Abstract
Ongoing discussions about the origin of elemental fractionation occurring during LA-ICP-MS analysis show that this problem is still far from being well understood. It is becoming accepted that all three possible sources (ablation, transport, excitation) contribute to elemental fractionation. However, experimental data about the vaporisation size limit of different particles in the ICP, as produced in laser ablation, have not been available until now. This information should allow one to determine the signal contributing mass within the ICP and would further clarify demands on suitable laser ablation systems and gas atmospheres in terms of their particle size distribution.
The results presented here show a vaporisation size limit of laser induced particles, which was found at particle sizes between 90 nm and 150 nm using an Elan 6000 ICP-MS. Due to the fact that the ICP-MS response was used as evaluation parameter, vaporisation and ionisation limits are not distinguishable.
The upper limit was determined by successively removing the larger particles from the aerosol, which was created by ablation of a NIST 610 glass standard at a wavelength of 266 nm, using a recently developed particle separation device. Various particle fractions were separated from the aerosol entering the ICP. The decrease in signal intensity is not proportional to the decrease in volume, indicating that particles above 150 nm in diameter are not completely ionised in the ICP. Due to the limited removal range of the particle separation device, which cannot remove particles smaller than 150 nm, single hole ablations were used to determine the lower vaporisation limit. This is based on measurements showing that larger particles occur dominantly during the first 100 laser pulses only. After this period, the ratio of ICP-MS counts and total particle volume was found to be constant while most of the particles are smaller than 90 nm, indicating complete vaporisation and ionisation of these particles.
To describe the influence of different plasma forward powers on the vaporisation limit, the range 1000–1600 W was studied. Results indicate that optimum vaporisation and ionisation occurs at 1300 W. However, an increase of the particle ionisation limit towards larger particles was not observed within the accuracy of this study using the full range of parameters available for optimisation on commonly used ICP-MS instruments.
Similar content being viewed by others
References
Becker JS (2002) J Anal At Spectrom 17:1172–1185
Jeffries TE, Perkins WT, Pearce NJG (1995) Analyst 120:1365–1371
Günther D, Frischknecht R, Heinrich CA, Kahlert HJ (1997) J Anal At Spectrom 12:939–944
Russo RE, Mao XL, Gonzalez JJ, Mao SS (2002) J Anal At Spectrom 17:1072–1075
Eggins SM, Kinsley LPJ, Shelley JMG (1998) Appl Surf Sci 129:278–286
Bleiner D, Günther D (2001) J Anal At Spectrom 16:449–456
Günther D (2002) Anal Bioanal Chem 372:31–32
Rodushkin I, Axelsson MD, Malinovsky D, Baxter DC (2002) J Anal At Spectrom 17:1223–1230
Rodushkin I, Axelsson MD, Malinovsky D, Baxter DC (2002) J Anal At Spectrom 17:1231–1239
Stirling CH, Lee DC, Christensen JN, Halliday AN (2000) Geochim Cosmochim Acta 64:3737–3750
Figg D, Kahr MS (1997) Appl Spectrosc 51:1185–1192
Guillong M, Günther D (2002) J Anal At Spectrom 17:831–837
Kuhn H-R, Günther D (2003) Anal Chem 75:747–753
Cromwell EF, Arrowsmith P (1995) Anal Chem 67(1):131–138
Jeong SH, Borisov OV, Yoo JH, Mao XL, Russo AE (1999) Anal Chem 71:5123–5130
Olesik JW, Dziewatkoski MP, Mcgowan GJ (1994) Abstr Pap Am Chem Soc 208:147-ANYL
Knight K, Chenery S, Zochowski SW, Thompson M, Flint CD (1996) J Anal Atom Spectrom 11(1):53–56
Horn I, Guillong M, Günther D (2001) Appl Surf Sci 182:91–102
Guillong M, Horn I, Günther D (2003) J Anal At Spectrom 18(10):1224–1230
Jackson SE, Günther D (2003) J Anal At Spetrom 18:205–212
Aeschliman DB, Houk RS (2003) J Anal At Spectrom (Advance Article)
Guillong M, Kuhn H-R, Günther D (2002) Spectrochim Acta Part B Atom Spectrosc 58:211–220
Horn I, Günther D (2003) Appl Surf Sci 207:144–157
Fryer BJ, Jackson SE, Longerich HP (1995) Can Mineral 33:303–312
Acknowledgements
This work was supported by ETH Zürich and the Swiss National Science Foundation.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Kuhn, HR., Guillong, M. & Günther, D. Size-related vaporisation and ionisation of laser-induced glass particles in the inductively coupled plasma. Anal Bioanal Chem 378, 1069–1074 (2004). https://doi.org/10.1007/s00216-003-2346-7
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00216-003-2346-7