar 

  14. Michels RG (1984) Surgery of retinal detachment with proliferative vitreoretinopathy. Retina 4: 63–83

    Google Scholar 

  15. Parel JM, O'Grady G, Machemer R (1981) A bipolar coaxial microprobe for safe transvitreal diathermy. Arch Ophthalmol 99: 494–495

    Google Scholar 

  16. The Retina Society Terminology Committee (1983) The classification of retinal detachment with proliferative vitreoretinopathy. Ophthalmology 90: 121–125

    Google Scholar 

  17. The Silicone Study Group (1987) Manual of procedures. Silicone Study Coordinating Center, University of Southern California, Los Angeles, pp 10.1–10.12

    Google Scholar 

  18. Zivojnovic R (1987) Silicone oil in vitreoretinal surgery. Nijhoff/Junk, Dordrecht, pp 45–60

    Google Scholar 

  19. Zivojnovic R, Vijfvinkel GJ (1983) A modified flute needle. Am J Ophthalmol 96: 548–549

    Google Scholar 

  20. Zivojnovic R, Vijfvinkel GJ (1988) A brush back-flush needle. Arch Ophthalmol 106: 695

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Additional information

Presented at the XVIth Meeting of the Club Jules Gonin, Bruges, 4–8 September 1988

Supported in part by the Helena Rubinstein Foundation, New York, and in part by the Veteran's Administration Medical Center, Miami, Florida, USA; the authors have no commercial or financial interest in the cannulated extrusion needle

Rights and permissions

Reprints and permissions

About this article

Cite this article

Flynn, H.W., Lee, W.G. & Parel, JM. Design features and surgical use of a cannulated extrusion needle. Graefe's Arch Clin Exp Ophthalmol 227, 304–308 (1989). https://doi.org/10.1007/BF02169401

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02169401

Keywords

Navigation