Abstract
Infrequent control, aging of components, may compromise the accuracy of ICU ventilators. In order to assess the reliability of ventilators during their clinical use, we bench tested a group of 20 CPU1 ventilators (Ohmeda) sampled at random in several ICU units. We found major leaks in 5 ventilators, attributable to the disposable tubings used in these systems. Mean error in expired tidal volume and corresponding standard deviation (precision) were greater than 100 ml in two. Positive end expiratory pressure measurement comprised a mean error higher than 2 cm H2O in 40% of the ventilators tested. The valve opening pressure threshold was correlated to the inspiratory flow (r=0.81) contrary to the valve opening delay (average 138±40 ms). These two parameters did not correlate with the age of the ventilator. Our study addresses the need for periodic control of ventilator performance in order to minimize the risks of errors and malfunctions.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Nunn, JF, Lyle JR (1986) Bench testing of the CPU-1 ventilator. Br J Anaesth 58:653–662
Cox D, Tinloi SF, Farrimond JG (1988) Investigation of the spontaneous modes of breathing of different ventilators. Intensive Care Med 14:532–537
Martin JD, Rafferty JF, Wetzel RC, Gioia FR (1989) Inspiratory work and response times of a modified pediatric volume ventilator during synchronized intermittent mandatory ventilation and pressure support. Anesthesiology 71:977–981
Synnott AJ, Wren WS (1988) Assessment of the Ohio 5400 volum monitor Ohmeda SE 302 spirometer and the flow meters of the Servo 900 C ventilator. Acta Anesthesiol Scand 32:30–32
Bartel LP, Bazik JR, Powner DJ 81985) Compression volume during mechanical ventilation: comparison of ventilators and tubing circuits. Crit Care Med 13:851–854
Clero B, Atinault A (1984) Le respirateur CPU1. Agressologie 25:129–134
Nickerson BG, Sarkisian C, Tremper K (1988) Bias and precision of pulse oximeters and arterial oximeters. Chest 93:515–517
Beydon L, Chassé M, Harf A, Lemaire F (1988) Inspiratory work of breathing during spontaneous ventilation using demand valves and continuous flow systems. Am Rev Respir Dis 138:300–306
Schwartz D (1963) Méthodes statistiques à l'usage des médecins et biologistes, 3rd edn. Paris Flammarion, pp 312–320
Turney SZ, Blumenfeld W (1973) Heated pneumotachometer: a calibration procedure. J Appl Physiol 34:117–121
Sykes MK, Vickers MD, Hull CD, Winterburn PJ (1981) Principles of clinical measurements. Blackwell, Oxford, pp 182–201
Cara M, Chanu D (1984) Spirométrie en ventilation artificielle. Agressologie 25:93–104
Christopher KL, Neff TA, Bowman JL, Eberle DJ, Irvin CG, Good JT (1985) Demand and continuous flow intermittent mandatory ventilation systems. Chest 87:625–630
Katz JA, Kraemer RW, Gjerde GE (1985) Inspiratory work and airway pressure with continuous positive airway pressure delivery systems. Chest 88:519–526
Gibney RT, Wilson RS, Pontoppidan H (1982) Comparison of work of breathing on high gas flow and demand valve continuous positive airway pressure systems. Chest 82:692–695
Viale JP, Annat G, Bertrand O, Godard J, Motin J (1985) Additional inspiratory work in intubated patients breathing with continuous positive airway pressure systems. Anesthesiology 63:536–539
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Beydon, L., Liu, N., Hassapopoulos, J. et al. Test of 20 similar intensive care ventilators in daily use conditions —evaluation of accuracy and performances. Intensive Care Med 18, 32–37 (1992). https://doi.org/10.1007/BF01706423
Received:
Revised:
Issue Date:
DOI: https://doi.org/10.1007/BF01706423