Diagnostic accuracy of simultaneous acquisition of transmission and emission data with technetium-99m transmission source on thallium-201 myocardial SPECT

  • Original Articles
  • Published:
Annals of Nuclear Medicine Aims and scope Submit manuscript

Abstract

Purpose

This study evaluates not only the clinical usefulness but also the problems in attenuation correction for thallium-201 (Tl-201) myocardial SPECT by means of simultaneous transmission and emission data acquisition in the detection of coronary artery disease (CAD).

Methods

A three-detector SPECT system equipped with a Tc-99m line source and fan-beam collimators was used for simultaneous transmission and emission data acquisition for Tl-201 myocardial SPECT in 73 patients (18 patients for normal database and 55 patients for the evaluation of diagnostic accuracy). Attenuation-corrected (AC) images and non-attenuation-corrected (NC) images were reconstructed with an iterative maximum-likelihood estimation-corrected (ML-EM) algorithm. Both sets of images were reoriented into the short axis. Normal database polar maps were constructed from the AC and NC images for quantitative analysis.

Results

There was a significant difference in specificity between NC and AC images in the RCA territory and those in specificity and accuracy in the LCX territory. There was no significant difference in sensitivity found between NC and AC images in either territory, but sensitivity in both territories tended to decrease with attenuation correction. In the LAD territory, there were various changes in sensitivity and specificity observed with attenuation correction in cases with each quantitative criterion.

Conclusions

Diagnostic performance of significant stenosis in the RCA and LCX territories quantitatively improved with attenuation correction because of an increase in specificity, but no significant improvement in diagnostic performance was obtained in the LAD territory with attenuation correction. We recommend combined interpretation of AC and NC images and careful evaluation of any SPECT image by means of transmission computed tomography.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Similar content being viewed by others

References

  1. DePasquale E, Nody C, DePuey G, Garcia E, Pilcher G, Bredlau C, et al. Quantitative rotational thallium-201 tomography for identifying and localizing coronary artery disease.Circulation 1988; 77: 316–327.

    PubMed  CAS  Google Scholar 

  2. Mahmarian JJ, Boyce TM, Goldberg RK, Cocanougher MK, Roberts BR, Verani MS. Quantitative exercise thallium-201 single-photon emission computed tomography for the enhanced diagnosis of ischemic heart disease.J Am Coll Cardiol 1990; 15: 318–329.

    Article  PubMed  CAS  Google Scholar 

  3. Tsui BMW, Gullberg GT, Edgerton ER, Ballard JG, Perry JR, McCartney WH, et al. Correction of nonuniform attenuation in cardiac SPECT imaging.J Nucl Med 1989; 30: 497–507.

    PubMed  CAS  Google Scholar 

  4. Frey EC, Tsui BMW, Perry JR. Simultaneous acquisition of emission and transmission data for improved Thallium-201 cardiac SPECT imaging using a Technetium-99m transmission source.J Nucl Med 1992; 33: 2238–2245.

    PubMed  CAS  Google Scholar 

  5. Tung CH, Gullberg GT, Zeng GL, Christian PE, Datz FL, Morgan HT. Non-uniform attenuation correction using simultaneous transmission and emission converging tomography.IEEE Trans Nucl Sci 1992; 39: 1134–1143.

    Article  CAS  Google Scholar 

  6. Jaszczak RJ, Gilland DR, Hanson MW, Jang S, Greer KL, Coleman RE. Fast transmission CT for determining attenuation maps using a collimated line source, rotatable air-copper-lead attenuators and fan-beam collimation.J Nucl Med 1993; 34: 1577–1586.

    PubMed  CAS  Google Scholar 

  7. Ficaro EP, Fessler JA, Rogers WL, Schwaiger M. Comparison of Americium-241 and Technetium-99m as transmission sources for attenuation correction of thallium-201 SPECT imaging of the heart.J Nucl Med 1994; 35: 652–663.

    PubMed  CAS  Google Scholar 

  8. Ficaro EP, Fessler JA, Ackermann RJ, Rogers WL, Corbett JR, Schwaiger M. Simultaneous transmission-emission Thallium-201 cardiac SPECT: effect of attenuation correction on myocardial tracer distribution.J Nucl Med 1995; 36: 921–931.

    PubMed  CAS  Google Scholar 

  9. Ficaro EP, Fessler JA, Shreve PD, Kritzman JN, Rose PA, Corbett JR. Simultaneous transmission/emission myocardial perfusion tomography: diagnostic accuracy of attenuation corrected Tc-99m sestamibi single-photon emission computed tomography.Circulation 1996; 93: 463–473.

    PubMed  CAS  Google Scholar 

  10. Vidal R, Buvat I, Darcourt J, Migneco O, Desvignes P, Baudouy M, et al. Impact of attenuation correction by simultaneous emission/transmission tomography on visual assessment of Tl-201 myocardial perfusion images.J Nucl Med 1999; 40: 1301–1309.

    PubMed  CAS  Google Scholar 

  11. Miron SD, Conant RG, Sodee DB, Amini SB. Clinical evaluation of simultaneous transmission-emission SPECT myocardial perfusion images (STEP). [Abstract]J Nucl Med 1995; 36: 12P.

    Google Scholar 

  12. Meikle SR, Hutton BF, Bailey DL. A transmission-dependent method for scatter correction in SPECT.J Nucl Med 1994; 35: 360–367.

    PubMed  CAS  Google Scholar 

  13. King MA, Xia W, DeVries DJ, Pan TS, Villegas BJ, Dahlberg S, et al. A Monte Carlo investigation of artifacts caused by liver uptake in single-photon emission computed tomography perfusion imaging with technetium 99m-labeled agents.J Nucl Cardiol 1996; 3 (1): 18–29.

    Article  PubMed  CAS  Google Scholar 

  14. Prvulovich EM, Lonn AHR, Bomanji JB, Jarritt PH, Ell PJ. Effect of attenuation correction on myocardial thallium-201 distribution in patients with a low likelihood of coronary artery disease.Eur J Nucl Med 1997; 24: 266–275.

    PubMed  CAS  Google Scholar 

  15. Maniawski PJ, Miller SD, Morgan HT. Combined attenuation and scatter correction significantly reduces the effect of extra-cardiac activity on Tl-201 myocardial perfusion SPECT. [Abstract]J Nucl Med 1996; 37: 214P.

  16. Case JA, King MA, Dalhberg ST. Results of attenuation and scatter correction in clinical cardiac Tl-201 SPECT perfusion imaging. [Abstract]J Nucl Med 1996; 37: 212P.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Ohyama, Y., Tomiguchi, S., Kira, T. et al. Diagnostic accuracy of simultaneous acquisition of transmission and emission data with technetium-99m transmission source on thallium-201 myocardial SPECT. Ann Nucl Med 15, 21–26 (2001). https://doi.org/10.1007/BF03012126

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF03012126

Key words

Navigation